Aution™ Hybrid AU-4050 Chemical Analyzer 14603Recalls & reported problems
Arkray USA, Inc. · he AUTION HYBRID AU-4050 Fully Automated Integrated Urine Analyzer System contains a test strip chemistry urine analyzer and a flow cytometry urine particle analyzer together in a single integrated device. The test strip chemistry module (CHM) is an automated urine analyzer intended for the in vitro measurement of the following parameters: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, specific gravity, turbidity, and color. The chemistry module is intended for use with the Uriflet S 9HA multi-parameter urine chemistry test strips. The flow cytometry module (FCM) is an automated urine particle analyzer intended to analyze the following parameters in urine samples: Red Blood Cells, White Blood Cells, Epithelial Cells, Casts, and Bacteria and flags the presence of the following: Pathological Casts, Crystals, Sperm, Small Round Cells, Yeast Like Cells, and Mucus. The AUTION HYBRID AU-4050 is intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00015482146030), received between Feb 11, 2019 and Mar 29, 2019 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Patient Data Problem” (4 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 3, 2019Class IITerminated
This correction is being initiated due to a software issue which results in the possibility of incorrect patient information being assigned to sample results when the following rare combination of three specific events were to occur: 1) No measurement results are generated due to an error by a urine sediment measurement. 2) The instrument is shutdown incorrectly. 3) An item rack is used for subsequent sample measurements.
Z-1850-2019 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Patient Data Problem4
Patient problems
- No Known Impact Or Consequence To Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 29, 2019Patient Data Problem; Patient Data Problem
PATIENT HAD HER DEMOGRAPHICS ASSIGNED TO ANOTHER PATIENT'S SAMPLE RESULTS. SAMPLES WERE RUN WITHOUT QUERY TO THE LIS MEANING NO ORDER WAS RECEIVED FROM THE LIS AND THEREFORE NO ORDERS WITH PATIENT DEMOGRAPHICS. ON (B)(6) 2018 IT WAS FOUND THAT THREE SAMPLES WITHOUT QUERY AND THEREFORE ORDERS STILL HAD DEMOGRAPHICS ASSOCIATED WITH THEM ON BOTH THE ANALYZER PRINTOUT AND THE LIS TRANSMISSION AND IN ALL THREE CASES THOSE DEMOGRAPHICS WERE FOR THE INCORRECT PATIENTS (THREE PATIENTS LISTED ABOVE). IN NONE OF THESE CASES WERE THE PATIENTS MISDIAGNOSED OR MISTREATED. THE PATIENT RESULTS WERE CORRECTLY TIED TO THE CORRECT PATIENTS AT THE LIS BECAUSE THE LIS ALREADY HAS THE CORRECT DEMOGRAPHIC INFORMATION. DEMOGRAPHIC INFORMATION SENT BY THE ANALYZER WITH THE RESULTS IS IGNORED BY THE LIS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8466042 ↗MalfunctionMar 29, 2019Patient Data Problem; Patient Data Problem
PATIENT HAD HER DEMOGRAPHICS ASSIGNED TO ANOTHER PATIENT'S SAMPLE RESULTS. SAMPLES WERE RUN WITHOUT QUERY TO THE LIS MEANING NO ORDER WAS RECEIVED FROM THE LIS AND THEREFORE NO ORDERS WITH PATIENT DEMOGRAPHICS. ON (B)(6) 2018 IT WAS FOUND THAT THREE SAMPLES WITHOUT QUERY AND THEREFORE ORDERS STILL HAD DEMOGRAPHICS ASSOCIATED WITH THEM ON BOTH THE ANALYZER PRINTOUT AND THE LIS TRANSMISSION AND IN ALL THREE CASES THOSE DEMOGRAPHICS WERE FOR THE INCORRECT PATIENTS (THREE PATIENTS LISTED ABOVE). IN NONE OF THESE CASES WERE THE PATIENTS MISDIAGNOSED OR MISTREATED. THE PATIENT RESULTS WERE CORRECTLY TIED TO THE CORRECT PATIENTS AT THE LIS BECAUSE THE LIS ALREADY HAS THE CORRECT DEMOGRAPHIC INFORMATION. DEMOGRAPHIC INFORMATION SENT BY THE ANALYZER WITH THE RESULTS IS IGNORED BY THE LIS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8466034 ↗MalfunctionFeb 11, 2019Patient Data Problem; Patient Data Problem
PATIENT HAD HIS DEMOGRAPHICS ASSIGNED TO ANOTHER PATIENT'S SAMPLE RESULTS. SAMPLES WERE RUN WITHOUT QUERY TO THE LIS MEANING NO ORDER WAS RECEIVED FROM THE LIS AND THEREFORE NO ORDERS WITH PATIENT DEMOGRAPHICS. ON (B)(6) 2018 IT WAS FOUND THAT THREE SAMPLES WITHOUT QUERY AND THEREFORE ORDERS STILL HAD DEMOGRAPHICS ASSOCIATED WITH THEM ON BOTH THE ANALYZER PRINTOUT AND THE LIS TRANSMISSION AND IN ALL THREE CASES THOSE DEMOGRAPHICS WERE FOR THE INCORRECT PATIENTS (THREE PATIENTS LISTED ABOVE). IN NONE OF THESE CASES WERE THE PATIENTS MISDIAGNOSED OR MISTREATED. THE PATIENT RESULTS WERE CORRECTLY TIED TO THE CORRECT PATIENTS AT THE LIS BECAUSE THE LIS ALREADY HAS THE CORRECT DEMOGRAPHIC INFORMATION. DEMOGRAPHIC INFORMATION SENT BY THE ANALYZER WITH THE RESULTS IS IGNORED BY THE LIS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8326305 ↗MalfunctionFeb 11, 2019Patient Data Problem; Patient Data Problem
PATIENT HAD HER DEMOGRAPHICS ASSIGNED TO ANOTHER PATIENT'S SAMPLE RESULTS. THE CUSTOMER CLAIMS THE SAME INCORRECT PATIENT DEMOGRAPHICS ARE SHOWN IN THE IPU SOFTWARE. PER THE CUSTOMER NO ACTION WAS TAKEN IN TREATMENT OF THE PATIENT INCORRECTLY IDENTIFIED BASED ON THESE RESULTS HAVING THE PATIENT'S DEMOGRAPHICS TIED TO THEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8326272 ↗
Other Automated Urinalysis System devices
- iQ200 Elite Plus (US) (2008 Version) (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls2 reports
- iQ200 Sprint Pro (NA)/SPRINT w load/unload (NA)/SPRINT Pro w load/unload (NA)/Sprint PBeckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls1 report
- iQ200 Select/Elite Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls0 reports
- iQ200 Select Plus Kit 2008 Analyzer (NA)/ Select Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System2 recalls4 reports
- iChem Velocity Back-Integrated Velocity (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System1 recall15 reports
- iChem Velocity Elite/Stand-Alone Velocity (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System1 recall0 reports
- UA Win10 Kit C52900Beckman Coulter Ireland Inc. · Automated Urinalysis System1 recall0 reports
- cobas u 411 Urine Analyzer 4906969001Roche Diagnostics GmbH · AUTOMATED URINALYSIS SYSTEM16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page