Skip to content

PC 55-01-0017Recalls & reported problems

Allied Medical, LLC · LITHOLYME,GEMULTISTYLE,6/CS · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00026072025260), received between Aug 26, 2025 and Mar 26, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2026Gas/Air Leak

    WHEN CHANGING THE CO2 ABSORBENT, IT BECOMES DIFFICULT TO SEAL. RESULTS IN A CIRCUIT LEAK. VERY DANGEROUS TO PATIENT. FLOW SENSORS O2 RINGS FELL OUT OF MACHINE ALONG WITH THE ABSORBENT CONTAINERS, CAUSING CRNA TO MANUALLY VENTILATE THE PATIENT UNTIL THE ANES. TECH WAS ABLE TO PUT PART BACK IN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24707078 ↗
  • MalfunctionAug 26, 2025Gas/Air Leak; Gas/Air Leak

    AS HAS BEEN REPORTED BEFORE, THE ANESTHESIA MACHINE HAD A MAJOR LEAK HALFWAY THROUGH A CARDIAC CASE. THIS IS DUE TO THE CO2 ABSORBERS NOT BEING COMPATIBLE WITH THE ANESTHESIA MACHINES THAT DIVISIONAL SUPPLY CHAIN HAD DECIDED TO CHANGE WITHOUT CONSULTATION OF THE ANESTHESIA DEPARTMENT. IN THIS EVENT THE ANESTHESIA GAS FLOWS HAD TO BE ADJUSTED SIGNIFICANTLY UNTIL THE PATIENT WAS ABLE TO BE ON CARDIOPULMONARY BYPASS SO THE MACHINE COULD BE CHANGED OUT SINCE THE LEAK COULD NOT BE CHECKED AND THE MACHINE COULD NOT BE TROUBLESHOOTED WHILE A PATIENT WAS ACTIVELY CONNECTED TO THE MACHINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22900484 ↗

Other Absorbent, Carbon-Dioxide devices

All Absorbent, Carbon-Dioxide devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page