Truform® 8848BG-SRecalls & reported problems
Surgical Appliance Industries, Inc. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Oct 7, 2011Terminated
During an audit with the FDA, it was discovered that a new compression testing device used for testing hosiery had been implemented and that the operators were not properly trained. Additionally, the acceptance criteria for the new device were not changed from the old testing device to reflect additional data (compression profile percentages). This resulted in the device history files containing s
Z-0391-2012 · Root cause: Employee error · Recall details · FDA record ↗
Other Stocking, Medical Support (To Prevent Pooling Of Blood In Legs) devices
- Truform® 0868BL-XLSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
- Truform® 8846-SSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
- Truform® 8868BG-LSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
- Truform® 1757BG-MSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
- Truform® 8868BL-LSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
- Truform® 0868BL-SSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
- Truform® 8864BG-SSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
- Truform® 8867BG-MSurgical Appliance Industries, Inc. · Stocking, Medical Support (To Prevent Pooling Of Blood In Legs)1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page