CURAD CUR8106Recalls & reported problems
Medline Industries, Inc. · GLOVE,EXAM,LATEX,CURAD,PF,LG · Class 1
All models of CURAD (85)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00080196319162), received between Sep 1, 2026 and Sep 1, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Puncture/Hole” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2026Material Puncture/Hole; Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50543064 ↗
Other Latex patient examination glove devices
- CURAD CUR8105Medline Industries, Inc. · Latex patient examination glove3 reports
- CURAD CUR8104Medline Industries, Inc. · Latex patient examination glove3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page