Skip to content

Medline MDS92250Recalls & reported problems

Medline Industries, Inc. · STETHOSCOPE, ADULT, STNLSS STL,GRY · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00080196326313), received between Sep 2, 2025 and Sep 2, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Damaged Thread” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Damaged Thread1
  • Sharp Edges1

Patient problems

  • Laceration(s)1
  • Perforation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 2, 2025Sharp Edges; Damaged Thread; Sharp Edges; Damaged Thread

    ACCORDING TO THE CUSTOMER, ON (B)(6) 2025, THE USER PLACED THE STETHOSCOPE TO THEIR EARS WITHOUT KNOWING THE "EAR BUD HAD FALLEN OFF" RESULTING IN A LACERATION OF THE LEFT EAR CANAL AND PERFORATED TYMPANIC MEMBRANE. THE CUSTOMER STATED THAT THE USER WENT TO URGENT CARE AND AN ENT PROVIDER FOR TREATMENT AND COMPLAINED OF SEVERE PAIN, RINGING IN THE EAR, MUFFLED HEARING, AND BLEEDING.

    Patient problems: Laceration(s); Perforation

    MAUDE report 22936859 ↗

Other STETHOSCOPE, MANUAL devices

All STETHOSCOPE, MANUAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page