Medline MDS9410Recalls & reported problems
Medline Industries, Inc. · ANEROID,HANDHELD,PREMIER,ADULT · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00080196773193), received between Sep 12, 2026 and Sep 30, 2026 — classified as 30 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak21
- Inflation Problem7
- Calibration Problem2
- Incorrect, Inadequate or Imprecise Result or Readings2
- Material Puncture/Hole2
- Deflation Problem1
- High Readings1
- Incorrect Measurement1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Gas/Air Leak
It was reported that customer received defective blood pressure cuffs.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951609 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
It was reported that "the newer blood pressure cuffs we are using are going out at a quicker rate than they used to" and "The cuffs will still look brand new but will have issues like: Leaking bladder, Meters out of configuration, The screw ball that keeps the air pressure will release at a faster than normal rate so they¿re unable to capture or ensure accuracy of the readings".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946830 ↗MalfunctionSep 30, 2026Deflation Problem; Inflation Problem; Gas/Air Leak
It was reported that "BP pressure cuff not deflaiting or inflating to it normal standards. Air stays in the cuff, and it does not inflate to its standard max pressure".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946030 ↗MalfunctionSep 29, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933007 ↗MalfunctionSep 25, 2026Material Puncture/Hole
It was reported that "THIS ARRIVED WITH A HOLE IN THE TUBING."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50904163 ↗MalfunctionSep 24, 2026Calibration Problem
It was reported that "THE NEEDLE EITHER STICKS OR THE UNIT DOESNT PUMP UP."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889211 ↗MalfunctionSep 23, 2026Inflation Problem; Gas/Air Leak
It was reported that "CUST THINKS THAT THERE MAY BE A HOLE, IT DOESN'T STAY INFLATED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872064 ↗MalfunctionSep 23, 2026Calibration Problem
It was reported that "item is not calibrating correctly."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868556 ↗MalfunctionSep 23, 2026Gas/Air Leak
It was reported that "None of them are working correctly" and "They are not holding air." "At least one of them has a noticeable hole in the tubing, not sure if there are pin point holes in the others."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50867987 ↗MalfunctionSep 22, 2026Incorrect Measurement
It was reported that BP GAUGE ON CUFF DOES NOT WORK, DOES NOT READ BP CORRECTLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50857653 ↗MalfunctionSep 21, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50842004 ↗MalfunctionSep 21, 2026Inflation Problem; Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50840496 ↗MalfunctionSep 21, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50838535 ↗MalfunctionSep 18, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829005 ↗MalfunctionSep 17, 2026Gas/Air Leak
It was reported that customer received defective blood pressure cuffs.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50806221 ↗MalfunctionSep 17, 2026Inflation Problem; Gas/Air Leak
It was reported that "The nurse went to blow the BP Cuff up and it is \"blown out.\" The nurse said the BP guage should be at 0 and it is reading 280 right now, won't pump up."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50804932 ↗MalfunctionSep 17, 2026Gas/Air Leak
It was reported that "I have an adult BP cuff item # (b)(6) that has a leak and is unable to hold air."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803325 ↗MalfunctionSep 17, 2026Gas/Air Leak
It was reported that "The bladders are leaking on all of them."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803204 ↗MalfunctionSep 16, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50794943 ↗MalfunctionSep 16, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50793830 ↗MalfunctionSep 16, 2026Material Puncture/Hole
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50793792 ↗MalfunctionSep 16, 2026High Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50792724 ↗MalfunctionSep 15, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50779398 ↗MalfunctionSep 15, 2026Incorrect, Inadequate or Imprecise Result or Readings; Material Split, Cut or Torn
It was reported that "PRODUCT NO LONGER HOLDING AIR WHEN USING. PRODUCT WILL DEFLATE AND NOT TAKE ACCURATE BLOOD PRESSURE READING."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50771474 ↗MalfunctionSep 15, 2026Gas/Air Leak
It was reported that bp cuff is not inflating or providing correct readings; stuck at 40 when completely deflated.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50766390 ↗
Other blood pressure cuff devices
- Medline MDS9380Medline Industries, Inc. · blood pressure cuff108 reports
- ONE-CUF ONE-A2-2AGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall1 report
- ONE-CUF ONE-A3-2AGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- Radius VSM 4826Masimo Corporation · blood pressure cuff1 recall0 reports
- ONE-CUF ONE-A2-2A-LGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- ONE-CUF ONE-A3-1BGE Medical Systems Information Technologies, Inc. · Blood Pressure Cuff1 recall0 reports
- Radius VSM 4825Masimo Corporation · blood pressure cuff1 recall0 reports
- Welch Allyn, Inc. 901091Welch Allyn, Inc. · blood pressure cuff1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page