MENTOR MemoryGel Breast Implant 3542501Recalls & reported problems
Mentor Texas L.P. · SILTEX Round Moderate Plus Profile Gel-Filled Breast Implant, 250cc |Textured|Round|Silicone| · Class 3 · Implant
All models of MENTOR MemoryGel Breast Implants (368) · MENTOR MemoryGel Breast Implant (211)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Mentor MemoryGel gel-filled breast implant has 105 reports, no deaths and 98 injuries. Many reports describe the implant rupturing, and some describe capsular contracture, where scar tissue tightens around the implant, and deformity. Several patients are reported to have had their implants removed or replaced. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 101
- Deaths reported
- 0
- Injuries reported
- 94
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 101 adverse event reports that list this exact device (DI 00081317000846), received between May 17, 2019 and Jan 7, 2026 — classified as 94 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (59 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture59
- Adverse Event Without Identified Device or Use Problem43
- Appropriate Device Problem Term/Code Not Available2
- Device Contamination with Chemical or Other Material1
- Device Ingredient or Reagent Problem1
- Migration1
Patient problems
- Capsular Contracture47
- Deformity/ Disfigurement28
- No Code Available19
- Seroma6
- Breast Discomfort/Pain5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- No Consequences Or Impact To Patient4
- Pain4
Recent adverse event reports
The 25 most recent of 101 reports, in the reporter's own words as published by the FDA.
InjuryJan 7, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 250CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA ULTRASOUND, WITH BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE III). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2025. THE REPLACEMENT DEVICES WERE TWO MENTOR GEL IMPLANTS. THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.
Patient problems: Capsular Contracture
MAUDE report 23995598 ↗InjuryJan 7, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 250CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA ULTRASOUND, WITH BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE III). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2025. THE REPLACEMENT DEVICES WERE TWO MENTOR GEL IMPLANTS. THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS.
Patient problems: Capsular Contracture
MAUDE report 23995597 ↗InjuryNov 23, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA ULTRASOUND, WITH LEFT BREAST IMPLANT RUPTURE AND RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE II). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2025. THE REPLACEMENT DEVICES WERE TWO MENTOR GEL IMPLANTS. THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.
Patient problems: Deformity/ Disfigurement
MAUDE report 23627097 ↗InjuryJul 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 250CC MENTOR MEMORYGEL BREAST IMPLANT EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE 3 ON THE LEFT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Capsular Contracture
MAUDE report 22555883 ↗InjuryJun 20, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A 51-YEAR-OLD HISPANIC FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH TWO UNSPECIFIED GEL BREAST PROSTHESES AND EXPERIENCED BILATERAL RUPTURES POST-OPERATIVELY, WHICH WAS DIAGNOSED MAMMOGRAM AND ULTRASOUND. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE LEFT SIDE DEVICE.
Patient problems: Capsular Contracture; Cyst(s); Deformity/ Disfigurement
MAUDE report 22302392 ↗InjuryJun 20, 2025Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 51-YEAR-OLD HISPANIC FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH TWO UNSPECIFIED GEL BREAST PROSTHESES AND EXPERIENCED BILATERAL RUPTURES POST-OPERATIVELY, WHICH WAS DIAGNOSED MAMMOGRAM AND ULTRASOUND. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE RIGHT SIDE DEVICE.
Patient problems: Capsular Contracture; Deformity/ Disfigurement
MAUDE report 22302386 ↗InjuryMay 15, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 400CC MENTOR MEMORYGEL BREAST IMPLANT EXPERIENCED BREAST PAIN AND A RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Breast Discomfort/Pain
MAUDE report 22028888 ↗InjuryMay 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 200CC MENTOR MEMORYGEL BREAST IMPLANT ON THE LEFT SIDE, AND WITH 250CC MENTOR MEMORYGEL BREAST IMPLANT ON THE RIGHT SIDE AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE; BAKER GRADE III. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REPLACEMENT SURGERY WITH CATALOG NUMBER SMPB290; SERIAL NUMBER (B)(6) ON THE LEFT SIDE, AND WITH CATALOG NUMBER SMPB375; SERIAL NUMBER (B)(6) ON THE RIGHT SIDE ON (B)(6) 2025. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE.
Patient problems: Capsular Contracture
MAUDE report 22001977 ↗InjuryMar 10, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 250CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED A RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2024, AND REPLACED WITH A 350CC MENTOR MEMORYGEL BREAST IMPLANT (CATALOG NUMBER 3503504BC) WITH SERIAL NUMBER (B)(6).
Patient problems: Deformity/ Disfigurement
MAUDE report 21562484 ↗InjuryMar 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION WITH MENTOR MEMORYGEL BREAST IMPLANTS 250CC AND EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE III ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT REMOVAL ON (B)(6) 2024.
Patient problems: Capsular Contracture
MAUDE report 21562005 ↗InjuryJan 31, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A 42-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH 250CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED BREAST IMPLANT RUPTURE ON HER LEFT SIDE POSTOPERATIVELY; DIAGNOSED IN THE OFFICE. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AS A RESULT, THE PATIENT IS SCHEDULED FOR SURGERY ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 21282324 ↗InjuryJan 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 79-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 250CC MENTOR MEMORYGEL BREAST IMPLANT ON BOTH SIDES AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE; BAKER GRADE III, AND RIGHT-SIDE BREAST IMPLANT RUPTURE POSTOPERATIVELY. AS A RESULT, THE PATIENT IS SCHEDULED FOR BILATERAL REPLACEMENT SURGERY ON (B)(6) 2025. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT-SIDED DEVICE.
Patient problems: Capsular Contracture
MAUDE report 21230347 ↗InjuryOct 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST RECONSTRUCTION WITH TWO MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT SUFFERED BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE IV ON THE RIGHT AND BAKER GRADE LEFT ON THE III). AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS.
Patient problems: Capsular Contracture
MAUDE report 20516894 ↗InjuryOct 15, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION WITH A 250CC MENTOR MEMORYGEL BREAST IMPLANT ON THE LEFT BREAST. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA ULTRASOUND, WITH LEFT BREAST IMPLANT RUPTURE. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024.
Patient problems: Deformity/ Disfigurement
MAUDE report 20448180 ↗InjuryOct 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST SURGERY WITH IMPLANTATION OF A 250CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT REPORTED EXPERIENCING A FIRM RIGHT BREAST. THE PATIENT REFUSED TO PROVIDE FURTHER INFORMATION. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THE DATE OF EVENT WAS PROVIDED TO MENTOR AS A BEST ESTIMATE.
Patient problems: Capsular Contracture
MAUDE report 20408699 ↗InjuryJul 8, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED BREAST SURGERY WITH A MENTOR MEMORYGEL BREAST IMPLANT 250CC GEL BREAST PROSTHESIS (LEFT DEVICE) AND A MENTOR MEMORYGEL BREAST IMPLANT 150CC GEL BREAST PROSTHESIS (RIGHT DEVICE) WHEN THE LEFT DEVICE RUPTURED POST IMPLANTATION. ADDITIONALLY, THE PATIENT DEVELOPED BAKER GRADE IV CAPSULAR CONTRACTURE IN BOTH BREASTS POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION WITH NO REPLACEMENT BREAST PROSTHESES ON (B)(6) 2024. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT BREAST PROSTHESIS.
Patient problems: Capsular Contracture
MAUDE report 19695055 ↗InjuryJun 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 250CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERAL BAKER GRADE III CAPSULAR CONTRACTURE OF THE BREASTS BY A MEDICAL PROFESSIONAL. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH 325CC MENTOR MEMORYGEL BREAST IMPLANTS ON (B)(6) 2024.
Patient problems: Capsular Contracture; Seroma
MAUDE report 19479461 ↗InjuryJun 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 250CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERAL BAKER GRADE III CAPSULAR CONTRACTURE OF THE BREASTS BY A MEDICAL PROFESSIONAL. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH 325CC MENTOR MEMORYGEL BREAST IMPLANTS ON (B)(6) 2024.
Patient problems: Capsular Contracture; Seroma
MAUDE report 19479456 ↗MalfunctionJun 3, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A 250CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS WAS UNBOXED AND OBSERVED TO BE DAMAGE DURING A BREAST SURGERY PRIOR TO BEING IMPLANTED INTO THE PATIENT. HOWEVER, THE SURGERY COMMENCED WITHOUT ANY REPORTED DELAYS AND A NEW IMPLANT WAS USED TO COMPLETE THE IMPLANTATION. NO ADVERSE EVENTS OR PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19455121 ↗InjuryMay 13, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT BREAST AUGMENTATION WITH 250CC MENTOR MEMORYGEL BREAST IMPLANT EXPERIENCED BILATERAL RUPTURE POST PROCEDURE. AS A RESULT, EXPLANT WAS PERFORMED ON (B)(6) 2024.
Patient problems: Deformity/ Disfigurement
MAUDE report 19306341 ↗InjuryMay 13, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT BREAST AUGMENTATION WITH 250CC MENTOR MEMORYGEL BREAST IMPLANT EXPERIENCED BILATERAL RUPTURE POST PROCEDURE. AS A RESULT, EXPLANT WAS PERFORMED ON (B)(6) 2024.
Patient problems: Deformity/ Disfigurement
MAUDE report 19306339 ↗InjuryApr 15, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 250CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT FELL AND WAS REPORTED TO HAVE EXPERIENCED BILATERALLY RUPTURED BREAST IMPLANTS. A CONSULTATION WITH A MEDICAL PROFESSIONAL HAS BEEN CONDUCTED, HOWEVER, AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THE DATE OF IMPLANTATION WAS PROVIDED TO MENTOR AS A BEST ESTIMATE. MENTOR BECAME AWARE THAT A DISCREPANCY WAS OBSERVED WHEN THE INFORMATION REPORTED INCLUDED A DATE OF IMPLANTATION THAT OCCURRED PRIOR TO THE MANUFACTURING DATE OF THE BREAST IMPLANT. AS THIS IS NOT POSSIBLE, MENTOR ASSOCIATES ARE REACHING OUT FOR FURTHER INFORMATION. FOLLOW-UPS ARE BEING CONDUCTED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY. REFER TO MANUFACTURING REPORT NUMBER 1645337-2024-04509 FOR THE CONTRALATERAL EVENT.
Patient problems: Deformity/ Disfigurement
MAUDE report 19110537 ↗InjuryApr 15, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 250CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT FELL AND WAS REPORTED TO HAVE EXPERIENCED BILATERALLY RUPTURED BREAST IMPLANTS. A CONSULTATION WITH A MEDICAL PROFESSIONAL HAS BEEN CONDUCTED, HOWEVER, AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THE DATE OF IMPLANTATION WAS PROVIDED TO MENTOR AS A BEST ESTIMATE. MENTOR BECAME AWARE THAT A DISCREPANCY WAS OBSERVED WHEN THE INFORMATION REPORTED INCLUDED A DATE OF IMPLANTATION THAT OCCURRED PRIOR TO THE MANUFACTURING DATE OF THE BREAST IMPLANT. AS THIS IS NOT POSSIBLE, MENTOR ASSOCIATES ARE REACHING OUT FOR FURTHER INFORMATION FOLLOW-UPS ARE BEING CONDUCTED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY. REFER TO MANUFACTURING REPORT NUMBER 1645337-2024-04508 FOR THE CONTRALATERAL EVENT.
Patient problems: Deformity/ Disfigurement
MAUDE report 19110520 ↗InjuryApr 10, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Material Rupture; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 300CC (LEFT), 250CC (RIGHT), MENTOR MEMORYGEL BREAST IMPLANT EXPERIENCED BILATERAL RUPTURES AND CAPSULAR CONTRACTURE BAKER GRADE 4 ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2024. THIS REPORT IS FOR THE RIGHT SIDE DEVICE.
Patient problems: Capsular Contracture; Seroma
MAUDE report 19082013 ↗InjuryJan 30, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT WHO HAD A 250CC MENTOR MEMORYGEL BREAST IMPLANT PLACED EXPERIENCED LEFT SIDED RUPTURE POST PROCEDURE. THE RUPTURE WAS DISCOVERED DURING EXPLANT. EXPLANT AND REPLACEMENT WITH A NEW 250CC MENTOR MEMORYGEL BREAST IMPLANT WAS PERFORMED ON (B)(6) 2023.
Patient problems: Deformity/ Disfigurement
MAUDE report 18607117 ↗
Other Prosthesis, breast, noninflatable, internal, silicone gel-filled devices
- MENTOR MemoryGel Breast Implant 3503501BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,167 reports
- MENTOR MemoryGel Breast Implant 3504001BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,130 reports
- MENTOR MemoryGel Breast Implant 3504004BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,012 reports
- MENTOR MemoryGel Breast Implant 3504504BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,754 reports
- MENTOR MemoryGel Breast Implant 3503001BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,752 reports
- MENTOR MemoryGel Breast Implant 3503504BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,749 reports
- MENTOR MemoryGel Breast Implant 3504501BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,690 reports
- MENTOR MemoryGel Breast Implant 3503251BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,684 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page