MENTOR MemoryGel Breast Implant 3544450Recalls & reported problems

Mentor Texas L.P. · SILTEX Round High Profile Gel-Filled Breast Implant, 450cc |Textured|Round|Silicone| · Class 3 · Implant

All models of MENTOR MemoryGel Breast Implants (368) · MENTOR MemoryGel Breast Implant (211)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Mentor MemoryGel Siltex Round silicone gel-filled breast implant appears in 74 adverse event reports, all describing injuries and no deaths. Most reports describe the implant breaking (rupture), often along with scar tissue tightening around the implant, swelling or fluid collection, and changes in breast shape; several patients underwent surgery to remove or replace the implants. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
74
Deaths reported
0
Injuries reported
74
Recalls naming it
0

Summary

The FDA's MAUDE database contains 74 adverse event reports that list this exact device (DI 00081317001102), received between Aug 21, 2018 and Apr 14, 2026 — classified as 74 injuries.

The most frequently reported specific device problem is “Material Rupture” (37 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem37
  • Material Rupture37
  • Migration3

Patient problems

  • Capsular Contracture31
  • No Code Available22
  • Deformity/ Disfigurement21
  • Granuloma8
  • Seroma5
  • Wrinkling4
  • Breast Implant Associated Anaplastic Large Cell Lymphoma (BIA ALCL)3
  • Breast Mass3

Recent adverse event reports

The 25 most recent of 74 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 14, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA MRI, WITH RIGHT BREAST IMPLANT RUPTURE. IN ADDITION, THE PATIENT ALSO DEVELOPED RIGHT BREAST CAPSULAR CONTRACTURE AND A MODERATELY THICKENED LEFT BREAST. AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025. THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 24876987 ↗
  • InjuryMar 16, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT WITH MENTOR MEMORYGEL BREAST IMPLANTS EXPERIENCED RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT DEVICE REMOVAL.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24604532 ↗
  • InjuryFeb 27, 2026Migration

    IT WAS REPORTED THAT A PATIENT WITH MENTOR MEMORYGEL BREAST IMPLANTS EXPERIENCED RUPTURE, CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN, SWELLING AND IMPLANT SITE FLUID COLLECTION ON THE LEFT SIDE ALONG WITH DEVICE MIGRATION ON THE RIGHT SIDE. AS A RESULT, THE PATIENT UNDERWENT DEVICE REMOVAL AND REPLACEMENT. THIS REPORT IS FOR THE RIGHT BREAST PROSTHESIS.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24473315 ↗
  • InjuryDec 26, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT WITH MENTOR MEMORYGEL BREAST IMPLANTS EXPERIENCED RUPTURE ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT IS SCHEDULED FOR REMOVAL ON (B)(6) 2026.

    Patient problems: Capsular Contracture; Deformity/ Disfigurement; Local Reaction; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899128 ↗
  • InjuryOct 24, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 450CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERALLY RUPTURED BREAST IMPLANTS USING MAGNETIC RESONANCE IMAGING. AS A RESULT, THE PATIENT WAS SCHEDULED FOR BILATERAL IMPLANT REMOVAL ON (B)(6) 2025.

    Patient problems: Capsular Contracture; Deformity/ Disfigurement

    MAUDE report 23383351 ↗
  • InjuryOct 24, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 450CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERALLY RUPTURED BREAST IMPLANTS USING MAGNETIC RESONANCE IMAGING. AS A RESULT, THE PATIENT WAS SCHEDULED FOR BILATERAL IMPLANT REMOVAL ON (B)(6) 2025.

    Patient problems: Capsular Contracture; Deformity/ Disfigurement

    MAUDE report 23383350 ↗
  • InjuryJun 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 44-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 450CC MENTOR MEMORYGEL BREAST IMPLANT ON BOTH SIDES AND EXPERIENCED BREAST IMPLANT RUPTURE ON HER RIGHT SIDE POSTOPERATIVELY. AS A RESULT, THE DEVICES WERE EXPLANTED ON (B)(6) 2025. ON (B)(6) 2025, MENTOR BECAME AWARE THAT THE PATIENT ALSO EXPERIENCED BILATERAL WRINKLING, AND LEFT SIDE BREAST MASS IN ADDITION TO RIGHT SIDE RUPTURE. RIGHT SIDE REPLACEMENT DEVICE USED ON (B)(6) 2025 WAS CATALOG NUMBER 3504504BC; SERIAL NUMBER (B)(6). THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT-SIDED DEVICE.

    Patient problems: Breast Mass; Wrinkling

    MAUDE report 22301236 ↗
  • InjuryMar 5, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A 35-YEAR-OLD HISPANIC FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 450CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED A RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 21532098 ↗
  • InjuryFeb 19, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 450CC MENTOR MEMORYGEL BREAST IMPLANT EXPERIENCED CAPSULAR CONTRACTURE AND A RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 21412396 ↗
  • InjuryFeb 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH MENTOR MEMORYGEL BREAST IMPLANTS AND EXPERIENCED RUPTURE ON THE LEFT SIDE ALONG WITH CAPSULAR CONTRACTURE BAKER GRADE III ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE RIGHT BREAST PROSTHESIS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Contracture

    MAUDE report 21334299 ↗
  • InjuryJan 15, 2025Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 450CC MENTOR MEMORYGEL BREAST IMPLANT EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE 2 ON THE LEFT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Capsular Contracture

    MAUDE report 21159321 ↗
  • InjuryDec 30, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT A 62-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 450CC MENTOR MEMORYGEL BREAST IMPLANT ON BOTH SIDES AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE; BAKER GRADE III, AND BILATERAL BREAST IMPLANT RUPTURE POSTOPERATIVELY; DIAGNOSED VIA MRI. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REPLACEMENT SURGERY WITH CATALOG NUMBER SMPX465; SERIAL NUMBER (B)(6) ON THE LEFT SIDE, AND WITH CATALOG NUMBER SMPX465; SERIAL NUMBER (B)(6) ON THE RIGHT SIDE ON (B)(6) 2024. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE.

    Patient problems: Capsular Contracture; Granuloma

    MAUDE report 21039260 ↗
  • InjuryDec 30, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT A 62-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 450CC MENTOR MEMORYGEL BREAST IMPLANT ON BOTH SIDES AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE; BAKER GRADE III, AND BILATERAL BREAST IMPLANT RUPTURE POSTOPERATIVELY; DIAGNOSED VIA MRI. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REPLACEMENT SURGERY WITH CATALOG NUMBER SMPX465; SERIAL NUMBER (B)(6) ON THE LEFT SIDE, AND WITH CATALOG NUMBER SMPX465; SERIAL NUMBER (B)(6) ON THE RIGHT SIDE ON (B)(6) 2024. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT-SIDED DEVICE.

    Patient problems: Capsular Contracture; Granuloma

    MAUDE report 21039241 ↗
  • InjuryNov 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH 450CC MENTOR MEMORYGEL BREAST IMPLANTS EXPERIENCED BILATERAL CAPSULAR CONTRACTURE POST PROCEDURE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Capsular Contracture

    MAUDE report 20731045 ↗
  • InjuryNov 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH 450CC MENTOR MEMORYGEL BREAST IMPLANTS EXPERIENCED BILATERAL CAPSULAR CONTRACTURE POST PROCEDURE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Capsular Contracture

    MAUDE report 20731043 ↗
  • InjuryOct 15, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 450 CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH A RUPTURED LEFT BREAST IMPLANT USING MAGNETIC RESONANCE IMAGING. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. SEVERAL FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 20452684 ↗
  • InjuryMay 7, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT A FEMALE WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH A MENTOR MEMORYGEL 450CC SILICONE PROSTHESIS EXPERIENCED LEFT SIDED SILENT RUPTURE POST PROCEDURE. THE ISSUE WAS CONFIRMED BY MAMMOGRAM AND ULTRASOUND EXAMINATIONS. AS A RESULT, PATIENT SCHEDULED FOR REMOVAL ON (B)(6) 2024.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 19261491 ↗
  • InjuryMar 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH A MENTOR MEMORYGEL, 450CC SILICONE, BREAST IMPLANT EXPERIENCED BILATERAL CAPSULAR CONTRACTURE GRADE IV ON THE RIGHT, AND GRADE III ON THE LEFT SIDE POSTOPERATIVE. THE HEALTH CARE PROFESSIONAL CONFIRMED THE ISSUE. AS A RESULT, PATIENT UNDERWENT BILATERAL REMOVAL ON (B)(6) 2024. THIS REPORT RELATES TO THE LEFT SIDE.

    Patient problems: Capsular Contracture

    MAUDE report 18895663 ↗
  • InjuryMar 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH A MENTOR MEMORYGEL, 450CC SILICONE, BREAST IMPLANT EXPERIENCED BILATERAL CAPSULAR CONTRACTURE GRADE IV ON THE RIGHT, AND GRADE III ON THE LEFT SIDE POSTOPERATIVE. THE HEALTH CARE PROFESSIONAL CONFIRMED THE ISSUE. AS A RESULT, PATIENT UNDERWENT BILATERAL REMOVAL ON (B)(6) 2024. THIS REPORT RELATES TO THE RIGHT SIDE.

    Patient problems: Capsular Contracture

    MAUDE report 18895587 ↗
  • InjuryFeb 13, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION PRIMARY WITH A MENTOR MEMORYGEL, 450CC SILICONE, BREAST IMPLANT EXPERIENCED BILATERAL SILENT RUPTURE POSTOPERATIVE. AS A RESULT, PATIENT UNDERWENT BILATERAL REMOVAL ON FEBRUARY 13, 2024. THIS REPORT RELATES TO THE RIGHT SIDE.

    Patient problems: Capsular Contracture; Deformity/ Disfigurement

    MAUDE report 18695987 ↗
  • InjuryFeb 13, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION PRIMARY WITH A MENTOR MEMORYGEL, 450CC SILICONE, BREAST IMPLANT EXPERIENCED BILATERAL SILENT RUPTURE POSTOPERATIVE. AS A RESULT, PATIENT UNDERWENT BILATERAL REMOVAL ON (B)(6) 2024. THIS REPORT RELATES TO THE LEFT SIDE.

    Patient problems: Capsular Contracture; Deformity/ Disfigurement

    MAUDE report 18695935 ↗
  • InjurySep 20, 2023Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH A (LEFT) 500CC MENTOR MEMORYGEL BREAST IMPLANT AND A (RIGHT) 450CC MENTOR MEMORYGEL BREAST IMPLANT. POST-OPERATIVELY, THE PATIENT SUFFERED BILATERAL BREAST FOLDING ON THE SIDES AND BOTTOM. THE BREAST FELT LIKE THEY ARE SHIFTING TO THE OUTSIDE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 17784887 ↗
  • InjurySep 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR MEMORYGEL BREAST IMPLANT 450CC GEL BREAST PROSTHESIS (LEFT DEVICE) AND A MENTOR MEMORYGEL BREAST IMPLANT 500CC GEL BREAST PROSTHESIS (RIGHT DEVICE) DEVELOPED BAKER GRADE III/IV CAPSULAR CONTRACTURE IN BOTH OF HER BREASTS POST IMPLANTATION. BILATERAL CAPSULAR CONTRACTURE WAS DIAGNOSED VIA A PHYSICAL EXAM. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 17777638 ↗
  • InjurySep 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH MENTOR MEMORYGEL BREAST IMPLANTS 450CC AND EXPERIENCED BILATERAL BREAST PAIN POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE LEFT BREAST PROSTHESIS. SEE MANUFACTURER REPORT NUMBER 1645337-2023-11034 FOR CONTRALATERAL SIDE.

    Patient problems: Breast Discomfort/Pain

    MAUDE report 17769648 ↗
  • InjurySep 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH MENTOR MEMORYGEL BREAST IMPLANTS 450CC AND EXPERIENCED BILATERAL BREAST PAIN POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE RIGHT BREAST PROSTHESIS. SEE MANUFACTURER REPORT NUMBER 1645337-2023-11033 FOR CONTRALATERAL SIDE.

    Patient problems: Breast Discomfort/Pain

    MAUDE report 17769646 ↗

Other Prosthesis, breast, noninflatable, internal, silicone gel-filled devices

All Prosthesis, breast, noninflatable, internal, silicone gel-filled devices →