MENTOR Smooth Round Moderate Plus Profile 3502300Recalls & reported problems

Mentor Texas L.P. · Saline Breast Implant with Diaphragm Valve, 300cc |Smooth|Round|Saline| · Class 3 · Implant

All models of MENTOR Smooth Round Moderate Plus Profile (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Mentor Smooth Round saline breast implant is used for breast augmentation and revision. Of its 565 reports, 556 are injuries, most describing the implant shell rupturing and leaving the implant deflated. Patients mention deformity in 222 reports and scar tightening (capsular contracture) in 104. There were no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
566
Deaths reported
0
Injuries reported
557
Recalls naming it
0

Summary

The FDA's MAUDE database contains 566 adverse event reports that list this exact device (DI 00081317001393), received between Jan 26, 2018 and Sep 24, 2026 — classified as 557 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (391 reports).

2 other recalls cover devices cleared under the same FDA submission (P990075); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 7 reportsJan 2024: 5 reportsFeb 2024: 4 reportsMar 2024: 3 reportsApr 2024: 1 reportsMay 2024: 4 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 5 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 2 reportsMar 2025: 7 reportsApr 2025: 3 reportsMay 2025: 3 reportsJun 2025: 4 reportsJul 2025: 8 reportsAug 2025: 6 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 3 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 4 reportsMar 2026: 2 reportsApr 2026: 3 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 4 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture391
  • Adverse Event Without Identified Device or Use Problem167
  • Migration16
  • Appropriate Device Problem Term/Code Not Available3
  • Deflation Problem3
  • No Apparent Adverse Event3
  • Off-Label Use3
  • Connection Problem2

Patient problems

  • Deformity/ Disfigurement224
  • No Code Available191
  • Capsular Contracture103
  • Pain27
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available16
  • Breast Discomfort/Pain16
  • Wrinkling14
  • Autoimmune Disorder10

Recent adverse event reports

The 25 most recent of 566 reports, in the reporter's own words as published by the FDA.

  • InjurySep 24, 2026Migration

    Patient problems: Deformity/ Disfigurement

    MAUDE report 50888975 ↗
  • InjurySep 24, 2026Migration

    It was reported that a female patient underwent a breast augmentation with two 300cc MENTOR Smooth Round Moderate Plus Profile saline breast prostheses and experienced bilateral device migration (superior positioning) post-operatively. At the time of this report, Mentor has received no information regarding explantation or an expected explantation date.This report is for the left side device.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 50888974 ↗
  • InjuryAug 27, 2026Material Rupture

    It was reported that a patient underwent primary breast augmentation with two 300cc MENTOR Smooth Round Moderate Plus Profile saline breast prostheses. Post-operatively, the patient suffered bilateral breast implant deflation. As a result, the patient has been scheduled for breast implant removal surgery on (b)(6) 2026.This Medwatch form is for the left breast prosthesis.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 50468011 ↗
  • InjuryAug 27, 2026Material Rupture

    It was reported that a patient underwent primary breast augmentation with two 300cc MENTOR Smooth Round Moderate Plus Profile saline breast prostheses. Post-operatively, the patient suffered bilateral breast implant deflation. As a result, the patient has been scheduled for breast implant removal surgery on (b)(6) 2026.This Medwatch form is for the right breast prosthesis.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 50468004 ↗
  • InjuryAug 25, 2026Material Rupture

    It was reported that a patient underwent a Primary Breast Augmentation Surgery with implantation of an undisclosed Mentor saline implant prosthesis. Post-operatively, the patient was diagnosed with a deflated right implant during a physical exam. As a result, the patient underwent implant removal on (b)(6)2026.The date of implantation was provided to Mentor as a best estimate. Follow-ups have been conducted, and no additional information has been received. If more information becomes available, Mentor will submit a supplemental report accordingly.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 50431235 ↗
  • InjuryAug 6, 2026Material Rupture

    Patient problems: Deformity/ Disfigurement

    MAUDE report 50165456 ↗
  • InjuryMay 4, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH A MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESIS THAT DEFLATED POST IMPLANTATION. DEFLATION OF THE PATIENT¿S RIGHT BREAST PROSTHESIS WAS REPORTED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL BREAST IMPLANT GEL BREAST PROSTHESES.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 25076220 ↗
  • InjuryMay 4, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT WITH SALINE MENTOR SMOOTH ROUND MODERATE PLUS PROFILE BREAST IMPLANTS EXPERIENCED BILATERAL DEFLATION POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE LEFT BREAST PROSTHESIS.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 25068151 ↗
  • InjuryMay 4, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT WITH SALINE MENTOR SMOOTH ROUND MODERATE PLUS PROFILE BREAST IMPLANTS EXPERIENCED BILATERAL DEFLATION POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE RIGHT BREAST PROSTHESIS.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 25068150 ↗
  • InjuryMay 1, 2026Migration

    IT WAS REPORTED THAT A 56-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED BREAST PAIN AND DEVICE MIGRATION ON THE LEFT SIDE POST-OPERATIVELY. THE PATIENT MENTIONED LATERAL DISPLACEMENT AND WHEN LAYING DOWN, THE BREAST PROSTHESIS MOVES UNDER THE ARMPIT. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Breast Discomfort/Pain

    MAUDE report 25057487 ↗
  • InjuryApr 16, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT REPORTED EXPERIENCING A DEFLATED RIGHT IMPLANT. A CONSULTATION WITH A MEDICAL PROFESSIONAL HAS YET TO BE CONDUCTED. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24901734 ↗
  • InjuryApr 2, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST IMPLANTS PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERALLY DEFLATED IMPLANTS DURING AN IN-OFFICE VISIT. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24767148 ↗
  • InjuryApr 2, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST IMPLANTS PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERALLY DEFLATED IMPLANTS DURING AN IN-OFFICE VISIT. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24767147 ↗
  • InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESES POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE UNSPECIFIED). AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Capsular Contracture

    MAUDE report 24647362 ↗
  • InjuryMar 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 43-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED CAPSULAR CONTRACTURE (BAKER GRADE 3) ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED WITH A PHYSICAL EXAM. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Capsular Contracture

    MAUDE report 24584161 ↗
  • InjuryFeb 20, 2026Material Rupture

    IT WAS REPORTED THAT A CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED A DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED WITH A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2026.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24405473 ↗
  • InjuryFeb 12, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH RIGHT BREAST IMPLANT DEFLATION. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2026. THE REPLACEMENT DEVICES WERE TWO MENTOR GEL IMPLANTS.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24324319 ↗
  • InjuryFeb 10, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH A DEFLATED RIGHT IMPLANT DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXCHANGE WITH MENTOR SALINE IMPLANTS ON (B)(6) , 2026.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24308714 ↗
  • InjuryFeb 5, 2026Material Rupture; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT WAS IN A SKIING ACCIDENT AND DEVELOPED LEFT BREAST IMPLANT MOVEMENT AND LEAK. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24274253 ↗
  • InjuryJan 23, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION WITH TWO 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT SUFFERED RIGHT BREAST PAIN AND POSSIBLE RIGHT BREAST IMPLANT LEAK. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2026.

    Patient problems: Breast Discomfort/Pain

    MAUDE report 24158620 ↗
  • InjuryJan 8, 2026Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH A BAKER GRADE 1 CAPSULAR CONTRACTURE AND RECEIVED A RECALL NOTICE. AS A RESULT, THE PATIENT UNDERWENT BILATERAL IMPLANT REMOVAL WITHOUT REPLACEMENTS ON (B)(6) 2025. HOWEVER, DURING THE SURGERY, A DEFLATED LEFT IMPLANT WAS DISCOVERED.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24013007 ↗
  • InjuryDec 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST RECONSTRUCTION REVISION SURGERY WITH IMPLANTATION OF A 300CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS REPORTED TO HAVE EXPERIENCED UNSPECIFIED HEALTH ISSUES AND PAIN. AS A RESULT, THE PATIENT UNDERWENT BILATERAL IMPLANT REMOVAL WITHOUT REPLACEMENTS ON (B)(6) 2025. FURTHERMORE, DURING THE SURGERY, THE EXPLANTED LEFT IMPLANT WAS OBSERVED TO BE DEFLATED. THIS REPORT IS FOR THE RIGHT BREAST PROSTHESIS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23763845 ↗
  • InjuryDec 9, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT WITH SALINE MENTOR SMOOTH ROUND MODERATE PLUS PROFILE BREAST IMPLANTS EXPERIENCED DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT IS SCHEDULED FOR REMOVAL ON (B)(6) 2026.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 23759161 ↗
  • InjuryNov 24, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR SMOOTH ROUND MODERATE PLUS PROFILE 300CC SALINE BREAST PROSTHESIS THAT DEFLATED POST IMPLANTATION. DEFLATION OF THE PATIENT¿S LEFT BREAST PROSTHESIS WAS REPORTED. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 23638115 ↗
  • InjuryNov 14, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 300CC MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED A DEFLATION ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 23559677 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990075. One submission often covers many models, so check the product and lot details in each record.

  • Z-0367-2022 · Sep 30, 2021 · Open, Classified
    MENTOR Saline-Filled Breast Implants are constructed from room temperature vulcanized silicone elastomer, made of polydimethylsiloxane. Surgeons inflate the silicone elastomer shel
  • Z-2070-2017 · Mar 27, 2017 · Terminated
    Mentor Smooth Round SPECTRUM Post- Operatively Adjustable Saline Breast Implant, Product Code: 350-1490, UDI/GTIN: 00081317002048

Other Prosthesis, breast, inflatable, internal, saline devices

All Prosthesis, breast, inflatable, internal, saline devices →