MENTOR Smooth Round High Profile 3503630Recalls & reported problems
Mentor Texas L.P. · Saline Breast Implant with Diaphragm Valve, 630cc |Smooth|Round|Saline| · Class 3 · Implant
All models of MENTOR Smooth Round High Profile (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Mentor Smooth Round High Profile is a saline breast implant. Its 217 reports mostly describe implants rupturing and deflating, often followed by removal, and many describe the breast becoming hard from scar tissue tightening. The reports include 212 injuries and no deaths. Reports fell from 16 in the prior 12 months to 9 in the latest 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 218
- Deaths reported
- 0
- Injuries reported
- 213
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 218 adverse event reports that list this exact device (DI 00081317001706), received between Jan 25, 2018 and Sep 18, 2026 — classified as 213 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (138 reports).
2 other recalls cover devices cleared under the same FDA submission (P990075); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture138
- Adverse Event Without Identified Device or Use Problem77
- Migration3
- Air/Gas in Device2
- Connection Problem2
- Break1
- Deflation Problem1
- Migration or Expulsion of Device1
Patient problems
- Deformity/ Disfigurement81
- No Code Available69
- Capsular Contracture50
- Pain11
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available10
- Breast Discomfort/Pain7
- Autoimmune Disorder6
- Wrinkling6
Recent adverse event reports
The 25 most recent of 218 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2026Material Rupture
It was reported that a female patient underwent unspecified Breast Augmentation with MENTOR Smooth Round High Profile, 630cc Saline implants, subsequently the patient experienced left sided deflation. As a result, the patient scheduled for removal surgery on (b)(6), 2026.
Patient problems: Deformity/ Disfigurement
MAUDE report 50818990 ↗InjuryApr 29, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS EXPERIENCED PTOSIS ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2026.
Patient problems: Deformity/ Disfigurement
MAUDE report 25032278 ↗InjuryMar 31, 2026Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT SUFFERED RIGHT BREAST IMPLANT DEFLATION. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Deformity/ Disfigurement
MAUDE report 24738764 ↗InjuryFeb 19, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 72-YEAR-OLD HISPANIC FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH TWO 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESES AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE (BAKER GRADE UNKNOWN) POST-OPERATIVELY. THE PATIENT DESCRIBED THE BREAST TO BE "REALLY HARD", PAINFUL, AND ONE SIDE BEING SMALLER THAN THE OTHER. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE LEFT SIDE DEVICE.
Patient problems: Capsular Contracture
MAUDE report 24394947 ↗InjuryFeb 19, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 72-YEAR-OLD HISPANIC FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH TWO 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESES AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE (BAKER GRADE UNKNOWN) POST-OPERATIVELY. THE PATIENT DESCRIBED THE BREAST TO BE "REALLY HARD", PAINFUL, AND ONE SIDE BEING SMALLER THAN THE OTHER. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE RIGHT-SIDE DEVICE.
Patient problems: Capsular Contracture
MAUDE report 24394945 ↗InjuryFeb 18, 2026Material Rupture
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS EXPERIENCED A DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2026.
Patient problems: Deformity/ Disfigurement
MAUDE report 24379033 ↗InjuryFeb 4, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST PROSTHESIS IMPLANTATION SURGERY WITH TWO 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT SUFFERED LEFT BREAST IMPLANT LEAK. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025. THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS.
Patient problems: Capsular Contracture
MAUDE report 24264397 ↗InjuryFeb 3, 2026Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH A MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS THAT DEFLATED POST IMPLANTATION. DEFLATION OF THE PATIENT¿S LEFT BREAST PROSTHESIS WAS DIAGNOSED VIA A HEALTHCARE PROFESSIONAL¿S EXAM. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL XTRA BREAST IMPLANT GEL BREAST PROSTHESES.
Patient problems: Deformity/ Disfigurement
MAUDE report 24251337 ↗InjuryDec 17, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT WITH SALINE MENTOR SMOOTH ROUND HIGH PROFILE BREAST IMPLANTS EXPERIENCED DEFLATION ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT DEVICE REMOVAL.
Patient problems: Deformity/ Disfigurement
MAUDE report 23830726 ↗InjuryDec 1, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH RIGHT BREAST IMPLANT DEFLATION. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 23687277 ↗InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST RECONSTRUCTION REVISION SURGERY WITH IMPLANTATION OF A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH REPORTED HAVING BREAST CANCER OF AN UNDISCLOSED SIDE. NO FURTHER DETAILS WERE PROVIDED. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. AS THE AFFECTED SIDE WAS NOT PROVIDED, THE LOT/SERIAL NUMBERS FOR BOTH PRODUCTS ARE BEING PROVIDED AS LEFT IMPLANT - LOT NUMBER: 9827568 / SERIAL NUMBER: (B)(6) AND RIGHT IMPLANT - LOT NUMBER: 9813544 / SERIAL NUMBER: (B)(6). THE CATALOG NUMBER IS THE SAME FOR BOTH DEVICES. FOLLOW-UPS WERE NOT PERFORMED PER THE PATIENT'S REQUEST.
Patient problems: Breast Cancer
MAUDE report 23319307 ↗InjurySep 12, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A 53-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED A DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED WITH PICTURE. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 23039815 ↗InjurySep 9, 2025Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WITH A SALINE MENTOR SMOOTH ROUND HIGH PROFILE BREAST IMPLANT (L) AND AN UNSPECIFIED MENTOR SALINE BREAST IMPLANT (R) EXPERIENCED BILATERAL DEFLATION ALONG WITH CAPSULAR CONTRACTURE BAKER GRADE IV ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE LEFT BREAST PROSTHESIS.
Patient problems: Capsular Contracture; Deformity/ Disfigurement
MAUDE report 23009296 ↗InjuryAug 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH MENTOR SMOOTH ROUND HIGH PROFILE, 630CC SILICONE PROSTHESIS EXPERIENCE LEFT SIDED CAPSULAR CONTRACTURE UNKNOWN GRADE POST PROCEDURE. THE ISSUE WAS DIAGNOSED THROUGH PHYSICAL EXAMINATION. AS A RESULT, THE PATIENT SCHEDULED FOR REMOVAL SURGERY ON (B)(6) 2025.
Patient problems: Capsular Contracture
MAUDE report 22752779 ↗InjuryAug 1, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A 55-YEAR-OLD FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED A DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Deformity/ Disfigurement
MAUDE report 22687142 ↗InjuryJul 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR SMOOTH ROUND HIGH PROFILE 630CC SALINE BREAST PROSTHESIS DEVELOPED ¿WEIRD¿ FEELINGS IN HER RIGHT BREAST POST IMPLANTATION. ADDITIONALLY, THE PATIENT¿S RIGHT BREAST FEELS HEAVIER, AND THE BREAST APPEARS LARGER THAN THE OTHER. THE PATIENT SUFFERED FROM POSSIBLE TRAUMA FROM A CAR ACCIDENT. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Breast Discomfort/Pain; Deformity/ Disfigurement
MAUDE report 22522475 ↗InjuryJul 10, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST RECONSTRUCTION SURGERY WITH IMPLANTATION OF A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH A DEFLATED RIGHT BREAST IMPLANT DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 22468417 ↗InjuryJun 2, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS EXPERIENCED A DEFLATION ON THE LEFT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Deformity/ Disfigurement
MAUDE report 22125286 ↗MalfunctionMay 27, 2025Break; Break
IT WAS REPORTED THAT A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BEING USED IN A BREAST AUGMENTATION REVISION WAS FOUND TO BE LEAKING AT THE FILL TUBE DURING THE OPERATION. NO ADVERSE EVENT WAS EXPERIENCED BY THE PATIENT. THE SURGEON USED A BACKUP DEVICE, AND THERE WERE NO REPORTED PATIENT CONSEQUENCES OR PROCEDURAL DELAYS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22091743 ↗InjuryMay 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT BREAST AUGMENTATION REVISION WITH MENTOR SMOOTH ROUND HIGH PROFILE 630CC SALINE PROSTHESES EXPERIENCED RIGHT SIDED CAPSULAR CONTRACTURE, RIGHT SIDED LOCATION, AND A LEFT SIDED ANISOMASTIA POST PROCEDURE. THIS WAS ACCOMPANIED BY BILATERAL PAIN. THERE WAS SWELLING ON THE RIGHT SIDE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Capsular Contracture; Local Reaction
MAUDE report 22014275 ↗InjuryMay 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT BREAST AUGMENTATION REVISION WITH MENTOR SMOOTH ROUND HIGH PROFILE 630CC SALINE PROSTHESES EXPERIENCED RIGHT SIDED CAPSULAR CONTRACTURE, RIGHT SIDED LOCATION, AND A LEFT SIDED ANISOMASTIA POST PROCEDURE. THIS WAS ACCOMPANIED BY BILATERAL PAIN. THERE WAS SWELLING ON THE RIGHT SIDE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Breast Discomfort/Pain; Deformity/ Disfigurement
MAUDE report 22014263 ↗InjuryMar 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE 2 ON THE LEFT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Capsular Contracture
MAUDE report 21708997 ↗InjuryMar 10, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESIS EXPERIENCED A DEFLATION ON THE LEFT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Deformity/ Disfigurement
MAUDE report 21561224 ↗InjuryMar 4, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 630CC MENTOR SMOOTH ROUND HIGH PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA MAMMOGRAM, WITH LEFT BREAST IMPLANT DEFLATION. AS A RESULT, THE PATIENT HAS BEEN SCHEDULED FOR BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 21517400 ↗InjuryJan 23, 2025Migration; Migration
A FEMALE PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SMOOTH ROUND HIGH PROFILE 630CC SALINE PROSTHESIS EXPERIENCED LEFT-SIDED DEFLATION POSTOPERATIVELY, DIAGNOSED DURING AN IN-OFFICE EXAMINATION. SUBSEQUENTLY, SHE WAS FOUND TO HAVE BILATERAL MALPOSITION, NECESSITATING SURGICAL INTERVENTION. ON (B)(6) 2024, THE PATIENT UNDERWENT CAPSULORRAPHY ON THE LEFT SIDE AND CAPSULOTOMY ON THE RIGHT, FOLLOWED BY BILATERAL REPLACEMENTS USING CAT 3503630, LOT NUMBERS 9958401 ON THE RIGHT AND 9730811 FOR THE LEFT SIDE. THIS REPORT RELATES TO THE RIGHT SIDE.
Patient problems: Deformity/ Disfigurement
MAUDE report 21217533 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990075. One submission often covers many models, so check the product and lot details in each record.
- Z-0367-2022 · Sep 30, 2021 · Open, ClassifiedMENTOR Saline-Filled Breast Implants are constructed from room temperature vulcanized silicone elastomer, made of polydimethylsiloxane. Surgeons inflate the silicone elastomer shel
- Z-2070-2017 · Mar 27, 2017 · TerminatedMentor Smooth Round SPECTRUM Post- Operatively Adjustable Saline Breast Implant, Product Code: 350-1490, UDI/GTIN: 00081317002048
Other Prosthesis, breast, inflatable, internal, saline devices
- MENTOR Smooth Round Moderate Profile 3501650Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline2,187 reports
- MENTOR Smooth Round Moderate Profile 3501660Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline1,972 reports
- MENTOR Smooth Round Moderate Profile 3501670Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline1,899 reports
- MENTOR Smooth Round Moderate Profile 3501645Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline1,655 reports
- MENTOR Smooth Round Moderate Profile 3501655Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline1,534 reports
- MENTOR Smooth Round Moderate Profile 3501640Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline1,294 reports
- MENTOR Smooth Round Moderate Profile 3501680Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline1,095 reports
- MENTOR Smooth Round Moderate Profile 3501635Mentor Texas L.P. · Prosthesis, breast, inflatable, internal, saline864 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page