MENTOR MemoryGel Xtra Breast Implant SHPX755Recalls & reported problems
Mentor Texas L.P. · Smooth High Profile Xtra 755cc |Smooth|Round|Silicone| · Class 3 · Implant
All models of MENTOR MemoryGel Breast Implants (368) · MENTOR MemoryGel Xtra Breast Implant (88)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is a Mentor MemoryGel Xtra 755cc silicone gel breast implant. The 63 reports mostly describe the scar tissue tightening around the implant (capsular contracture), which makes the breast hard, raised and painful, and implants rupturing or leaking, sometimes leading to removal surgery. 60 injuries and no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 63
- Deaths reported
- 0
- Injuries reported
- 60
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 00081317025368), received between Jun 5, 2019 and Apr 1, 2026 — classified as 60 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (22 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem43
- Material Rupture22
- Appropriate Device Problem Term/Code Not Available2
- Air/Gas in Device1
- Device Contamination with Chemical or Other Material1
- Migration1
Patient problems
- Capsular Contracture33
- Deformity/ Disfigurement16
- No Code Available7
- No Clinical Signs, Symptoms or Conditions3
- Post Operative Wound Infection3
- Skin Inflammation/ Irritation3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Breast Discomfort/Pain2
Recent adverse event reports
The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.
InjuryApr 1, 2026Material Rupture
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT EXPERIENCED A RUPTURE ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2026.
Patient problems: Deformity/ Disfigurement
MAUDE report 24757779 ↗InjuryFeb 13, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WITH MENTOR MEMORYGEL XTRA BREAST IMPLANTS EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE III ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT DEVICE REMOVAL.
Patient problems: Capsular Contracture
MAUDE report 24335282 ↗InjuryDec 9, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT WITH MENTOR MEMORYGEL XTRA BREAST IMPLANTS EXPERIENCED BREAST PAIN AND RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Breast Discomfort/Pain
MAUDE report 23761557 ↗InjurySep 19, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST SURGERY WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED A RUPTURE AND CAPSULAR CONTRACTURE (BAKER GRADE UNKNOWN) ON THE LEFT SIDE POST-OPERATIVELY. THE PATIENT REPORTED THAT THE BREAST WAS HARD AND WAS PINCHING A NERVE CAUSING MAJOR PAIN. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Capsular Contracture
MAUDE report 23098251 ↗InjuryAug 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE 4 ON THE RIGHT SIDE POST-OPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Capsular Contracture
MAUDE report 22830968 ↗InjuryJul 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 35-YEAR-OLD RACE FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED A SKIN REACTION ON THE LEFT SIDE POST-OPERATIVELY. THE PATIENT HAD SOME REDNESS NEAR HER INCISION SITE IN THE INFERIOR PORTION OF THE BREAST. THE PATIENT ALSO HAD MINOR DISCOLORATION IN THE INFEROLATERAL ASPECT OF HER BREAST. THIS COULD HAVE BEEN DUE TO SLIGHT BRUISING OR POSSIBLE REACTION TO THE PATIENT'S BREAST MESH. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Skin Inflammation/ Irritation
MAUDE report 22613016 ↗InjuryMar 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 35-YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED CAPSULAR CONTRACTURE (BAKER GRADE 3) ON THE RIGHT SIDE POST-OPERATIVELY. THE BREAST WAS GETTING MORE FIRM AND HIGHER. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025, AND REPLACED WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT (CATALOG NUMBER SHPX755) WITH SERIAL NUMBER (B)(6). IT WAS REPORTED THAT THE PATIENT HAD A PRIOR SURGICAL INTERVENTION ON THE RIGHT SIDE. THE PATIENT HAD A CAPSULORRHAPHY WITH A MESH. THE PATIENT WENT FROM BEING TOO LOOSE TO TOO TIGHT MORE THAN EXPECTED. THE PATIENT HAD A PALPABLE NODULE OF THE MESH WHICH WAS BUGGING HER.
Patient problems: Capsular Contracture; Skin Inflammation/ Irritation
MAUDE report 21669566 ↗InjuryFeb 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 51-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 700CC MENTOR MEMORYGEL XTRA BREAST IMPLANT ON THE LEFT SIDE, AND WITH 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT ON THE RIGHT SIDE AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE; BAKER GRADE UNKNOWN POSTOPERATIVELY. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE.
Patient problems: Capsular Contracture
MAUDE report 21474622 ↗InjuryDec 10, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT ON THE LEFT BREAST. POST-OPERATIVELY, THE PATIENT SUFFERED LEFT BREAST CAPSULAR CONTRACTURE (BAKER GRADE IV). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024. DURING THE SURGERY, THE LEFT BREAST IMPLANT WAS ALSO IDENTIFIED TO HAVE RUPTURED. SUBSEQUENTLY, THE IMPLANT WAS REPLACED WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT (CATALOG: SHPX755, LOT: 9642863, SN: (B)(6).
Patient problems: Capsular Contracture
MAUDE report 20897893 ↗InjuryJun 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH A MENTOR MEMORYGEL XTRA BREAST IMPLANT 755CC GEL BREAST PROSTHESIS DEVELOPED BAKER GRADE III/IV CAPSULAR CONTRACTURE IN HER RIGHT BREAST POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR SMOOTH ROUND HIGH PROFILE 630CC SALINE BREAST PROSTHESES ON (B)(6) 2024.
Patient problems: Capsular Contracture
MAUDE report 19534303 ↗InjuryMay 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE III) AND A DROPPED LEFT BREAST. THE RIGHT BREAST HAS FIRMNESS AND LEFT BREAST IS MORE PTOTIC THAN THE RIGHT. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.
Patient problems: Deformity/ Disfigurement
MAUDE report 19233015 ↗InjuryMay 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE III) AND A DROPPED LEFT BREAST. THE RIGHT BREAST HAS FIRMNESS AND LEFT BREAST IS MORE PTOTIC THAN THE RIGHT. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS.
Patient problems: Capsular Contracture
MAUDE report 19232990 ↗InjuryApr 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT PRIMARY BREAST RECONSTRUCTION WITH A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT EXPERIENCED LEFT SIDED BAKER GRADE IV CAPSULAR CONTRACTURE POST PROCEDURE. AS A RESULT, EXPLANT WAS PERFORMED ON (B)(6) 2024.
Patient problems: Capsular Contracture
MAUDE report 19198340 ↗InjuryApr 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 54-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT ON BOTH SIDES AND EXPERIENCED BREAST IMPLANT RUPTURE, AND CAPSULAR CONTRACTURE; BAKER GRADE UNKNOWN ON HER RIGHT SIDE. ON MARCH 26, 2024, MENTOR BECAME AWARE THAT THE RIGHT SIDE DEVICE IS STILL IMPLANTED, CLARIFICATION WAS RECEIVED THAT BILATERAL MASTOPEXY WAS PERFORMED WITH SERIAL NUMBER (B)(6) ON THE LEFT SIDE, AND NUMBER (B)(6) ON THE RIGHT SIDE. WHILE RIGHT DEVICE WAS STILL IMPLANTED, LEFT SIDE IMPLANT WAS REPLACED WITH A MENTOR SMX-605 SILICONE ON (B)(6) 2022. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT-SIDED DEVICE.
Patient problems: Deformity/ Disfigurement
MAUDE report 19093073 ↗InjuryMar 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION REVISION SURGERY WITH 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT EXPERIENCED SWELLING AND SORENESS IN THE RIGHT BREAST AS WELL AS CAPSULAR CONTRACTURE OF AN UNSPECIFIED BAKER GRADE ON THE RIGHT SIDE. THE PATIENT STATED THAT HER RIGHT NIPPLE WAS SINKING INTO HER AREOLA. IT WAS ALSO REPORTED THAT THE PATIENT¿S LEFT BREAST HAD CRINKLES AND WAS DROOPING. AT THE TIME OF THIS REPORT, THE PATIENT HAS CONSULTED WITH A MEDICAL PROFESSIONAL, BUT MENTOR HAS NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE RIGHT IMPLANT. REFER TO MANUFACTURING REPORT NUMBER 1645337-2024-03114 FOR THE CONTRALATERAL EVENT.
Patient problems: Capsular Contracture; Local Reaction
MAUDE report 18897191 ↗InjuryMar 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION REVISION SURGERY WITH 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT EXPERIENCED SWELLING AND SORENESS IN THE RIGHT BREAST AS WELL AS CAPSULAR CONTRACTURE OF AN UNSPECIFIED BAKER GRADE ON THE RIGHT SIDE. THE PATIENT STATED THAT HER RIGHT NIPPLE WAS SINKING INTO HER AREOLA. IT WAS ALSO REPORTED THAT THE PATIENT¿S LEFT BREAST HAD CRINKLES AND WAS DROOPING. AT THE TIME OF THIS REPORT, THE PATIENT HAS CONSULTED WITH A MEDICAL PROFESSIONAL, BUT MENTOR HAS NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THIS REPORT IS FOR THE LEFT IMPLANT. REFER TO MANUFACTURING REPORT NUMBER 1645337-2024-03112 FOR THE CONTRALATERAL EVENT.
Patient problems: Wrinkling
MAUDE report 18897190 ↗InjuryMar 10, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 46-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST RECONSTRUCTION SURGERY WITH 7555CC MENTOR MEMORYGEL XTRA BREAST IMPLANT ON BOTH SIDES AND EXPERIENCED BREAST IMPLANT RUPTURE ON HER LEFT SIDE POSTOPERATIVELY; DIAGNOSED VIA MRI. ON (B)(6) 2024, MENTOR BECAME AWARE THAT THE DATE OF SURGERY IS (B)(6) 2024. ON FEBRUARY 21, 2024, MENTOR BECAME AWARE THAT THE PATIENT EXPERIENCED BILATERAL BREAST IMPLANT RUPTURES. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE.
Patient problems: Capsular Contracture; Deformity/ Disfigurement
MAUDE report 18872135 ↗InjuryFeb 23, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Material Rupture; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A 54-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED BREAST IMPLANT RUPTURE, AND CAPSULAR CONTRACTURE; BAKER GRADE UNKNOWN ON HER RIGHT SIDE POSTOPERATIVELY. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Capsular Contracture; Deformity/ Disfigurement; No Clinical Signs, Symptoms or Conditions
MAUDE report 18767525 ↗InjuryFeb 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT SUFFERED RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE III). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024. THE REPLACEMENT DEVICE WAS A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT (CATALOG: SHPX755, LOT: 9595719 & SN: (B)(6)).
Patient problems: Capsular Contracture
MAUDE report 18723149 ↗InjuryFeb 6, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A 46-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT REVISION BREAST RECONSTRUCTION SURGERY WITH 7555CC MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED BREAST IMPLANT RUPTURE ON HER LEFT SIDE POSTOPERATIVELY; DIAGNOSED VIA MRI. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Deformity/ Disfigurement
MAUDE report 18651385 ↗InjuryJan 29, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST SURGERY WITH IMPLANTATION OF A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT REPORTED HER IMPLANTS WERE COLLAPSING. NO FURTHER DETAILS WERE PROVIDED. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THE DATE OF EVENT WAS NOT PROVIDED TO MENTOR. SEVERAL FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY. REFER TO MANUFACTURING REPORT NUMBER 1645337-2024-01019 FOR THE CONTRALATERAL EVENT.
Patient problems: Deformity/ Disfigurement
MAUDE report 18596577 ↗InjuryJan 29, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST SURGERY WITH IMPLANTATION OF A 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT REPORTED HER IMPLANTS WERE COLLAPSING. NO FURTHER DETAILS WERE PROVIDED. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. THE DATE OF EVENT WAS NOT PROVIDED TO MENTOR. SEVERAL FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY. REFER TO MANUFACTURING REPORT NUMBER 1645337-2024-01018 FOR THE CONTRALATERAL EVENT.
Patient problems: Deformity/ Disfigurement
MAUDE report 18596576 ↗InjuryDec 20, 2023Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A FEMALE WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH A MENTOR MEMORYGEL, XTRA 755CC SILICONE PROSTHESIS EXPERIENCED LEFT-SIDED CAPSULAR CONTRACTURE GRADE IV, UNACCEPTABLE COSMETIC APPEARANCE OF THE BILATERAL BREASTS, RIGHT SIDED INFRA-AREOLAR SCAR CONTRACTURE, AND CHRONIC LEFT BREAST INFECTION POSTOPERATIVELY. THE ISSUE OF CAPSULAR CONTRACTURE WAS DIAGNOSED THROUGH PHYSICAL EXAMINATION. AS A RESULT, PATIENT UNDERWENT PERI AREOLAR MASTOPEXY, RELEASE OF RIGHT BREAST INFRA-AREOLAR MASTOPEXY, LEFT TOTAL CAPSULECTOMY WITH WASHOUT OF THE LEFT BREAST, AND BILATERAL REPLACEMENTS AS FOLLOW; L) REF: (B)(4); LOT-SN: (B)(6) AND R) REF: (B)(4); LOT-SN: (B)(6) ON (B)(6) 2023. THIS REPORT RELATES TO THE RIGHT SIDE.
Patient problems: Deformity/ Disfigurement
MAUDE report 18365461 ↗InjuryNov 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH A MENTOR MEMORYGEL XTRA BREAST IMPLANT 755CC GEL BREAST PROSTHESIS DEVELOPED CAPSULAR CONTRACTURE (BAKER GRADE UNKNOWN) IN HER LEFT BREAST POST IMPLANTATION. THE CAPSULAR CONTRACTURE WAS DIAGNOSED DURING AN OFFICE VISIT. AS A RESULT, THE PATIENT UNDERWENT UNILATERAL EXPLANTATION AND REPLACEMENT WITH A MENTOR MEMORYGEL XTRA BREAST IMPLANT 605CC GEL BREAST PROSTHESIS ON (B)(6) 2023.
Patient problems: Capsular Contracture
MAUDE report 18207490 ↗InjuryOct 31, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WHO UNDERWENT BREAST AUGMENTATION REVISION WITH 755CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS EXPERIENCED BILATERAL BAKER GRADE IV CAPSULAR CONTRACTURE POST PROCEDURE. THE IS WAS ACCOMPANIED BY PAIN, HARDNESS, AND THE DEVICES SITTING HIGHER. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. SEE 1645337-2023-12883 FOR CONTRALATERAL PROSTHESIS REPORT.
Patient problems: Capsular Contracture
MAUDE report 18043929 ↗
Other Prosthesis, breast, noninflatable, internal, silicone gel-filled devices
- MENTOR MemoryGel Breast Implant 3503501BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,167 reports
- MENTOR MemoryGel Breast Implant 3504001BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,130 reports
- MENTOR MemoryGel Breast Implant 3504004BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,012 reports
- MENTOR MemoryGel Breast Implant 3504504BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,754 reports
- MENTOR MemoryGel Breast Implant 3503001BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,752 reports
- MENTOR MemoryGel Breast Implant 3503504BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,749 reports
- MENTOR MemoryGel Breast Implant 3504501BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,690 reports
- MENTOR MemoryGel Breast Implant 3503251BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,684 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page