Skip to content

INTRAVIA 2B8012Recalls & reported problems

Baxter International Inc. · INTRAVIA Container, 250 mL Capacity · Class 2

Quick summary

This Baxter INTRAVIA container, a plastic intravenous fluid bag, has 114 adverse event reports on file. Many describe small plastic pieces or other material found inside the bag before use, and some describe leaks or ports coming apart. The reports include no deaths and no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
119
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 119 adverse event reports that list this exact device (DI 00085412003191), received between Dec 28, 2017 and Sep 30, 2026 — classified as 119 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (65 reports).

3 other recalls cover devices cleared under the same FDA submission (K964853); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 6 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 4 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material65
  • Fluid/Blood Leak41
  • Disconnection8
  • Detachment of Device or Device Component5
  • Material Puncture/Hole4
  • Break2
  • Degraded2
  • Material Split, Cut or Torn2

Patient problems

  • No Clinical Signs, Symptoms or Conditions78
  • No Patient Involvement40
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 119 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Fluid/Blood Leak

    It was reported that an Empty INTRAVIA Container was actively leaking from the port and seam during inspection, labeling, and packaging activities. There was no patient involvement. No additional information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941981 ↗
  • MalfunctionSep 18, 2026Device Contamination with Chemical or Other Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829550 ↗
  • MalfunctionSep 17, 2026Device Contamination with Chemical or Other Material

    It was reported that four (4) Empty INTRAVIA Containers had pieces of the port membrane floating in the fluorouracil solution. This occurred during preparation of fluorouracil home infusions. There was no patient involvement. No additional information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805438 ↗
  • MalfunctionSep 15, 2026Device Contamination with Chemical or Other Material

    It was reported that an INTRAVIA Container contained particulate matter inside the solution/fluid path. The particulate matter was described as 'very fine, almost hair-like' material floating within the solution. The issue was observed during setup after Sterile Water for Injection, dextrose 50% vial, sodium chloride 4 mEq/mL intravenous solution, and heparin (PF) 100 unit/mL were added for a neonatal custom intravenous infusion. The bag was quarantined and a new bag was compounded to resolve the issue. There was no patient involvement. No additional information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50766681 ↗
  • MalfunctionAug 20, 2026Device Contamination with Chemical or Other Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50355488 ↗
  • MalfunctionAug 20, 2026Device Contamination with Chemical or Other Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50355486 ↗
  • MalfunctionMay 1, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AT LEAST TWO (2) 250 ML INTRAVIA CONTAINERS HAD CLEAR, STRAND-LIKE CONTAMINANTS FLOATING IN THE BAGS. THIS WAS OBSERVED WHEN PERFORMING BIANNUAL MEDIA FILLS. THE CONTAMINANTS WERE FURTHER DESCRIBED AS "SMALL WISPY STRANDS" AND THE STRANDS APPEARED TO BE PLASTIC. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25059722 ↗
  • MalfunctionApr 29, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A 250ML INTRAVIA CONTAINER HAD PARTICULATE MATTER/FOREIGN MATERIAL LOCATED INSIDE THE SOLUTION/FLUID PATH. AFTER THE PRODUCT WAS FILLED WITH SOLUTION (AFTER ADDING ANOTHER MEDICATION OR AFTER RECONSTITUTION), THE PARTICULATE MATTER WAS OBSERVED NEAR THE PERIPHERY. THIS ISSUE WAS OBSERVED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25037716 ↗
  • MalfunctionApr 15, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT TWO (2) 250 ML INTRAVIA CONTAINERS HAVE SMALL PLASTIC PIECES INSIDE OF THE BAGS. THIS WAS OBSERVED DURING USE IN THE PHARMACY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887757 ↗
  • MalfunctionMar 25, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN INTRAVIA CONTAINER HAD SMALL, HARD PLASTIC PIECES INSIDE OF THE BAG. THE ISSUE WAS NOTICED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694949 ↗
  • MalfunctionMar 25, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN INTRAVIA CONTAINER HAD SMALL, HARD PLASTIC PIECES INSIDE OF THE BAGS. THE ISSUE WAS NOTICED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694948 ↗
  • MalfunctionMar 25, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN INTRAVIA CONTAINER HAD SMALL, HARD PLASTIC PIECES INSIDE OF THE BAGS. THE ISSUE WAS NOTICED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694271 ↗
  • MalfunctionMar 3, 2026Fluid/Blood Leak; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PORT OF AN INTRAVIA CONTAINER 250 ML CAPACITY WAS SEPARATED. THIS OCCURRED AFTER REMOVING THE TAMPER PROOF SEAL FROM THE BAG PORT AND CLEANING THE PORT WITH ALCOHOL. THE PORT THEN SEPARATED FROM THE BAG, CAUSING SOME OF THE DRUG TO DRIP OUT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24492893 ↗
  • MalfunctionFeb 10, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THERE WAS A VERY SMALL PIECE OF PLASTIC WAS FLOATING IN A 250ML INTRAVIA CONTAINER. THE PARTICULATE MATTER WAS DISCOVERED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24309224 ↗
  • MalfunctionSep 25, 2025Failure to Disconnect; Failure to Disconnect

    IT WAS REPORTED THAT AN UNSPECIFIED SPIKE/TUBING WAS DIFFICULT TO REMOVE FROM AN UNSPECIFIED AMOUNT OF 250ML INTRAVIA CONTAINERS. THIS ISSUE OCCURRED WHILE CHANGING THE EPIDURAL BAGS. IT WAS UNKNOWN IF A PATIENT WAS CONNECTED AT THE TIME OF THIS EVENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23151944 ↗
  • MalfunctionSep 19, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT TWO (2) 250 ML INTRAVIA CONTAINERS HAD PLASTIC PARTICLES INSIDE THE BAG. THIS ISSUE WAS IDENTIFIED DURING COMPOUNDING WHILE ADDING SOLUTION/DRUG TO THE BAG AND/OR WHEN REMOVING AIR FROM BAG. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23098176 ↗
  • MalfunctionSep 19, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT TWO (2) 250 ML INTRAVIA CONTAINERS HAD PLASTIC PARTICLES INSIDE THE BAG. THIS ISSUE WAS IDENTIFIED DURING COMPOUNDING WHILE ADDING SOLUTION/DRUG TO THE BAG AND/OR WHEN REMOVING AIR FROM BAG. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23098051 ↗
  • MalfunctionJun 13, 2025Disconnection; Loose or Intermittent Connection; Disconnection; Loose or Intermittent Connection

    IT WAS REPORTED THAT THE TIP PROTECTORS (BLUE PORT) OF SIXTY-THREE (63) 250 ML INTRAVIA CONTAINERS WERE FALLING OFF. THIS ISSUE WAS FURTHER DESCRIBED AS LOOSE TIP PROTECTORS FOUND DURING AN INSPECTION IN THE POST COMPOUNDING DEPARTMENT PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22202559 ↗
  • MalfunctionJun 9, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THE BLUE PORT CAPS WERE FALLING OFF FROM FIFTY SEVEN (57) 250 ML CAPACITY INTRAVIA CONTAINERS. THE ISSUE WAS FURTHER DESCRIBED AS, "THE BLUE PORT WAS DETACHING FROM THE BAGS". THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22170799 ↗
  • MalfunctionApr 17, 2025Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn

    IT WAS REPORTED THAT AN INTRAVIA CONTAINER 250 ML CAPACITY LEAKED. THE ISSUE WAS FURTHER DESCRIBED AS; THE BOTTOM RIGHT CURVE WAS OBSERVED TO BE TORN. THIS WAS DISCOVERED DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21856352 ↗
  • MalfunctionApr 2, 2025Contamination; Contamination

    IT WAS REPORTED THAT A 250ML INTRAVIA CONTAINER HAD A BLACK MARKING INSIDE THE BAG. THIS WAS OBSERVED AFTER COMPOUNDING WITH CYCLOPHOSPHAMIDE. THE TECHNICIAN ATTEMPTED TO REMOVE THE BLACK MARKING AND REALIZED THE MARKING WAS INSIDE THE BAG ON THE TOP (EMBEDDED). THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21756408 ↗
  • MalfunctionFeb 25, 2025Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    IT WAS REPORTED THAT A 250 ML INTRAVIA CONTAINER HAD A SMALL PUNCTURE/PINHOLE ON THE BACK OF THE BAG WHICH RESULTED IN A LEAK. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21462179 ↗
  • MalfunctionFeb 12, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT UNSPECIFIED PARTICULATES WERE OBSERVED IN FOUR (04) INTRAVIA CONTAINERS 250 ML CAPACITY. THE DEVICES CONTAINED UNSPECIFIED SOLUTIONS. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21365667 ↗
  • MalfunctionFeb 3, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 250ML INTRAVIA CONTAINER LEAKED FROM THE NEEDLE STRIKE (SEPTUM) ON THE DOSING PORT. THIS WAS NOTICED BEFORE PATIENT USE. THE BAG CONTAINED FENTANYL 2MCG/ML-BUPIVACAINE 0.1%. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21294686 ↗
  • MalfunctionJan 6, 2025Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn

    IT WAS REPORTED THAT AN INTRAVIA CONTAINER 250 ML CAPACITY LEAKED FROM A TEAR ON THE SIDE OF THE DOSING PORT WHERE THE PORT TUBING FROM THE BAG MEETS THE PORT INSERT. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21080136 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K964853. One submission often covers many models, so check the product and lot details in each record.

  • Z-0485-2018 · Jul 27, 2017 · Terminated
    Empty IntraVia Container with Non-DEHP Fluid Path, 500mL Containers are intended for use in the preparation and administration of drug admixtures.
  • Z-0084-2015 · Sep 16, 2014 · Terminated
    Baxter 2B8011 Empty INTRA VIA Container with PVC Ports, Sterile fluid path, 150 ml, Admixture Products and Accessories. The lntraVia Empty Plastic Container is intended for use i
  • Z-0085-2015 · Sep 16, 2014 · Terminated
    Baxter 2B8013 Empty INTRA VIA Container with PVC Ports (Sterile, nonpyrogenic fluid path) ,500 ml. Admixture Products and Accessories The lntraVia Empty Plastic Container is in

Other CONTAINER, I.V. devices

All CONTAINER, I.V. devices →