Baxter Healthcare SA · HomeChoice Automated PD Set with Cassette and 4-Prong Luers · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Baxter HomeChoice peritoneal dialysis set has 373 reports, with no deaths and 3 injuries. Most describe malfunctions such as fluid leaking, contamination, loose connections and air-in-line alarms, and 2 reports describe peritonitis that a reporter linked to a cassette. No recalls name this device. Recent reports rose from 30 to 62 compared with the period before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 373
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 373 adverse event reports that list this exact device (DI 00085412030531), received between Apr 2, 2018 and Sep 23, 2026 — classified as 2 injuries, 371 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (175 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak175
- Device Contamination with Chemical or Other Material70
- Protective Measures Problem25
- Incomplete or Inadequate Connection23
- Loose or Intermittent Connection22
- Disconnection21
- Break20
- Component Missing19
Patient problems
- No Clinical Signs, Symptoms or Conditions311
- No Consequences Or Impact To Patient56
- No Patient Involvement5
- Abdominal Pain1
- No Known Impact Or Consequence To Patient1
- Peritonitis1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 373 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50875526 ↗MalfunctionJun 12, 2026Fluid/Blood Leak; Protective Measures Problem
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM. IT WAS NOT SPECIFIED IF THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING AN UNSPECIFIED STEP OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS REPORTED THAT THERE WAS A LEAK FROM AN UNSPECIFIED LOCATION OF THE CASSETTE THAT LED TO THIS ALARM. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25484374 ↗MalfunctionMay 27, 2026Scratched Material
IT WAS REPORTED THAT THERE WERE SCRATCHES ON AN UNSPECIFIED QUANTITY OF HOMECHOICE CASSETTES. THIS WAS IDENTIFIED DURING SETUP OF THE DEVICES FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25302611 ↗MalfunctionMay 22, 2026Fitting Problem; Protective Measures Problem
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM. IT WAS NOT REPORTED IF THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS REPORTED THAT THE "CONNECTOR DOES NOT SCREW PROPERLY ONTO THE LINES" BETWEEN THE SUPPLY LINE OF THE HOMECHOICE CASSETTE AND THE SUPPLY BAG THAT LED TO THIS ALARM. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25262804 ↗MalfunctionMay 22, 2026Loose or Intermittent Connection; Protective Measures Problem
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM. IT WAS NOT REPORTED IF THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS REPORTED THAT THERE WAS A LOOSE CONNECTION BETWEEN THE SUPPLY LINE OF THE HOMECHOICE CASSETTE AND THE SUPPLY BAG THAT LED TO THIS ALARM. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257351 ↗MalfunctionMay 15, 2026Disconnection
IT WAS REPORTED THAT THE PULL RING CAP ON AN UNSPECIFIED LINE OF THE HOMECHOICE CASSETTE DETACHED. THIS OCCURRED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25182506 ↗MalfunctionMay 14, 2026Component Missing
IT WAS REPORTED THAT THE PULL RING CAP ON AN UNSPECIFIED LINE OF THE HOMECHOICE CASSETTE WAS MISSING. THIS WAS IDENTIFIED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25173700 ↗MalfunctionMay 7, 2026Disconnection
IT WAS REPORTED THAT THE PULL RING CAP DETACHED FROM THE TUBING OF THE HOMECHOICE CASSETTE. THIS OCCURRED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25113588 ↗MalfunctionMay 5, 2026Incomplete or Inadequate Connection
IT WAS REPORTED THAT THERE WAS A CONNECTION ISSUE BETWEEN THE HOMECHOICE CASSETTE AND PHYSIONEAL BAG. THE CONNECTION WAS ¿FUTILE AND NOT TIGHT.¿ THIS OCCURRED DURING PRIME OF PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25085564 ↗MalfunctionMay 5, 2026Fitting Problem
IT WAS REPORTED THAT FOUR (4) HOMECHOICE CASSETTES HAD A CONNECTION ISSUE; DESCRIBED AS "WHILE OPENING THE CASSETTE PATIENT CABLE LID AND CHECKING, THERE IS NO THREADED GROOVE ON THE INSIDE OF THE PATIENT CABLE INLET, SO IT CANNOT BE CONNECTED TO THE INSIDE OF THE BODY CONNECTOR¿. THIS WAS IDENTIFIED DURING SETUP AND PREPARATION FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25082970 ↗MalfunctionMay 4, 2026Fluid/Blood Leak; Protective Measures Problem
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS REPORTED THAT THERE WAS A LEAK AT THE CONNECTION BETWEEN SUPPLY LINE OF THE HOMECHOICE CASSETTE AND THE SUPPLY BAG THAT LED TO THIS ALARM. THE PATIENT WAS ADVISED TO END THE THERAPY SESSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25069099 ↗MalfunctionMay 1, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A HOMECHOICE CASSETTE LEAKED AT THE DIALYSIS SOLUTION INLET. THIS OCCURRED DURING PREPARATION OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY; DESCRIBED AS THIS OCCURRED IN THE PROCESS OF CONNECTING THE PHYSIONEAL BAG ¿TO THE REFILL LINE¿. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25056371 ↗MalfunctionApr 29, 2026Material Puncture/Hole
IT WAS REPORTED THAT THERE WAS A TEAR IN THE HOSE (TUBING) OF A HOMECHOICE CASSETTE. THIS WAS IDENTIFIED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25033066 ↗MalfunctionApr 28, 2026Fitting Problem
IT WAS REPORTED THAT THERE WAS A CONNECTION ISSUE BETWEEN THE HOMECHOICE CASSETTE AND PHYSIONEAL BAG; DESCRIBED AS ¿THE INJECTION PART WAS SEPARATED.¿ THIS OCCURRED DURING SETUP OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25016097 ↗MalfunctionApr 27, 2026Fluid/Blood Leak; Protective Measures Problem
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS REPORTED THAT THERE WAS A LEAK FROM AN UNSPECIFIED LOCATION THAT LED TO THIS ALARM. DIALYSATE WAS FLOWING ON THE FLOOR UNDER THE CYCLER. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25004917 ↗MalfunctionApr 21, 2026Material Puncture/Hole
IT WAS REPORTED THAT THERE WAS A HOLE IN THE SET OF THE HOMECHOICE CASSETTE. THIS WAS IDENTIFIED DURING SETUP OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24942272 ↗MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT TWO (2) HOMECHOICE CASSETTES WERE ¿SQUASHED.¿ THIS WAS IDENTIFIED BEFORE USE OF THE DEVICES FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24940613 ↗MalfunctionApr 20, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A HOMECHOICE CASSETTE LEAKED FROM AN UNSPECIFIED LOCATION. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24925477 ↗MalfunctionApr 16, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS A LEAK AT THE CONNECTION OF THE HEATER LINE ON THE HOMECHOICE CASSETTE AND PHYSIONEAL BAG. THIS OCCURRED DURING FILL OF PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24897109 ↗MalfunctionApr 16, 2026Fluid/Blood Leak; Loose or Intermittent Connection; Protective Measures Problem
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING FILL TWO OF FOUR OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS REPORTED THAT THERE WAS A LOOSE CONNECTION BETWEEN THE HEATER LINE OF THE HOMECHOICE CASSETTE AND THE HEATER BAG, WHICH RESULTED IN A LEAK THAT LED TO THIS ALARM. RENAL THERAPY SERVICES (RTS) ADVISED THE PATIENT TO DISCONNECT AND TO CONTACT THEIR NURSE REGARDING THE ISSUE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24895130 ↗MalfunctionApr 9, 2026Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THERE WAS A BLACK STAIN INSIDE THE TUBING OF A HOMECHOICE CASSETTE. THIS WAS IDENTIFIED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24829747 ↗MalfunctionApr 3, 2026Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THERE WERE ¿BLACK DOTS¿ INSIDE THE STERILE FLUID PATHWAY OF A HOMECHOICE AUTOMATED PD SET WITH CASSETTE. THIS WAS IDENTIFIED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781344 ↗MalfunctionApr 3, 2026Crack
IT WAS REPORTED THAT THERE WERE CRACKS IN SIX (6) HOMECHOICE CASSETTES. THIS WAS IDENTIFIED DURING USE OF THE DEVICES FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24777794 ↗MalfunctionApr 2, 2026Loose or Intermittent Connection
IT WAS REPORTED THAT THERE WAS A CONNECTION ISSUE BETWEEN THE HOMECHOICE CASSETTE AND PHYSIONEAL BAG. THE CONNECTION WAS NOT FASTENED WELL. THIS OCCURRED DURING SETUP OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24767255 ↗MalfunctionApr 2, 2026Break; Fluid/Blood Leak
IT WAS REPORTED THAT A HOMECHOICE CASSETTE WAS DAMAGED WHICH RESULTED IN A LEAK. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765604 ↗
Other System, peritoneal, automatic delivery devices
- Liberty® Select Cycler 180343Fresenius Medical Care Holdings, Inc. · System, peritoneal, automatic delivery1 recall545 deaths15,896 reports
- Liberty® 050-87216Fresenius Medical Care Holdings, Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY16 deaths9,148 reports
- Liberty® 050-87212Fresenius Medical Care Holdings, Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY12 deaths2,169 reports
- Liberty® 180111Fresenius Usa, Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY126 deaths1,716 reports
- Amia Automated PD Set with Cassette 5C5479Baxter International Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY1 recall976 reports
- INTERGRATED APD SET L5C4531Baxter International Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY811 reports
- Liberty 050-87216AFresenius Medical Care Renal Therapies Group, LLC · System, Peritoneal, Automatic Delivery1 death678 reports
- Liberty® 050-87215Fresenius Medical Care Holdings, Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY1 death279 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page