INTERLINK 2C6700Recalls & reported problems

Baxter Healthcare Corporation · Straight Type Blood Set with Standard Blood Filter, 170 to 260 Micron Filter, Injection Site, Male Luer lock Adapter · Class 2

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00085412068152), received between Apr 18, 2018 and Dec 11, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

10 other recalls cover devices cleared under the same FDA submission (K993120); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Break2
  • Backflow1
  • Contamination1
  • Detachment of Device or Device Component1
  • Loose or Intermittent Connection1
  • Material Separation1
  • No Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • Insufficient Information1
  • No Patient Involvement1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2025Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TUBING OF AN INTERLINK STRAIGHT TYPE BLOOD SET SEPARATED FROM THE DRIP CHAMBER. THE SET WAS SPIKED INTO THE BLOOD BAG (RBC) AND HUNG. WHILE PRIMING THE DRIP CHAMBER, THE INTRAVENOUS (IV) TUBING SEPARATED FROM THE DRIP CHAMBER, CAUSING BLOOD TO POUR FREELY FROM BOTH THE CHAMBER AND THE BAG UNTIL IT COULD BE RE-SPIKED WITH NEW STERILE BLOOD TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23780412 ↗
  • MalfunctionNov 14, 2025Contamination; Fluid/Blood Leak; Material Separation

    NURSE WAS ABOUT TO ADMINISTER PACKED RBC'S [RED BLOOD CELLS]. AFTER SPIKING THE BAG, NURSE WENT TO HANG AND PRIME DRIP CHAMBER. AS HE WAS PRIMING DRIP CHAMBER, IV TUBING SEPARATED FROM THE DRIP CHAMBER, CAUSING BLOOD TO POUR FREELY FROM CHAMBER AND BAG UNTIL IT COULD BE RE-SPIKED WITH NEW STERILE BLOOD TUBING. BLOOD CONTAMINATION OCCURRED OF IV PUMPS, IV POLE, IV POWER SHELF, PATIENT BED, AND FLOOR. CONTAMINATION DISINFECTED AND STERILIZED. IV TUBING NUMBER (01) 00085412068152 LOT (10) R24J19024

    Patient problems: Insufficient Information

    MAUDE report 23554090 ↗
  • MalfunctionNov 15, 2022Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT AN INTERLINK STRAIGHT TYPE BLOOD SET WOULD NOT CONNECT TO A BAG OF PLATELETS. THIS ISSUE WAS FURTHER DESCRIBED AS, ¿CONNECTOR ON END OF TUBING WOULD NOT CONNECT TO MALE ADAPTER PLUG, OR ANY OTHER PORT ON ANY ANOTHER TUBING TRIED¿. THIS ISSUE WAS IDENTIFIED DURING PRIMING. TO RESOLVE THIS ISSUE, A NEW SET WAS USED AND WORK APPROPRIATELY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15802589 ↗
  • MalfunctionMar 17, 2022Backflow; Fluid/Blood Leak; Backflow; Fluid/Blood Leak

    IT WAS REPORTED THAT AN INTERLINK STRAIGHT TYPE BLOOD SET LEAKED DURING INFUSION OF BLOOD. ALL CLAMPS, INCLUDING THE CENTRAL LINE CLAMP, WERE OPEN. THE BOTTOM INJECTION SITE PORT WAS CAPPED WITH A GREEN ALCOHOL CAP. DURING INFUSION, AN ESTIMATED 22CC OF BLOOD BACKFLOWED OUT OF THE INJECTION SITE PORT AND ONTO THE PATIENT AND BED. THE REST OF BLOOD IN THE TUBING WAS TRANSFERRED BACK INTO THE BLOOD BAG, AND A NEW SET WAS UTILIZED TO COMPLETE THE INFUSION WITH NO ISSUES NOTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13794848 ↗
  • MalfunctionFeb 8, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN INTERLINK STRAIGHT TYPE BLOOD SET WAS LEAKING FROM THE INJECTION SITE PORT. BLOOD WAS HUNG UTILIZING THE BLOOD SET AND IT WAS OBSERVED THAT BLOOD BACK FLOWED OUT OF THE INJECTION SITE PORT. THIS ISSUE WAS IDENTIFIED DURING PATIENT USE. TO RESOLVE THE ISSUE, A NEW SET WAS UTILIZED WITH NO ISSUES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13488360 ↗
  • MalfunctionSep 30, 2019No Flow; No Flow

    IT WAS REPORTED THAT THE LINES OF AN UNSPECIFIED QUANTITY OF INTERLINK BLOOD RECIPIENT SETS WERE UNABLE TO BE CLEARED. IT WAS FURTHER REPORTED THAT THE TUBING WAS TOO LONG. THIS WAS IDENTIFIED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9136000 ↗
  • MalfunctionApr 23, 2019Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. IT WAS REPORTED THAT THE SPIKE OF AN INTERLINK STRAIGHT TYPE BLOOD SET BROKE OFF IN THE BLOOD TRANSFUSION BAG, LEADING TO A LEAK. THIS OCCURRED WHEN THE NURSE SPIKED THE BAG DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 8541743 ↗
  • MalfunctionApr 18, 2018Break; Break

    IT WAS REPORTED THAT AN INTERLINK BLOOD SET FILTER BROKE OFF DURING PATIENT USE. THERE WAS NO REPORT OF A LEAK. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7441451 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K993120. One submission often covers many models, so check the product and lot details in each record.

  • Z-0193-05 · Apr 28, 2004 · Terminated
    Baxter Clearlink Intravenous (IV) Delivery System; the system consists of solution sets, extension sets and blood sets that have a luer activated valve for IV access, which allows
  • Z-1177-03 · Jul 16, 2003 · Terminated
    Fenwal Y-Type Blood/Solution Set with Large Standard Blood Filter, 10 drops/mL, 82'', 170 to 260 Micron Filter, Large Bore Tubing, Male Luer Lock Adapter, catalog 4C7776; Baxter He
  • Z-1176-03 · Jul 16, 2003 · Terminated
    Fenwal Y-Type Blood/Solution Set with Large Standard Blood Filter, 15 drops/mL, 107'', 2 Injection Sites, Male Luer Lock Adapter, catalog 4C7774; Baxter Healthcare Corporation, Dee
  • Z-1162-03 · Jul 16, 2003 · Terminated
    Fenwal Y-Type Blood/Solution Set with Large Standard Blood Filter and Pressure Pump, 15 drops/mL, 99'', 170 to 260 Micro Filter, Large Standard Blood Filter, 2 Injection Sites, Mal
  • Z-1175-03 · Jul 16, 2003 · Terminated
    Fenwal Straight Type Blood Set with Large Standard Blood filter, 15 drops/mL, 80'', 170 to 260 Micron Filter, Injection Site, Male Luer Lock Adapter, catalog 4C6766; Baxter Healthc
  • Z-1171-03 · Jul 16, 2003 · Terminated
    Baxter Y-Type Blood/Solution Set, 10 drops/mL, 115'', 80 Micron Filter, Injection Site, Male Luer Lock Adapter, catalog 2C7617; Baxter Healthcare Corporation, Deerfield, IL 60015 U
  • Z-1178-03 · Jul 16, 2003 · Terminated
    Fenwal 3-Lead Blood/Solution Set with Large Standard Blood Filter, 15 drops/mL, 82'', 170 to 260 Micron Filter, Injection Site, Male Luer Lock Adapter, catalog 4C7784; Baxter Healt
  • Z-1172-03 · Jul 16, 2003 · Terminated
    Baxter Y-Type Blood/Solution Set with Pressure Pump, 10 drops/mL, 99'', Large Standard Blood Filter, 170 to 260 Micron Filter, 2 Injection Sites, Male Luer Lock Adapter, catalog 2C
  • Z-1161-03 · Jul 16, 2003 · Terminated
    Clearlink System Y-Type Blood/Solution Set, 10 drops/mL, 112'', Standard Blood Filter, 170 to 260 Micron Filter, Luer Activated Valve, Male Luer Lock Adapter, catalog 2C8750; Baxte
  • Z-1160-03 · Jul 16, 2003 · Terminated
    Baxter Y-Type Blood/Solution Set, 10 drops/mL, 112'', Standard Blood Filter, 170 to 260 Micron Filter, Injection Site, Male Luer Lock Adapter, catalog 2C6750; Baxter Healthcare Cor

Other Set, blood transfusion devices

All Set, blood transfusion devices →