CLEARLINK 2C8921Recalls & reported problems

Baxter International Inc. · 3-Port Manifold with Check Valves, Large Bore Stopcock Extension Set with Rotating Male Luer Lock, 2 Luer Activated Valves, Male Luer Lock Adapter with Retractable Collar · Class 2

All models of CLEARLINK (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00085412071718), received between Dec 13, 2019 and Sep 10, 2020 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Disconnection” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Disconnection2
  • Fluid/Blood Leak1
  • Loose or Intermittent Connection1
  • Material Deformation1
  • No Flow1

Patient problems

  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2020Material Deformation; No Flow; Material Deformation; No Flow

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM 3-PORT MANIFOLD SETS WOULD NOT FLOW. THE EVENT WAS FURTHER DESCRIBED AS TUBING WOULD KINK EASILY AND WOULD HAVE TO BE TAPED IN SEVERAL PLACES TO KEEP IT FROM KINKING; THE PRIMARY AREA OF KINKED TUBING WAS ON EACH END OF THE MANIFOLD. THE KINKED TUBING ALSO PREVENTED NEEDED MEDICATIONS OF SALINE FLUSH FROM FULLY BEING ADMINISTERED. THIS WAS IDENTIFIED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10515040 ↗
  • MalfunctionSep 9, 2020Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM 3-PORT MANIFOLD SETS LEAKED AND FELL APART. THIS WAS FURTHER DESCRIBED AS ¿THE FLUIDS WERE LEAKING OUT OF THE EXTENSION TUBING ONTO THE FLOOR, AND ONTO THE PATIENT¿S GURNEY AND OR TABLE¿. IT WAS STATED THAT ¿THE LEAKS WERE COMING FROM WHERE THE ANESTHESIA EXTENSION TUBING CONNECTS AND MANY TIMES THE TUBING JUST FELL APART¿. THIS WAS IDENTIFIED DURING PATIENT INFUSIONS. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THESE EVENTS. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10511273 ↗
  • MalfunctionDec 13, 2019Disconnection; Disconnection

    IT WAS REPORTED AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM 3-PORT MANIFOLD GOT DISCONNECTED FROM THE DISTAL END OF AN UNKNOWN QUANTITY OF ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET. IT WAS FURTHER REPORTED, "AFTER CONNECTION WAS ESTABLISHED" THE USER WOULD HAVE TO MAKE ¿THREE QUARTERS OF TURN TO ESTABLISH CONNECTION¿. THIS ISSUE WAS IDENTIFIED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9469112 ↗
  • MalfunctionDec 13, 2019Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED LOOSE CONNECTIONS WERE OBSERVED IN AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM 3-PORT MANIFOLDS. THE EVENT WAS FURTHER DESCRIBED AS "LOOSE CONNECTIONS BETWEEN THE STOPCOCK MANIFOLD AND THE TUBING" WERE OBSERVED. THE EVENT WAS OCCURRED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9467911 ↗

Other Stopcock, i.V. Set devices