CLEARLINK/CONTINU-FLO 2C8931Recalls & reported problems
Baxter International Inc. · CONTINU-FLO Solution Set, 2 Luer Activated Valves, Male Luer Lock Adapter, 3-Port Manifold with Check Valves, Large Bore Stopcock Extension Set with Rotating Male Luer Lock, 2 Luer Activated Valves, Male Luer Lock Adapter with Retractable Collar · Class 2
All models of CLEARLINK/CONTINU-FLO (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00085412071725), received between Jan 12, 2018 and Nov 11, 2025 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Disconnection” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Disconnection10
- Fluid/Blood Leak8
- Loose or Intermittent Connection4
- Crack2
- Material Twisted/Bent2
- Air/Gas in Device1
- Connection Problem1
- Failure to Prime1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Patient Involvement5
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 11, 2025Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT THERE WAS AIR ENTERING THE LINE OF CLEARLINK CONTINU-FLO SOLUTION SET THROUGH THE MANIFOLD LOCATED DIRECTLY ABOVE THE STOP COCK. IT WAS UNKNOWN IF A PATIENT WAS CONNECTED DURING THIS EVENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523414 ↗MalfunctionOct 28, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET LEAKED AT THE Y-SITE. THE ISSUE WAS FURTHER DESCRIBED AS, THIS OCCURRED WHILE ADMINISTERING MEDICATION DURING INTUBATION THROUGH ONE OF THE PORTS CLOSEST TO THE PATIENT, RESULTING IN THE MEDICATION LEAKING FROM THE PORT DIRECTLY BELOW IT INSTEAD OF BEING DELIVERED TO THE PATIENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23408300 ↗MalfunctionAug 4, 2023Disconnection; Loose or Intermittent Connection; Disconnection; Loose or Intermittent Connection
IT WAS REPORTED THAT BOTH SIDES OF THE MANIFOLD OF A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET WAS LOOSELY CONNECTED. THIS WAS FOUND UPON REMOVING THE TUBING FROM THE PACKAGE. THE CONNECTIONS WERE TIGHTENED AND CHECKED PRIOR TO STARTING THE INFUSION. SUBSEQUENTLY, THE CONNECTIONS WERE FOUND LOOSE AGAIN AND THE ANESTHESIOLOGIST FOUND THE TUBING DISCONNECTED FROM THE MANIFOLD AND A BLOOD/MEDICATION PUDDLE ON THE FLOOR BECAUSE THE MANIFOLD CONNECTIONS HAD DISCONNECTED. THE PATIENT DID NOT MOVE FROM THE PRE-OP CHAIR WHEN THIS WAS FOUND. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17466498 ↗MalfunctionJul 27, 2023Disconnection; Loose or Intermittent Connection; Disconnection; Loose or Intermittent Connection
IT WAS REPORTED THAT BOTH SIDES OF THE MANIFOLD OF A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET WAS LOOSELY CONNECTED. THIS ISSUE WAS FURTHER DESCRIBED AS, ¿THE CONNECTIONS WERE HAND TIGHTENED TWICE PRIOR TO THE TUBING BECOMING DISCONNECTED¿. IT WAS NOT SPECIFIED WHEN IN THE PROCESS STEP THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17406919 ↗MalfunctionJul 3, 2023Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT THERE WAS NO INNER VALVE ON ONE OF THE PORTS OF A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET. THE MANIFOLD CONNECTIONS CAME LOOSE EVENT AFTER TIGHTENING.THE ISSUE OCCURRED AFTER PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17249416 ↗MalfunctionMar 15, 2022Disconnection; Disconnection
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM CONTINU-FLO SOLUTION SETS EXPERIENCED DISCONNECTIONS FROM THE MANIFOLD PORT. THE CONNECTION ISSUES OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13769157 ↗MalfunctionMar 9, 2022Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET EXPERIENCED A DISCONNECTION FROM THE MANIFOLD PORT WHICH RESULTED IN A BLOOD LEAK. THE CONNECTION ISSUE OCCURRED DURING PATIENT THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13711404 ↗MalfunctionMar 9, 2022Disconnection; Disconnection
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM CONTINU-FLO SOLUTION SETS EXPERIENCED DISCONNECTIONS FROM THE MANIFOLD PORT. THE CONNECTION ISSUES OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13711129 ↗MalfunctionDec 29, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM CONTINU-FLO SOLUTION SETS LEAKED FROM AN UNSPECIFIED LOCATION. THIS OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11091056 ↗MalfunctionDec 9, 2020Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED FIFTEEN (15) CLEARLINK SYSTEM CONTINU-FLO SOLUTION SETS WOULD NOT CONNECT. THE DEFECT WAS FURTHER DESCRIBED AS "THE END WOULD NOT SCREW TOGETHER". THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 10976814 ↗MalfunctionNov 13, 2019Crack; Material Twisted/Bent; Crack; Material Twisted/Bent
IT WAS REPORTED THAT THE MANIFOLD OF A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET WAS CRACKED, BENT AND NOT WORKING. THIS WAS IDENTIFIED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 9318326 ↗MalfunctionNov 11, 2019Crack; Material Twisted/Bent; Crack; Material Twisted/Bent
IT WAS REPORTED THAT "THE MANIFOLD ON EITHER END" OF A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET WAS "CRACKED, BENT AND NOT WORKING". THIS WAS IDENTIFIED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 9305882 ↗MalfunctionApr 29, 2019Failure to Prime; Failure to Prime
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. IT WAS REPORTED THAT A CLEARLINK CONTINU-FLO SOLUTION SET WOULD NOT PRIME. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8560704 ↗MalfunctionApr 24, 2019Connection Problem; Connection Problem
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF CLEARLINK CONTINU-FLO SOLUTION SETS EXPERIENCED CONNECTION ISSUES. THE CLEAR PLASTIC PIECE ON THE END OF THE TUBING WHICH ATTACHES TO THE PATIENT WOULD NOT TWIST TO LOCK ONTO THE LUER LOCK. THIS OCCURRED DURING AN UNSPECIFIED PROCESS STEP. IT WAS NOT SPECIFIED WHETHER THERE WAS PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8547299 ↗MalfunctionJan 12, 2018Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK CONTINU-FLO SOLUTION SET DISCONNECTED DURING INFUSION, LEADING TO A BLOOD LEAK. THE AMOUNT OF PATIENT BLOOD LOSS WAS UNKNOWN. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7187599 ↗MalfunctionJan 12, 2018Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK CONTINU-FLO SOLUTION SET SEPARATED FROM THE ADMINISTRATION PORT. THIS OCCURRED DURING INFUSION. THE REPORTER STATED THAT THE PATIENT ¿BLED OUT¿ (EVENT NOT FURTHER CLARIFIED). THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7187587 ↗MalfunctionJan 12, 2018Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK CONTINU-FLO SOLUTION SET SEPARATED FROM THE ADMINISTRATION PORT. THIS OCCURRED DURING INFUSION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7187549 ↗MalfunctionJan 12, 2018Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK CONTINU-FLO SOLUTION SET SEPARATED FROM THE ADMINISTRATION PORT. THIS OCCURRED DURING INFUSION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7187522 ↗MalfunctionJan 12, 2018Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK CONTINU-FLO SOLUTION SET SEPARATED FROM THE ADMINISTRATION PORT. THIS OCCURRED DURING INFUSION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7187520 ↗
Other Stopcock, i.V. Set devices
- Connecta, Luer-Lok 394910BD Switzerland Sàrl · Stopcock, i.V. Set3 recalls8 reports
- Outlook® 490361B. Braun Medical Inc. · Stopcock, i.V. Set3 recalls0 reports
- Blood Administration Set 490530B. Braun Medical Inc. · Stopcock, i.V. Set2 recalls11 reports
- Medex MX9341LIcu Medical, Inc. · Stopcock, i.V. Set63 reports
- ICU Medical AC135Icu Medical, Inc. · Stopcock, i.V. Set1 recall10 reports
- IV Administration Set 490067B. Braun Medical Inc. · Stopcock, i.V. Set1 recall9 reports
- ICU Medical MC330174Icu Medical, Inc. · Stopcock, i.V. Set1 recall4 reports
- ICU Medical Z2587Icu Medical, Inc. · Stopcock, i.V. Set1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page