UltraSet CAPD Disposable Disconnect Y-Set L5C4366Recalls & reported problems
Baxter Healthcare Corporation · The UltraSet products are single use devices that are intended for the drainage and infusion of peritoneal dialysis solution. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Baxter UltraSet CAPD disposable disconnect Y-set is used to drain and infuse peritoneal dialysis solution. The 100 reports mostly describe fluid leaks and connection problems, with 3 injuries and no deaths. There are no recalls naming this device. Report volume was similar to the previous 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 100
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 100 adverse event reports that list this exact device (DI 00085412090085), received between Mar 6, 2018 and Aug 25, 2026 — classified as 3 injuries, 97 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (55 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak55
- Incomplete or Inadequate Connection11
- Material Puncture/Hole9
- Fitting Problem8
- Disconnection6
- Break5
- Component Missing5
- Loose or Intermittent Connection4
Patient problems
- No Clinical Signs, Symptoms or Conditions71
- No Consequences Or Impact To Patient24
- No Patient Involvement2
- Peritonitis2
- Abdominal Pain1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 100 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 25, 2026Fitting Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50441795 ↗MalfunctionJun 23, 2026Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT THERE WAS A HOLE IN THE BAG OF AN ULTRA SET CAPD DISPOSABLE DISCONNECT Y-SET WHICH RESULTED IN A LEAK. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25598032 ↗MalfunctionJun 9, 2026Fitting Problem
IT WAS REPORTED THAT THERE WAS A CONNECTION ISSUE WITH AN UNSPECIFIED QUANTITY OF CAPD DISCONNECT Y-SETS; DESCRIBED AS ¿THE MANUAL DRAIN BAGS WERE NOT FITTING PROPERLY¿. THIS OCCURRED DURING USE OF THE DEVICES FOR AUTOMATED PERITONEAL DIALYSIS (PD) THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25445420 ↗InjuryJan 12, 2026Fluid/Blood Leak
A PERITONEAL DIALYSIS (PD) PATIENT EXPERIENCED PERITONITIS MANIFESTED BY CLOUDY EFFLUENT AND ABDOMINAL PAIN. THE PATIENT WAS USING CAPD (CONTINUOUS AMBULATORY PD) STERILE FLUID PATH Y SET. IT WAS REPORTED THAT THE "SAMPLE PORT WITH THE BLUE PIECE WAS LEAKING WHEN DOING THE EXCHANGE", FURTHER REPORTED AS "WHILE FLUSHING THE LINE BEFORE IT STARTS IT LEAKS". THE PATIENT RECEIVED UNSPECIFIED TREATMENT FOR THE EVENT. IT WAS NOT REPORTED IF THE PATIENT WAS HOSPITALIZED FOR THE EVENT. AT THE TIME OF THIS REPORT, THE PATIENT OUTCOME WAS NOT REPORTED. IT WAS REPORTED THAT PD THERAPY WAS PUT ON HOLD. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Abdominal Pain; Peritonitis
MAUDE report 24044562 ↗MalfunctionJan 7, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A CAPD DISCONNECT Y-SET LEAKED ON THE CARPET FROM THE SEAM OF THE BAG. THIS OCCURRED DURING USE OF THE DEVICE FOR CONTINUOUS AMBULATORY PERITONEAL DIALYSIS (CAPD) THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23993932 ↗MalfunctionOct 16, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DRAIN BAG OF AN ULTRA SET CAPD DISPOSABLE DISCONNECT Y-SET LEAKED FLUID ONTO THE FLOOR. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23311884 ↗MalfunctionSep 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS A LEAK BETWEEN THE BAG AND TUBING OF AN ULTRA SET CAPD DISPOSABLE DISCONNECT Y-SET. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038890 ↗MalfunctionSep 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS A LEAK BETWEEN THE BAG AND TUBING OF AN ULTRA SET CAPD DISPOSABLE DISCONNECT Y-SET. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038713 ↗MalfunctionJun 26, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY (AT LEAST A FEW) OF ULTRA SET CAPD DISPOSABLE DISCONNECT Y-SETS LEAKED. THIS OCCURRED DURING SETUP/PREPARATION OF THE DEVICES FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22337643 ↗MalfunctionMay 7, 2025Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THERE WAS A CONNECTION ISSUE BETWEEN THIRTY (30) UNITS OF THE CAPD DISCONNECT Y-SET AND THE MINICAP TRANSFER SET. THE Y-SETS WERE NOT PROPERLY CONNECTING TO THE TRANSFER SET. THIS OCCURRED "WHILE ATTEMPTING TO CONNECT THE BAG TO THE PATIENT'S TRANSFER SET" FOR USE IN PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21977884 ↗MalfunctionApr 24, 2025Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THERE WAS A CONNECTION ISSUE WITH A CAPD DISCONNECT Y-SET; DESCRIBED AS ¿UNABLE TO HOOK UP THE Y SET TO DRAIN- IT WOULD NOT FIT ON¿ THE PATIENT¿S CATHETER. THIS OCCURRED DURING SETUP/PREPARATION OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21904805 ↗MalfunctionNov 25, 2024Break; Break
IT WAS REPORTED THAT THE WATER BAG OF TWO (2) CAPD DISPOSABLE DISCONNECT Y-SETS WERE BROKEN. THIS WAS OBSERVED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20773896 ↗MalfunctionSep 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE BAG OF A CAPD DISCONNECT Y SET LEAKED. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20295413 ↗MalfunctionSep 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE BAG OF A CAPD DISCONNECT Y SET LEAKED. THIS OCCURRED DURING DRAINING OF PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20294218 ↗MalfunctionSep 6, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CAPD DISCONNECT Y SET LEAKED. THIS OCCURRED DURING DRAIN OF PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20157389 ↗MalfunctionAug 28, 2024Unsealed Device Packaging; Unsealed Device Packaging
IT WAS REPORTED THAT TWELVE (12) CAPD DISPOSABLE DISCONNECT Y-SETS "ARRIVED OPENED." THE BAGS WERE NOT SEALED PROPERLY. THIS WAS OBSERVED PRIOR TO USE OF THE DEVICES FOR PERITONEAL DIALYSIS (PD) THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20091214 ↗MalfunctionAug 27, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CAPD DISCONNECT Y SET LEAKED; FURTHER DESCRIBED AS "LEAKING UPON INITIAL DRAIN AT THE TOP OF THE DRAIN BAG WHERE THE TUBING CONNECTS TO THE BAG" AND "FLUID ON THE FLOOR". THIS WAS OBSERVED DURING THE INITIAL DRAIN STEP OF CONTINUOUS AMBULATORY PERITONEAL DIALYSIS (CAPD) THERAPY; THE DEVICE WAS CLAMPED AND DISCONNECTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20076916 ↗MalfunctionAug 18, 2024Disconnection; Disconnection
IT WAS REPORTED THAT AN ULTRA SET CAPD DISPOSABLE DISCONNECT Y-SET CAME APART AT THE Y-SET. THIS WAS NOTICED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19999316 ↗MalfunctionAug 15, 2024Disconnection; Disconnection
IT WAS REPORTED THAT THE ULTRA SET CAPD DISPOSABLE DISCONNECT Y-SET CAME APART AT THE YSET. THIS OCCURRED DURING USE OF THE DEVICES FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19992090 ↗MalfunctionAug 14, 2024Break; Incomplete or Inadequate Connection; Break; Incomplete or Inadequate Connection
IT WAS REPORTED THAT THE "THREADS" OF A CAPD DISCONNECT Y SET T WERE STRIPPED AND WOULD NOT LOCK IN PLACE OR TIGHTEN. THIS OCCURRED DURING TRAINING FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19985869 ↗MalfunctionJul 5, 2024Fitting Problem; Fitting Problem
IT WAS REPORTED THAT A CAPD DISCONNECT Y SET HAD A CONNECTION ISSUE; FURTHER DESCRIBED AS ¿THE THREADS SEEMED TO SLIP AROUND THE END OF THE TRANSFER SET AND NEVER MAKE A TIGHT CONNECTION." UPON INSPECTION OF THE THREADS OF THE Y SET IT APPEARED THAT THE THREADS WERE NOT CORRECT. THIS OCCURRED DURING SET-UP AND PREPARATION OF THE DEVICE FOR PERITONEAL DIALYSIS (PD) THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19683287 ↗MalfunctionJun 27, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED A PERITONEAL DIALYSIS PATIENT OBSERVED HOLES IN 23 ULTRASET CAPD DISPOSABLE DISCONNECT Y-SETS PRIORI TO USE. THE CAUSE OF THE HOLES WAS UNKNOWN. THE CAREGIVER REPORTED THEY DID NOT NOTICE ANY DAMAGE WITH THE OVERPOUCH PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19629338 ↗InjuryJun 27, 2024Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED A PERITONEAL DIALYSIS PATIENT EXPERIENCED A LEAK FROM AN ULTRASET CAPD DISPOSABLE DISCONNECT Y-SET. IT WAS FURTHER REPORTED THE PATIENT NOTICED A HOLE ON THE SET. THE CAUSE OF THE HOLE WAS UNKNOWN. SUBSEQUENTLY, THE PATIENT EXPERIENCED PERITONITIS. HOSPITALIZATION, TREATMENT, PATIENT OUTCOME AND ACTION TAKEN WITH PD THERAPY WERE NOT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Peritonitis
MAUDE report 19629209 ↗MalfunctionJun 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CAPD DISCONNECT Y SETS WERE DAMAGED; FURTHER DESCRIBED AS "COMING APART AT THE Y-SET". THIS WAS OBSERVED DURING AN UNSPECIFIED PROCESS STEP OF PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19581412 ↗MalfunctionJun 6, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT A CAPD DISCONNECT Y SET HAD A PINHOLE IN THE BOTTOM CORNER OF THE DRAIN BAG. THIS WAS OBSERVED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19476858 ↗
Other SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE devices
- MiniCap 5C4482Baxter Healthcare Corporation · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE3 recalls5 deaths2,961 reports
- MiniCap 5C4483Baxter Healthcare Corporation · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE4 recalls433 reports
- Safe-Lock® 050-95018Fresenius Medical Care Holdings, Inc. · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE2 recalls509 reports
- MINICAP WITH POVIDONE-IODINE SOLUTION 5C4466PBaxter Healthcare Corporation · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE1 recall4 deaths347 reports
- MINICAP EXTENDED LIFE PD TRNS 5C4482ABaxter Healthcare Corporation · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE193 reports
- stay.safe® 050-95012Fresenius Medical Care Holdings, Inc. · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE1 recall48 reports
- stay.safe® 050-95004Fresenius Medical Care Holdings, Inc. · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE1 recall43 reports
- stay.safe® 050-95005Fresenius Medical Care Holdings, Inc. · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE1 recall34 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page