EXACTAMIX EXACTA-MRecalls & reported problems
Baxter Healthcare Corporation · ExactaMix Pharmacy Compounding Device · Class 2
All models of EXACTAMIX (28)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00085412586557), received between Aug 1, 2023 and Sep 23, 2026 — classified as 2 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Device Emits Odor” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Emits Odor3
- Contamination2
- Excess Flow or Over-Infusion2
- Break1
- Component Missing1
- Material Split, Cut or Torn1
- Precipitate in Device or Device Ingredient1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Device Emits Odor
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50875525 ↗MalfunctionApr 22, 2026Device Emits Odor
IT WAS REPORTED THAT A EXACTAMIX PHARMACY COMPOUNDING DEVICE HAD A BURNING SMELL. THIS OCCURRED DURING PROGRAMMING/SETUP. THE EVENT OCCURRED IN THE PHARMACY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24960602 ↗InjuryFeb 11, 2026Excess Flow or Over-Infusion
IT WAS REPORTED A NEONATAL PATIENT DEVELOPED HYPERMAGNESEMIA WHEN AN EXACTAMIX PHARMACY COMPOUNDING DEVICE DISPENSED MAGNESIUM WHEN IT WAS NOT PROGRAMMED. A NEONATE IN THE NEONATAL INTENSIVE CARE UNIT (NICU) WAS FOUND TO HAVE ¿AN EXTREMELY ELEVATED MAGNESIUM LEVEL¿ (ACTUAL MAGNESIUM LEVELS NOT REPORTED). THE ORDER IN THE HOSPITAL ELECTRONIC SYSTEM WAS ZERO FOR MAGNESIUM IN THE TOTAL PARENTERAL NUTRITION BAG; HOWEVER, THE LAB TESTING SHOWED THE MAGNESIUM LEVEL WAS 121.61 MG/DL. BASED ON THE TOTAL VOLUME OF 229.01, THE AMOUNT OF MAGNESIUM DISPENSED INTO THE BAG WAS 277.08MG OR 0.55ML. THE STAFF COMPARED THE VOLUME AND ANALYTES BETWEEN THE VALUES GENERATED DURING TESTING TO ALL OTHER BAGS PREPARED ON THE DATE OF THE EVENT. THEY ELIMINATED THE POSSIBILITY OF 2 LINES/PORTS BEING MIXED UP. ADDITIONALLY, THEY DID NOT IDENTIFY ANY OTHER ORDERS THAT MATCHED THE CALCULATED VOLUMES ASIDE FROM ONE DOSE THAT MATCHED THE 1.6ML WHICH WAS FOR AN ADULT ORDER. THE ADULT BAG WAS 980ML AND THE BAG OF THE PATIENT INVOLVED WAS 500ML. THE REPORTER HAS INSTRUCTED STAFF TO REMOVE THE MAGNESIUM FROM THE COMPOUNDER AND ADD IT MANUALLY. TREATMENT AND PATIENT OUTCOME WERE NOT REPORTED. NO FURTHER INFORMATION W
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24316949 ↗InjuryFeb 11, 2026Excess Flow or Over-Infusion
IT WAS REPORTED A NEONATAL PATIENT DEVELOPED HYPERMAGNESEMIA WHEN AN EXACTAMIX PHARMACY COMPOUNDING DEVICE DISPENSED TOO MUCH MAGNESIUM. A NEONATE IN THE NEONATAL INTENSIVE CARE UNIT (NICU) WAS FOUND TO HAVE A MAGNESIUM LEVEL ¿2.5 TIMES¿ MORE THAN THE ACCEPTABLE LEVELS (ACTUAL MAGNESIUM LEVELS NOT REPORTED). THE ORDER IN THE HOSPITAL ELECTRONIC SYSTEM WAS 0.59ML TO BE ADDED TO THE TOTAL PARENTERAL NUTRITION BAG. THE LAB TESTED THE BAG FOUR TIMES WITH AN AVERAGE VALUE OF 179.44 MG/100ML. THE TOTAL VOLUME OF THE BAG WAS 445.36 ML. THE REPORTER STATED THAT APPROXIMATELY 1.6 ML OF MAGNESIUM WAS DISPENSED DURING PREPARATION BASED ON THEIR CALCULATION OF 799.15MG X 1ML/500MG = 1.598 ML. THE STAFF COMPARED THE VOLUME AND ANALYTES BETWEEN THE VALUES GENERATED DURING TESTING TO ALL OTHER BAGS PREPARED ON THE DATE OF THE EVENT. THEY ELIMINATED THE POSSIBILITY OF 2 LINES/PORTS BEING MIXED UP. ADDITIONALLY, THEY DID NOT IDENTIFY ANY OTHER ORDERS THAT MATCHED THE CALCULATED VOLUMES ASIDE FROM ONE DOSE THAT MATCHED THE 1.6ML WHICH WAS FOR AN ADULT ORDER. THE ADULT BAG WAS 980ML AND THE BAG OF THE PATIENT INVOLVED WAS 500ML. THE REPORTER HAS INSTRUCTED STAFF TO REMOVE THE MAGNESIUM FROM THE COMPO
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24316948 ↗MalfunctionJan 20, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THE WIRES OF THE POWER CORD OF AN EM2400, MAIN MODULE WERE EXPOSED. THE ISSUE WAS NOTICED DURING SETUP. POWER CORD WAS SWAPPED WITH BACKUP TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24123236 ↗MalfunctionJul 29, 2025Device Emits Odor; Device Emits Odor
IT WAS REPORTED THAT AN EM2400, MAIN MODULE HAD A BURNING SMELL. THIS OCCURRED DURING TROUBLESHOOTING PROCESS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22649626 ↗MalfunctionOct 31, 2024Component Missing; Component Missing
IT WAS REPORTED THAT ONE OF THE MAGNETS ON THE ROTOR DOOR OF AN EXACTAMIX COMPOUNDING DEVICE WAS MISSING. THE MISSING MAGNET ON THE ROTOR DOOR WAS DISCOVERED DURING PROGRAMMING/SETUP IN THE PHARMACY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20582002 ↗MalfunctionOct 31, 2024Contamination; Contamination
IT WAS REPORTED THAT WHILE USING AN EXACTAMIX PHARMACY COMPOUNDING DEVICE THE BAGS WERE CLOUDY. THIS WAS OBSERVED DURING DELIVERY IN A PHARMACY. TO RESOLVE THIS ISSUE, A VALVE SET AND A SECOND COMPOUNDER WERE SETUP WITH NO ISSUES NOTED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20580129 ↗MalfunctionSep 30, 2024Contamination; Contamination
IT WAS REPORTED THAT WHILE USING AN EXACTAMIX PHARMACY COMPOUNDING DEVICE THE BAGS WERE CLOUDY. THE CUSTOMER WAS TRYING RUN A ¿CLEAR¿ BAG BUT NOTICED THAT THE BAG LOOKED CLOUDY AFTER PUMPING AND THERE WERE LIPIDS ON PORT 1 AND 2. THIS ISSUE WAS RESOLVED AFTER CORRECTING THE CONFIGURATION AND RE-SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20334177 ↗MalfunctionSep 23, 2024Precipitate in Device or Device Ingredient; Precipitate in Device or Device Ingredient
IT WAS REPORTED THAT PRECIPITATE CONTAMINATION WAS OBSERVED ASSOCIATED WITH AN EXACTAMIX PHARMACY COMPOUNDING DEVICE. THE PRECIPITATE WAS OBSERVED BRIEFLY WHILE PRIMING AND VERIFYING. TO RESOLVE THIS ISSUE, RE-PRIMING WAS PERFORMED WITH NO FURTHER ISSUES. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20281191 ↗MalfunctionAug 1, 2023Break; Break
IT WAS REPORTED THAT THE MAIN MODULE DOOR OF AN EXACTAMIX AUTOMATED COMPOUNDING DEVICE (ACD) WAS DAMAGED. THE DAMAGED DOOR WAS FURTHER DESCRIBED AS, ¿IN FAULT STATE / DOOR THAT HOUSES ROTOR AND TUBING WON'T CLOSE. IS UNHINGED¿. THIS ISSUE WAS OBSERVED DURING AN UNSPECIFIED PROCESS STEP PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17436554 ↗
FDA clearance and classification
The FDA regulates product code NEP (system/device, pharmacy compounding) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Common questions about the EXACTAMIX EXACTA-M
Has the EXACTAMIX EXACTA-M been recalled?
No FDA recall record in our data names this exact device identifier (DI 00085412586557), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EXACTAMIX EXACTA-M?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Aug 1, 2023 and Sep 23, 2026. In those reports, the most common specific device problems are “Device Emits Odor” (3), “Contamination” (2) and “Excess Flow or Over-Infusion” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EXACTAMIX EXACTA-M?
None of the reports are classified as deaths, 2 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the EXACTAMIX EXACTA-M FDA approved?
Devices of this type (product code NEP) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Baxter Healthcare Corporation.
Who makes the EXACTAMIX EXACTA-M, and is it still sold?
The FDA's device database lists Baxter Healthcare Corporation as the company responsible for it (brand EXACTAMIX, model EXACTA-M). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System/device, pharmacy compounding devices
- EXACTAMIX H938175Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING3 recalls263 reports
- EXACTAMIX H938174Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING2 recalls235 reports
- B.BRAUN 2112550B. Braun Medical Inc. · SYSTEM/DEVICE, PHARMACY COMPOUNDING73 reports
- IVX StationOmnicell, Inc. · System/Device, Pharmacy Compounding1 recall0 reports
- ABACUS 83003393Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING1 recall0 reports
- ABACUS 83003392Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING1 recall0 reports
- EXACTAMIX EXM12DYBaxter Healthcare Corporation · System/device, pharmacy compounding1 recall0 reports
- EXACTAMIX EXM24DYBaxter Healthcare Corporation · System/device, pharmacy compounding1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page