OPTIS™ Integrated SystemRecalls & reported problems

Lightlab Imaging, Inc. · Imaging System · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00183739000937), received between Nov 6, 2018 and Oct 18, 2023 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem1
  • Connection Problem1
  • Erratic or Intermittent Display1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Signal Artifact/Noise1
  • Smoking1
  • Thermal Decomposition of Device1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • No Patient Involvement1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING THE PROCEDURE WHEN USING THE OPTIS INTEGRATED SYSTEM , AN UNSPECIFIED STENT WAS TO BE IMPLANTED. HOWEVER, THE RENDERING STENT SEEMS TO BE OFF 10MM. THE OUTCOME OF THE PROCEDURE IS UNKNOWN AT THIS TIME. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. FIELD SERVICE WILL BE PERFORMED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17959089 ↗
  • InjuryNov 16, 2021Communication or Transmission Problem; Communication or Transmission Problem

    DURING THE PROCEDURE, A ONE HOUR DELAY OCCURRED WHEN TWO CATHETERS COULD NOT BE CALIBRATED WITH THE SYSTEM. THE ERROR MESSAGE "PLEASE ADJUST THE CALIBRATION MANUALLY (CAL-002)." APPEARED. THE PROCEDURE WAS COMPLETED WITHOUT OCT AND THE PATIENT WAS STABLE. THE SYSTEM WAS REPAIRED TO RESOLVE THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12819311 ↗
  • InjuryOct 1, 2020Thermal Decomposition of Device; Thermal Decomposition of Device

    THE DOC HOLSTER CAME OFF OF TABLE CLAMP AND THE WASHERS INSIDE SHORTED OUT THE CIRCUIT CAUSING THE PLASTIC TO MELT. NO PATIENT WAS INVOLVED IN THIS EVENT AND NO ADVERSE CONSEQUENCES OCCURRED.

    Patient problems: No Patient Involvement

    MAUDE report 10616201 ↗
  • MalfunctionNov 15, 2019Smoking; Smoking

    AFTER A POWER SURGE IN THE HOSPITAL, AN AUDIBLE "BANG" OCCURRED AND SMOKE WAS NOTED TO BE COMING FROM THE SYSTEM. THERE WAS INTERMITTENT ANGIO CO-REGISTRATION FEED RESULTING IN GRAINY IMAGES BEING DISPLAYED. THE HOSPITAL EXPERIENCED A POWER SURGE AND THE SYSTEM BECAME UNABLE TO FUNCTION PROPERLY. THE FUSES WERE REPLACED AND THE KIT WAS ATTEMPTED TO BE BOOTED, BUT AN AUDIBLE "BANG" OCCURRED AND THE SYSTEM WOULD NO LONGER POWER UP. SMOKE WAS NOTED TO BE COMING FROM THE SYSTEM. THERE WAS NO PATIENT INVOLVED IN THIS EVENT. AFTER INVESTIGATION, THE POWER SUPPLY UNIT FUSES WERE FOUND TO BE BLOWN. THE FUSES WERE REPLACED AND THE ISSUE WAS RESOLVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9328659 ↗
  • InjuryApr 24, 2019Connection Problem; Connection Problem

    THE PROCEDURE WAS POSTPONED DUE TO A CONNECTION ISSUE THAT OCCURRED AFTER A CATHETER WAS DISCONNECTED WITHOUT USING THE EJECT BUTTON. ANOTHER CATHETER WAS USED BUT THE ISSUE PERSISTED. THE ISSUE WAS ISOLATED TO THE SYSTEM. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8546870 ↗
  • InjuryNov 6, 2018Signal Artifact/Noise; Erratic or Intermittent Display; Signal Artifact/Noise; Erratic or Intermittent Display

    A DRAGONFLY OPTIS CATHETER WAS ADVANCED TO THE LESION, BUT WHEN IN LIVE VIEW, THE VIBRATION FROM THE DEVICE WAS SIGNIFICANT. A "PINCH TEST" WAS PERFORMED WHICH SHOWED ARTIFACTS. WHEN A PULLBACK WAS PERFORMED, THE OPTIS INTEGRATED SYSTEM FAILED TO ACQUIRE AN IMAGE. AFTER TWO ATTEMPTS, A SECOND DRAGONFLY OPTIS CATHETER WAS PREPARED. AFTER A PULLBACK WAS PERFORMED WITH THIS DEVICE, THE SYSTEM WAS NOT ABLE TO ACQUIRE AN IMAGE EITHER. THE PROCEDURE WAS RESCHEDULED AND COMPLETED ON (B)(6) 2018 WITH NO ISSUES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8042682 ↗

Other SYSTEM, IMAGING, OPTICAL COHERENCE TOMOGRAPHY (OCT) devices

All SYSTEM, IMAGING, OPTICAL COHERENCE TOMOGRAPHY (OCT) devices →