PedFuse® 01-80101Recalls & reported problems
Spinefrontier, Inc. · PedFuse, Set Screw, 5-7mm · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00190361010586), received between May 17, 2019 and May 23, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K092420); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method2
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2019Mechanical Problem; Mechanical Problem
ON (B)(6) 2018, IT WAS REPORTED THAT THE PEDFUSE RESET LOCKING CAPS SHED A STRIP OF METAL WHILE TIGHTENING DURING SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8638695 ↗MalfunctionMay 17, 2019Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
SPINEFRONTIER SALES REPRESENTATIVE REPORTED THAT THE SURGEON ENCOUNTERED DIFFICULTY ADVANCING THE SET SCREW INTO THE TULIP DURING INITIAL PLACEMENT. THE SET SCREW GOT CROSS THREADED DURING THE PROCESS AND HAD TO BE REPLACED. THIS WAS THE SECOND SET SCREW CROSS THREADING OCCURRENCE DURING THIS CASE. THE FIRST OCCURRENCE WAS REPORTED UNDER MDR 3005977257-2019-00004.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8622112 ↗MalfunctionMay 17, 2019Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
SPINEFRONTIER SALES REPRESENTATIVE REPORTED THAT THE SURGEON ENCOUNTERED DIFICULTY ADVANCING THE SET SCREW INTO THE TULIP DURING INITIAL PLACEMENT. THE SET SCREW GOT CROSS THREADED DURING THE PROCESS AND HAD TO BE REPLACED. THIS WAS THE FIRST OF TWO SET SCREW CROSS THREADING OCCURRENCE DURING THIS CASE. THE SECOND OCCURRENCE WAS REPORTED UNDER MDR 3005977257-2019-00005.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8622107 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092420. One submission often covers many models, so check the product and lot details in each record.
- Z-1926-2013 · Mar 14, 2013 · TerminatedRev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instruments which capture, hold and drive pedicle screw assemblies. There is a capturing feature th
- Z-1737-2013 · Dec 30, 2011 · TerminatedKRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicl
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page