Invictus 17950-225Recalls & reported problems

Alphatec Spine, Inc. · MIS T27 Polyaxial Screwdriver, Locking · Class 1

All models of Invictus (96)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Alphatec Spine Inviscus polyaxial locking screwdriver is used to place screws during minimally invasive spine surgery. The 53 reports mostly describe the tip snapping off during a procedure, sometimes leaving a fragment in the patient that had to be retrieved. Five reports mention a foreign body left in the patient, and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 00190376473017), received between Feb 13, 2024 and Sep 24, 2026 — classified as 5 injuries, 49 malfunctions.

The most frequently reported specific device problem is “Break” (54 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 3 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 6 reportsOct 2024: 3 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break54
  • Material Separation23

Patient problems

  • No Clinical Signs, Symptoms or Conditions45
  • Foreign Body In Patient5
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Break; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892126 ↗
  • MalfunctionJun 29, 2026Break; Material Separation

    THE TIP OF THE SCREWDRIVER SHEARED OFF. THE BROKEN FRAGMENT WAS RETRIEVED WITHOUT INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25659264 ↗
  • MalfunctionJun 16, 2026Break

    DURING THE PROCEDURE, THE TIP OF THE SCREWDRIVER SHEARED OFF. IT WAS SUCCESSFULLY RETRIEVED AND DISCARDED BY THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25524093 ↗
  • MalfunctionJun 10, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF INTO A SCREW. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25464962 ↗
  • MalfunctionMay 27, 2026Break

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25304452 ↗
  • MalfunctionMay 20, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF WHEN PLACING A SCREW. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25233147 ↗
  • MalfunctionMay 6, 2026Break; Material Separation

    IT WAS REPORTED THAT THE DRIVER IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25105574 ↗
  • MalfunctionMar 6, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER HAS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24539491 ↗
  • MalfunctionFeb 20, 2026Break; Material Separation

    IT WAS REPORTED THAT THE DRIVER HAS A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24403000 ↗
  • MalfunctionFeb 18, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER IS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377864 ↗
  • MalfunctionFeb 3, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER SHEARED OFF DURING FINAL TIGHTENING OF A SCREW. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24250398 ↗
  • MalfunctionDec 30, 2025Break; Break

    IT WAS REPORTED THAT THE DRIVER BROKE DURING SCREW REMOVAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23930223 ↗
  • MalfunctionDec 30, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23926442 ↗
  • MalfunctionOct 22, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23358593 ↗
  • MalfunctionSep 3, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22958851 ↗
  • MalfunctionJul 31, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER HAS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22676594 ↗
  • MalfunctionJul 29, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER HAS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22650200 ↗
  • MalfunctionJul 7, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER SNAPPED OFF DURING A CASE. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22420475 ↗
  • MalfunctionMay 29, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22107173 ↗
  • MalfunctionMay 27, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER HAS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22092306 ↗
  • MalfunctionMay 9, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER SHEARED OFF DURING REMOVAL OF A SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21998779 ↗
  • MalfunctionFeb 20, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER HAS BROKEN OFF.

    Patient problems: Insufficient Information

    MAUDE report 21427550 ↗
  • MalfunctionNov 12, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF INTO A SCREW DURING A CASE. THE SCREW WAS ATTEMPTED TO BE REMOVED, AND THE TULIP HEAD BROKE OFF OF THE SCREW SHANK WITH THE DRIVER TIP INSIDE. BOTH THE TULIP AND DRIVER TIP WERE REMOVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20668548 ↗
  • MalfunctionNov 8, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP BROKE DURING SCREW INSERTION. THE TIP WAS REMOVED FROM THE PATIENT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20644368 ↗
  • MalfunctionNov 5, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF WHEN INSERTING THE SCREW INTO THE ILIUM. ALL BROKEN PIECES WERE REMOVED FROM THE PATIENT. NO FURTHER COMPLICATIONS WERE REPORTED. THIS IS REPORT 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20613048 ↗