SafeOp 3 AIX11240Recalls & reported problems

Alphatec Spine, Inc. · SafeOp 3 Patient Interface · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00190376517506), received between Jul 10, 2025 and May 27, 2026 — classified as 3 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 13, 2026Open, Classified

    Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.

    Z-2793-2026 · Root cause: Software Design Change · FDA record ↗

Most reported problems

Device problems

  • Unable to Obtain Readings6
  • Malposition of Device2
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Limb Pain1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2026Unable to Obtain Readings

    DURING A SPINAL PROCEDURE USING INTRAOPERATIVE NEUROMONITORING, THE SYSTEM WOULD NOT DISPLAY FREE RUN EMG WAVEFORMS. THE CASE PROCEEDED WITHOUT THE USE OF FREE RUN EMG.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25306989 ↗
  • MalfunctionMar 6, 2026Unable to Obtain Readings

    DURING A SPINAL PROCEDURE USING INTRAOPERATIVE NEUROMONITORING, THE LEFT HAND FREE-RUN EMG REPORTEDLY WOULD ONLY DISPLAY IN ONE VIEW AFTER ABOUT 30-40 MINUTES OF NORMAL ACTIVITY. THE MEPS RAN PROPERLY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531926 ↗
  • InjuryJan 27, 2026Malposition of Device

    A PATIENT UNDERWENT A SPINAL PROCEDURE ON (B)(6) 2025 IN WHICH INTRAOPERATIVE NEUROMONITORING WAS UTILIZED. DURING THE PROCEDURE TO4 TESTING WAS PASSING WITH 4/4, AND TEMG SCREW TESTING RESULTS SHOWED > 20MA. POSTOPERATIVE CT SCANS REVEALED A MEDIAL PEDICLE WALL BREACH OF THE S1 SCREW. THE PATIENT HAD REVISION SURGERY ON (B)(6) 2025 IN WHICH THE SCREW WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24193393 ↗
  • MalfunctionDec 15, 2025Unable to Obtain Readings; Unable to Obtain Readings

    DURING A SPINAL PROCEDURE, THE UNIT WAS NOT DISPLAYING SEMG ACTIVITY. THE TABLET WAS RESTARTED, WHICH RESOLVED THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23803902 ↗
  • MalfunctionDec 5, 2025Unable to Obtain Readings; Unable to Obtain Readings

    DURING A SPINAL PROCEDURE USING INTRAOPERATIVE NEUROMONITORING, FREE-RUN EMG REPORTEDLY FAILED TO DISPLAY. THE MEPS WERE STUCK IN A STIMULATING STATE AND WOULD NOT REGISTER. IN ADDITION, THE ELECTRODE TEST WOULD NOT REGISTER AFTER A REBOOT. THE PATIENT INTERFACE AND TABLE WERE RESTARTED TWICE TO RESOLVE THE ISSUE. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23733330 ↗
  • MalfunctionDec 5, 2025Unable to Obtain Readings; Unable to Obtain Readings

    DURING A SPINAL PROCEDURE, THE NEUROMONITORING SYSTEM WOULD NOT DISPLAY ANY FREE-RUN EMG ACTIVITY. THE SYSTEM WAS SWITCHED OUT TO CONTINUE THE CASE SUCCESSFULLY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23733255 ↗
  • InjuryJul 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Malposition of Device; Incorrect, Inadequate or Imprecise Result or Readings; Malposition of Device

    A PATIENT UNDERWENT A SPINAL PROCEDURE AT THE L4-L5 LEVEL, DURING WHICH INTRAOPERATIVE NEUROMONITORING WAS UTILIZED. POSTOPERATIVELY, THE PATIENT DEVELOPED NEW-ONSET LEFT LEG PAIN. RADIOGRAPHIC IMAGING REVEALED A MEDIAL BREACH OF THE LEFT L5 PEDICLE SCREW. FOUR MONTHS AFTER THE INITIAL SURGERY, REVISION SURGERY WAS PERFORMED TO REMOVE AND REPLACE THE HARDWARE.

    Patient problems: Limb Pain

    MAUDE report 22522090 ↗
  • InjuryJul 10, 2025Unable to Obtain Readings; Unable to Obtain Readings

    AT THE START OF A SPINAL PROCEDURE, THE NEUROMONITORING SYSTEM FAILED TO DISPLAY ANY FREE-RUN EMG ACTIVITY. MULTIPLE ATTEMPTS TO RESTART BOTH THE PATIENT INTERFACE AND THE TABLET WERE UNSUCCESSFUL, RESULTING IN THE PROCEDURE BEING ABORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22471935 ↗

Other Stimulator, electrical, evoked response devices

All Stimulator, electrical, evoked response devices →