Calibrate 420-PRO-07102500-SRecalls & reported problems
Alphatec Spine, Inc. · Calibrate NanoTec PSX, 7 x 10 x 25mm, 20° · Class 2 · Implant
All models of Calibrate (44)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00190376529714), received between Sep 11, 2025 and Dec 4, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Collapse” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Collapse1
- Fracture1
Patient problems
- Failure of Implant2
- Insufficient Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 4, 2025Fracture; Fracture
AN EXPANDABLE INTERBODY CAGE WAS IMPLANTED AND EXPANDED AS INTENDED DURING A SPINAL PROCEDURE PERFORMED ON AN UNKNOWN DATE. NO INTRAOPERATIVE COMPLICATIONS WERE REPORTED. ABOUT 2 MONTHS POSTOPERATIVELY, RADIOGRAPHIC IMAGING REVEALED THAT THE CAGE HAD FRACTURED. THE PATIENT IS ASYMPTOMATIC, AND THERE ARE NO PLANS FOR REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 23717007 ↗MalfunctionSep 11, 2025Collapse; Collapse
A PATIENT UNDERWENT A LUMBAR INTERVERTEBRAL BODY FUSION PROCEDURE ON (B)(6) 2024. AN EXPANDABLE INERBODY CAGE WAS IMPLANTED AND EXPANDED AS INTENDED. NO INTRAOPERATIVE COMPLICATIONS WERE REPORTED. AT THE 1-YEAR POSTOPERATIVE VISIT, RADIOGRAPHIC IMAGES REVEALED THAT THE CAGE HAD COLLAPSED.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 23031584 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page