DJO SURGICAL 804-06-311Recalls & reported problems

Encore Medical, L.P. · ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00190446843825), received between May 21, 2025 and May 21, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

7 other recalls cover devices cleared under the same FDA submission (K233481); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 18, 2025Class IIOpen, Classified

    Their is a potential that the reamer may kick or bind up during or immediately prior to use.

    Z-2546-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 21, 2025Mechanical Problem; Mechanical Problem

    COMPLAINT - DURING OUR CASE WITH DR, WE WERE COMPLETING A REVERSE AUGMENTED GLENOID WHEN THE MAIN OFF AXIS REAMER AND ITS COMPONENTS SEEMINGLY BOUND UP ON EACH OTHER WHEN ATTEMPTING TO REAM FOR THE WEDGE BASEPLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22066793 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K233481. One submission often covers many models, so check the product and lot details in each record.

  • Z-2548-2025 · Jul 18, 2025 · Open, Classified
    Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER SLEEVE Model/Catalog Number: 804-06-310 Product Description: Material: S.S/CoCrMo, Non-Sterile
  • Z-2549-2025 · Jul 18, 2025 · Open, Classified
    Brand Name: DJO SURGICAL Product Name: FA S Altivate Reverse Glenoid Tray Model/Catalog Number: None Software Version: NA Product Description: Enovis" shoulder devices are intended
  • Z-2547-2025 · Jul 18, 2025 · Open, Classified
    Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER BOSS DRILL, 6.5mm Model/Catalog Number: 804-06-312 Product Description: Material: S.S., Non Sterile
  • Z-2010-2025 · May 28, 2025 · Open, Classified
    Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 18mm Model/Catalog Number: 506-04-118 Software Version: NA Product Description: Encore Medical L.
  • Z-2009-2025 · May 28, 2025 · Open, Classified
    Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm Model/Catalog Number: 506-04-122 Software Version: NA Product Description: Encore Medical L.
  • Z-2011-2025 · May 28, 2025 · Open, Classified
    Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm Model/Catalog Number: 506-04-130 Software Version: NA Product Description: Encore Medical L.
  • Z-2008-2025 · May 28, 2025 · Open, Classified
    Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm Model/Catalog Number: 506-04-126 Software Version: NA Product Description: Encore Medical L.