Skip to content

Presidio Anterior Lumbar Fixation System 125-603Recalls & reported problems

Life Spine, Inc. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00190837043247), received between Jun 23, 2020 and Apr 1, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2026Fracture

    IT WAS STATED THAT, "BROKE DURING A CASE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24756115 ↗
  • MalfunctionMar 5, 2026Fracture

    IT WAS STATED, "IT BROKE IN THE VERTEBRAL BODY WHILE BEING INSERTED TO MAKE THE PILOT HOLE FOR THE SCREW. YES THE BROKEN PIECE WAS REMOVED BY HEAVY NEEDLE DRIVER"

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24519494 ↗
  • InjuryJun 23, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THA THE TIP OF THE AWL BROKE OFF DURING THE PROCEDURE BECAUSE OF HARD BONE AND A STEEP ANGLE. THE DOCTOR FELT IT WAS BEST TO LEAVE THE BROKEN PIECE IN THE THAT POSITION AND SAFELY BURIED COMPLETELY IN THE VERTEBRAL BONE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10186626 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page