Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Nov 29, 2023Class IIOpen, Classified

    During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

    Z-0826-2024 · Root cause: Process control · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code OKE (port introducer kit) under 21 CFR 880.5965 as a Class II device. 21 CFR 880.5965 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ACS UIPO82AB

Has the ACS UIPO82AB been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0826-2024 (Class II), began on Nov 29, 2023: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked"… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the ACS UIPO82AB, and is it still sold?

The FDA's device database lists American Contract Systems, Inc. as the company responsible for it (brand ACS, model UIPO82AB). Its record shows it as in commercial distribution.