Skip to content

ACS SMGY46ARecalls & reported problems

American Contract Systems, Inc. · GYN PACK · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 6, 2024Open, Classified

    Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility

    Z-2518-2024 · Root cause: Under Investigation by firm · FDA record ↗

  2. Nov 29, 2023Open, Classified

    During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

    Z-0828-2024 · Root cause: Process control · FDA record ↗