ACS PERP25AQRecalls & reported problems
American Contract Systems, Inc. · PEI RETINAL PACK - 206025 · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 4, 2025Open, Classified
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Z-0185-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
Other Eye tray devices
- Medline Industries, Inc. DYNJ47859CMedline Industries, Inc. · Eye tray5 recalls0 reports
- Medline Industries, Inc. DYNJ68875FMedline Industries, Inc. · Eye tray3 recalls1 report
- Medline Industries, Inc. DYNJ65808CMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ64572AMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ41207CMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ902256CMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ53019AMedline Industries, Inc. · Eye tray2 recalls1 report
- Medline Industries, Inc. DYKMBNDL92BMedline Industries, Inc. · Eye tray2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.