ACS MSKA45ARecalls & reported problems
American Contract Systems, Inc. · ARTHROSCOPY PACK · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Aug 7, 2024Open, Classified
ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Z-3229-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Other Orthopedic tray devices
- Medline Industries, Inc. DYNJ905330BMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ17452SMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ902473LMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ900718CMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. CDS780119VMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ21826RMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ906893DMedline Industries, Inc. · Orthopedic tray4 recalls2 reports
- Medline Industries, Inc. DYNJ909505AMedline Industries, Inc. · Orthopedic tray4 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page