ACS CMEY17AIRecalls & reported problems
American Contract Systems, Inc. · EYE PACK - 213187 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00191072211514), received between Apr 21, 2025 and Apr 21, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Contamination” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 4, 2025Class IIOpen, Classified
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Z-0185-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Contamination1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 21, 2025Contamination; Contamination
THE CUSTOMER STATES MULTIPLE PIECES OF BLACK MATERIAL, POSSIBLY INK, WAS COMING OFF LABELS ONTO SURGEON'S GLOVES DURING AN IMPLANT PROCEDURE. THE PRODUCT WAS USED DURING IOL IMPLANT PROCEDURE AND WAS NOT NOTICED UNTIL DEBRIS WAS SEEN IN THE EYE BY THE SURGEON. THE DEBRIS WAS REMOVED FROM PATIENT'S EYE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876493 ↗
Other Eye tray devices
- Medline Industries, Inc. DYNJ47859CMedline Industries, Inc. · Eye tray5 recalls0 reports
- Medline Industries, Inc. DYNJ68875FMedline Industries, Inc. · Eye tray3 recalls1 report
- Medline Industries, Inc. DYNJ65808CMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ902256CMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ64572AMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ41207CMedline Industries, Inc. · Eye tray3 recalls0 reports
- Medline Industries, Inc. DYNJ53019AMedline Industries, Inc. · Eye tray2 recalls1 report
- Medline Industries, Inc. DYNJ909065AMedline Industries, Inc. · Eye tray2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page