Skip to content

ACS UIVT47ASRecalls & reported problems

American Contract Systems, Inc. · VITRECTOMY SUPPLY (PS 907699) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Sep 4, 2025Open, Classified

    Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

    Z-0185-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Other Eye tray devices

All Eye tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page