Skip to content

Streamline MIS Spinal Fixation System 05-PA-65-50Recalls & reported problems

Xtant Medical Holdings, Inc. · CANNULATED SCREW, POLYAXIAL · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00191083019604), received between Feb 6, 2026 and Feb 6, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2026Detachment of Device or Device Component

    THE MANUFACTURER WAS CONTACTED ON 01/08/2026 TO REPORT AN ISSUE WITH A MIS SCREW. IT WAS REPORTED THAT DURING A POSTERIOR FUSION AT L3-L5, DURING FINAL TIGHTENING OF THE LAST SCREW IN A CONSTRUCT, THE TULIP OF THE SCREW DISSOCIATED FROM THE SHANK. SINCE THE OTHER SCREWS IN THE CONSTRUCT WERE FINAL TIGHTENED AND THE SCREW WITH THE DISSOCIATED TULIP WAS NOT AT EITHER END OF THE CONSTRUCT, THE DISSOCIATED TULIP AND SCREW SHANK WERE LEFT IN PLACE.LATERAL X-RAY IMAGES PROVIDED BY THE COMPLAINANT SHOWED 6 IMPLANTED SCREWS CONNECTED BY TWO RODS, ONE OF THE SCREWS HAD ITS TULIP DISSOCIATED FROM THE SHANK OF THE SCREW. THE DISSOCIATED TULIP OF THE SCREW REMAINED SEATED IN PLACE ON THE IMPLANTED ROD. THE SCREW WITH THE DISSOCIATED TULIP WAS THE MIDDLE SCREW IN A SERIES OF THREE. THE SHANK OF THE SCREW WITH THE DISSOCIATED TULIP APPEARED TO BE SEATED FURTHER INTO THE PATIENT'S BONE THAN THE OTHER SCREWS USED FOR FIXATION. ADDITIONALLY, UNKNOWN INTERVERTEBRAL IMPLANTS APPEAR IN BETWEEN THE LEVELS OF FIXATION .

    Patient problems: Failure of Implant

    MAUDE report 24283672 ↗

Other Orthosis, spinal pedicle fixation devices

All Orthosis, spinal pedicle fixation devices →