Skip to content

Agile™ Esophageal M00517710Recalls & reported problems

Boston Scientific Corporation · Partially Covered OTW Stent System · Class 2 · Implant

All models of Agile Esophageal (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00191506006587), received between Feb 1, 2023 and Feb 1, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 1, 2024Class IIOpen, Classified

    WallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.

    Z-3287-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 1, 2023Difficult to Remove; Difficult to Remove

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2023, THAT AN AGILE ESOPHAGEAL OTW PARTIALLY COVERED STENT WAS TO BE IMPLANTED IN THE ESOPHAGUS TO TREAT AN ESOPHAGEAL CANCER DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS NOT TORTUOUS. DURING THE PROCEDURE, THE AGILE ESOPHAGEAL STENT WAS DEPLOYED THROUGH A PREVIOUSLY IMPLANTED NON- BOSTON SCIENTIFIC STENT. DURING REMOVAL OF THE DELIVERY SYSTEM, THE AGILE ESOPHAGEAL STENT CAME OFF TOGETHER WITH THE DELIVERY CATHETER. THE PROCEDURE WAS COMPLETED WITH A 23MM X 10CM AGILE ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16277832 ↗

Other PROSTHESIS, ESOPHAGEAL devices

All PROSTHESIS, ESOPHAGEAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page