Skip to content

Agile™ Esophageal M00517720Recalls & reported problems

Boston Scientific Corporation · Partially Covered OTW Stent System · Class 2 · Implant

All models of Agile Esophageal (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00191506006594), received between Jun 18, 2024 and Dec 31, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 1, 2024Class IIOpen, Classified

    WallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.

    Z-3288-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Remove1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 31, 2025Positioning Problem; Positioning Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL PARTIALLY COVERED STENT WAS TO BE IMPLANTED IN THE ESOPHAGUS TO TREAT ESOPHAGEAL CANCER DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON DECEMBER 1, 2025. THE PATIENT'S ANATOMY WAS TORTUOUS (TIGHT) AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PHYSICIAN WAS NOT PLEASED WITH THE PLACEMENT OF THE 10CM AGILE ESOPHAGEAL STENT AS IT "JUMPED" FORWARD UPON PLACEMENT AND WAS NOT ABLE TO REPOSITION CORRECTLY. THE STENT WAS THEN REMOVED WITH RAT TOOTH FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER STENT OF THE SAME MODEL BUT WITH A LONGER SIZE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23935704 ↗
  • MalfunctionJun 18, 2024Difficult to Remove; Difficult to Remove

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL OTW PARTIALLY COVERED STENT WAS TO BE IMPLANTED TO TREAT ESOPHAGEAL CANCER DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, AFTER THE GUIDEWIRE WAS INSERTED ACROSS THE STRICTURE AND A PAPERCLIP PLACED ON THE HALFWAY POINT, THE STENT WAS SENT DOWN OVER THE GUIDEWIRE. THE STENT WAS THEN ABLE TO BE DEPLOYED; HOWEVER, UPON REMOVING THE DELIVERY SYSTEM, THE STENT ALONG WITH THE DELIVERY SYSTEM CAME OUT OF THE PATIENT'S MOUTH. THE PROCEDURE WAS COMPLETED WITH ANOTHER AGILE ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19556837 ↗

Other PROSTHESIS, ESOPHAGEAL devices

All PROSTHESIS, ESOPHAGEAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page