Skip to content

Vercise™ DB-9208-15Recalls & reported problems

Boston Scientific Neuromodulation Corporation · S8 Adapter 15cm · Class 3 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00191506014339), received between Feb 18, 2026 and Feb 18, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “High impedance” (1 report).

4 other recalls cover devices cleared under the same FDA submission (P150031); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • High impedance1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 18, 2026High impedance; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED INADEQUATE STIMULATION WHICH RESULTED IN A RETURN OF THEIR PARKINSON'S DISEASE SYMPTOMS BRADYKINESIA AND RIGIDITY. IT WAS DISCOVERED THAT THERE WERE HIGH IMPEDANCE READINGS ON THE NON-BOSTON SCIENTIFIC RIGHT-SIDED LEAD EXTENSION. THE SUBPECTORAL IMPLANTABLE PULSE GENERATOR (IPG) WAS PALPATED BY THE PHYSICIAN TO TROUBLESHOOT THE ISSUE HOWEVER, THIS DID NOT RESOLVE THE IMPEDANCE ISSUE. X-RAY IMAGING WAS TAKEN WHICH DID NOT REVEAL ANY APPARENT ANOMALIES. THEREFORE, THE PATIENT UNDERWENT A REVISION PROCEDURE DURING WHICH ONE OF THE BSC ADAPTERS WAS EXPLANTED AND REPLACED WHICH DID NOT RESOLVE THE IMPEDANCE ISSUE. THE PHYSICIAN THEN DECIDED TO REPLACE THE NON-BSC LEAD EXTENSION AFTER WHICH THE IMPEDANCE ISSUE WAS THEN RESOLVED AND THE PATIENT HAS FULLY RECOVERED POSTOPERATIVELY. THE EXPLANTED ADAPTER WILL NOT BE RETURNED DUE TO HOSPITAL POLICY.

    Patient problems: Insufficient Information

    MAUDE report 24384403 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P150031. One submission often covers many models, so check the product and lot details in each record.

  • Z-2340-2025 · Jul 8, 2025 · Open, Classified
    Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
  • Z-2558-2025 · Jul 8, 2025 · Open, Classified
    Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
  • Z-1890-2024 · Apr 18, 2024 · Open, Classified
    Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
  • Z-2125-2020 · Apr 24, 2019 · Terminated
    Fifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the

Other Stimulator, electrical, implanted, for parkinsonian symptoms devices

All Stimulator, electrical, implanted, for parkinsonian symptoms devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.