Vercise™ DB-9208-15Recalls & reported problems
Boston Scientific Neuromodulation Corporation · S8 Adapter 15cm · Class 3 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00191506014339), received between Feb 18, 2026 and Feb 18, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “High impedance” (1 report).
4 other recalls cover devices cleared under the same FDA submission (P150031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- High impedance1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 18, 2026High impedance; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED INADEQUATE STIMULATION WHICH RESULTED IN A RETURN OF THEIR PARKINSON'S DISEASE SYMPTOMS BRADYKINESIA AND RIGIDITY. IT WAS DISCOVERED THAT THERE WERE HIGH IMPEDANCE READINGS ON THE NON-BOSTON SCIENTIFIC RIGHT-SIDED LEAD EXTENSION. THE SUBPECTORAL IMPLANTABLE PULSE GENERATOR (IPG) WAS PALPATED BY THE PHYSICIAN TO TROUBLESHOOT THE ISSUE HOWEVER, THIS DID NOT RESOLVE THE IMPEDANCE ISSUE. X-RAY IMAGING WAS TAKEN WHICH DID NOT REVEAL ANY APPARENT ANOMALIES. THEREFORE, THE PATIENT UNDERWENT A REVISION PROCEDURE DURING WHICH ONE OF THE BSC ADAPTERS WAS EXPLANTED AND REPLACED WHICH DID NOT RESOLVE THE IMPEDANCE ISSUE. THE PHYSICIAN THEN DECIDED TO REPLACE THE NON-BSC LEAD EXTENSION AFTER WHICH THE IMPEDANCE ISSUE WAS THEN RESOLVED AND THE PATIENT HAS FULLY RECOVERED POSTOPERATIVELY. THE EXPLANTED ADAPTER WILL NOT BE RETURNED DUE TO HOSPITAL POLICY.
Patient problems: Insufficient Information
MAUDE report 24384403 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P150031. One submission often covers many models, so check the product and lot details in each record.
- Z-2340-2025 · Jul 8, 2025 · Open, ClassifiedBurr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
- Z-2558-2025 · Jul 8, 2025 · Open, ClassifiedVercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
- Z-1890-2024 · Apr 18, 2024 · Open, ClassifiedVercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
- Z-2125-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the
Other Stimulator, electrical, implanted, for parkinsonian symptoms devices
- ACTIVA RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS2,839 reports
- Activa® 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS1,325 reports
- Percept™ RC B35300Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 recall3 deaths796 reports
- Vercise™ Cartesia™ DB-2202-45Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS18 deaths718 reports
- Activa® PC 37601Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS3 deaths521 reports
- Activa® RC 37612Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 death469 reports
- Activa® RC 37612Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms410 reports
- NA NM-3138-55Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS198 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.