ACUVUE Abiliti Overnight Therapeutic lenses for myopia managementRecalls & reported problems
Menicon CO.,Ltd. · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00192538951357), received between Jul 14, 2023 and Mar 19, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Use of Device Problem1
Patient problems
- Corneal Edema2
- Corneal Infiltrates2
- Corneal Scar2
- Eye Pain2
- Corneal Ulcer1
- Eye Irritation1
- Photophobia1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 27FEB2025, JOHNSON AND JOHNSON VISION CARE (JJVC) EMPLOYEE SENT AN EMAIL TO THE ACUVUE ABILITI¿ OVERNIGHT BRAND CONTACT LENS (CL) MANUFACTURING SITE (MENICON B.V.) INFORMING THEM OF THE FOLLOWING ADVERSE EVENT. MENICON B.V. RECORDED IT AS COMPLAINT ID: (B)(4) AND INFORMED THE MANUFACTURER (MENICON CO.). ADVERSE EVENT INFORMATION: ON 13FEB2025, AN EMAIL SENT BY A JOHNSON AND JOHNSON EMPLOYEE ADVISED A PATIENT (PT) WAS DIAGNOSED WITH A "POSSIBLE CORNEAL INFECTION" WITH A RESULTING SCAR THAT DOES NOT IMPACT THE VISUAL ACUITY (VA) WHILE WEARING AN ACUVUE ABILITI¿ OVERNIGHT BRAND CL. ON (B)(6) 2025, ADDITIONAL INFORMATION WAS RECEIVED. THE TREATING EYE CARE PROFESSIONAL (ECP) REPORTED THE PT EXPRESSED SYMPTOMS OF "UNCOMFORTABLE AND IRRITATED" WHILE WEARING THE LEFT EYE (OS) CL ON (B)(6) 2024. THE PT WAS SEEN ON (B)(6) 2024 WITH FINDINGS OF CENTRAL CORNEAL EDEMA AND LIGHT SENSITIVITY. THE PT WAS PRESCRIBED LOTEMAX FOUR TIMES DAILY (QID). THE PT WAS SEEN BY THE ECP FOR ADDITIONAL FOLLOW-UP (FU) VISITS ON (B)(6) 2024. THE ECP REPORTS THAT THE PT WAS ALSO SEEN ON (B)(6) 2025 AND IS 20/20 WITH OU CORRECTION. ON 28FEB2025, WE OBTAINED ADDITIONAL INFORMATION FROM JJVC. THE TREATING ECP REPO
Patient problems: Corneal Edema; Corneal Infiltrates; Corneal Scar; Eye Irritation; Eye Pain; Photophobia
MAUDE report 21639997 ↗InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 28FEB2025, JOHNSON AND JOHNSON VISION CARE (JJVC) EMPLOYEE SENT AN EMAIL TO THE ACUVUE ABILITI¿ OVERNIGHT BRAND CONTACT LENS (CL) MANUFACTURING SITE (MENICON B.V.) INFORMING THEM OF THE FOLLOWING ADVERSE EVENT. MENICON B.V. RECORDED IT AS COMPLAINT ID (B)(4) AND INFORMED THE MANUFACTURER (MENICON CO.). ADVERSE EVENT INFORMATION: ON 13FEB2025, AN EMAIL SENT BY A JOHNSON AND JOHNSON EMPLOYEE ADVISED A PATIENT (PT) WAS DIAGNOSED WITH A "POSSIBLE CORNEAL INFECTION" WITH A RESULTING SCAR THAT DOES NOT IMPACT THE VISUAL ACUITY (VA) WHILE WEARING AN ACUVUE ABILITI¿ OVERNIGHT BRAND CL. ON 26FEB2025, ADDITIONAL INFORMATION WAS RECEIVED. THE TREATING EYE CARE PROFESSIONAL (ECP) REPORTED THE PT WAS IN THE OFFICE ON (B)(6) 2025, ASYMPTOMATIC, AND WAS PRESCRIBED LOTEMAX FOUR TIMES DAILY (QID). THE PT WAS SEEN BY THE ECP FOR ADDITIONAL FOLLOW-UP (FU) VISITS ON (B)(6) 2025 AND (B)(6) 2025 WITH "SLOW IMPROVEMENTS OF EDEMA". ON 28FEB2025, ADDITIONAL INFORMATION WAS PROVIDED: THE TREATING ECP REPORTED THAT THE PT EXPERIENCED A "CORNEAL INFILTRATE LIKE AN IMMUNE RESPONSE WITH WHITE CELLS, BUT NOT INFECTIOUS AS PT WAS TREATED WITH LOTEMAX QID FOR 1 WEEK AND NEVER USED ANTIBIOTICS TO TREAT THE PT."
Patient problems: Corneal Edema; Corneal Infiltrates; Corneal Scar
MAUDE report 21639991 ↗InjuryJul 14, 2023Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Adverse Event Without Identified Device or Use Problem
ON 10JUL2023, THE SUSPECT PRODUCT WAS RECEIVED BY JOHNSON AND JOHNSON VISION CARE, INC., AND RETURNED TO MENICON FOR EVALUATION ON 13JUL2023. ON 14JUL2023, MENICON CONDUCTED THE EVALUATION. DETAILS OF THE EVALUATION RESULTS ARE SHOWN IN B6. THE CONCLUSION OF THE EVALUATION WAS THAT NO DEVIATIONS WERE DETECTED DURING THE SPECIFICATION AND GENERAL INSPECTIONS. AS A RESULT OF THE CORNEA TREATMENT, THE PATIENT (PT) RECOVERED WITH LITTLE SCARRING AND RETURNED TO WEARING CONTACT LENSES. PROPER HANDLING AND REGULAR EYE EXAMINATIONS ARE DESCRIBED IN THE INSTRUCTIONS FOR PATIENTS. UNCLEAR HOW THE PT WAS GUIDED BY THE EYE CARE PROFESSIONAL.
Patient problems: Corneal Ulcer; Eye Pain
MAUDE report 17325409 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page