MICROPORT INLNBRHNRecalls & reported problems
Microport Orthopedics Inc. · INLINE BROACH HANDLE INSTRUMENT · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00192629140622), received between Jan 15, 2026 and May 22, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Remove1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2026Break; Difficult to Remove
ALLEGEDLY, THE TIP OF THE BROACH HANDLE THAT ENGAGES THE BROACH WAS BROKEN OFF. NOT EXACTLY SURE WHEN THAT HAPPENED. WE ONLY LEARNED THAT IT WAS BROKEN OFF WHEN THE SURGEON WENT TO WITHDRAW THE BROACH FROM THE FEMUR. WHEN HE WOULD STRIKE THE BROACH HANDLE WITH THE MALLET IT WOULD COME OUT OF THE BROACH, LEAVING THE BROACH IN THE FEMUR. WE HAD TO USE A COKER TO PULL THE BROACH OUT. WE DID TAKE AN EX-RAY OF THE HIP AND FOUND NO DEBRIS IN THE WOUND.
Patient problems: Insufficient Information
MAUDE report 25257841 ↗InjuryJan 15, 2026Break; Adverse Event Without Identified Device or Use Problem
ALLEGEDLY, THE HANDLE WAS REPORTED AS SOMEWHAT LOOSE WHEN ATTACHING TO EACH RASP, BUT IT WAS NOT DETACHING WITH EVERY SEQUENTIAL RASP. ACCORDING TO THE DOCTOR, PREVIOUS CASES HAVE SHOWN RASPS COMING LOOSE APPROXIMATELY EVERY 50 PROCEDURES. DESPITE BEING LOOSE, THE HANDLE WAS FUNCTIONING AND DID NOT SLIP OFF THE RASP WHEN TRYING TO BACK IT OUT. WHEN THE SIZE 2 RASP WAS IMPACTED OUT TO PROCEED TO THE NEXT RASP, THE TIP OF THE INLNBRHN BROKE OFF. THE BROKEN TIP COULD NOT BE REMOVED BECAUSE IT WAS LODGED IN THE STEM POCKET. IN PREVIOUS SITUATIONS, THE BROKEN TIP HAS COME OUT OF THE POCKET AND A REPLACEMENT HANDLE WAS USED FOR REMOVAL. A LOCKING PLIERS-TYPE CLAMP WAS APPLIED TO GRIP THE RASP POCKET AT THE EDGE AND TAP IT OUT. AS A RESULT, THE SURGERY WAS EXTENDED BY MORE THAN 30 MINUTES. THIS ISSUE OCCURRED DURING THE SAME SURGICAL CASE AS THE INCIDENT GROUP (B)(4), SO THE TIME EXTENSION INCLUDES BOTH EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24087260 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.