Skip to content

GUARDIAN G1-503WCPX1Recalls & reported problems

Medline Industries, Inc. · SHOWER COMMODE WC, ALUMINUM, 275 LB · Class 1

All models of Guardian (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00193489028754), received between Aug 11, 2026 and Sep 25, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Break1
  • Corroded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Fracture

    It was reported that "Customer has a shower chair and one of the front wheels the bearings are seized up and because of that the wheels will not turn."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898766 ↗
  • MalfunctionSep 23, 2026Corroded

    It was reported that "both front wheel rusting" and "heavy rust in the bearings of the front casters. Casters will no longer swivel."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873351 ↗
  • MalfunctionAug 11, 2026Break

    It was reported that the two front wheel bearings got rusted and it fell off and the wheels are broken.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50213287 ↗
  • MalfunctionAug 11, 2026Fracture

    It was reported that the back of the chair has detached from the frame and is broken in half.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50201808 ↗
  • MalfunctionAug 11, 2026Fracture

    It was reported that "I purchased a Medline aluminum shower wheelchair from Amazon on (B)(6) 2026, about five months later the two front wheel bearings got rusted and it fell off and the wheels are broken."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50201468 ↗

Other CHAIR, WITH CASTERS devices

All CHAIR, WITH CASTERS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page