AVANOS 13557Recalls & reported problems
Avanos Medical, Inc. · HOMEPUMP C-SERIES* Elastomeric Pump, 60 ml, 2 ml/hr · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00193494135577), received between Dec 15, 2020 and Oct 7, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (4 reports).
3 other recalls cover devices cleared under the same FDA submission (K052117); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Excess Flow or Over-Infusion4
Patient problems
- No Known Impact Or Consequence To Patient3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 7, 2022Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
FILL VOLUME: 48 ML. FLOW RATE: 2ML/H. PROCEDURE: UNKNOWN. CATHPLACE: UNKNOWN. IT WAS REPORTED "THE ADMINISTRATION OF THE INFUSION WAS PERFORMED SIX HOURS EARLIER THAN SCHEDULED." THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 19-SEP-2022 STATED "THE INFUSER WAS PREPARED AND APPARENTLY THEY DID NOT FOLLOW THE CAPACITY INSTRUCTIONS (THEY GAUGED AT 48 ML, AS IF THE INFUSION RATE WERE LINEAR); INFORMATION INDICATES THAT FOR A 60ML INFUSER, IF YOU WANT IT TO BE INFUSED IN 24 HOURS, IT MUST BE CALIBRATED AT 51ML."
Patient problems: Insufficient Information
MAUDE report 15567695 ↗MalfunctionDec 15, 2020Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING AN UNKNOWN NUMBER OF PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 2026095-2020-00182 FOR THE FIRST REPORT ; REFER TO 2026095-2020-00183 FOR THE SECOND REPORT. IT WAS REPORTED THAT, "SEVERAL PUMPS EMPTIED TOO QUICKLY."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11010746 ↗MalfunctionDec 15, 2020Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING AN UNKNOWN NUMBER OF PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 2026095-2020-00182 FOR THE FIRST REPORT ; REFER TO 2026095-2020-00184 FOR THE THIRD REPORT . IT WAS REPORTED THAT ,"SEVERAL PUMPS EMPTIED TOO QUICKLY."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11010675 ↗MalfunctionDec 15, 2020Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING AN UNKNOWN NUMBER OF PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 2026095-2020-00183 FOR THE SECOND REPORT; REFER TO 2026095-2020-00184 FOR THE THIRD REPORT. IT WAS REPORTED THAT, "SEVERAL PUMPS EMPTIED TOO QUICKLY."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11010653 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K052117. One submission often covers many models, so check the product and lot details in each record.
- Z-2111-2009 · Feb 11, 2008 · TerminatedI-Flow Corporation Easy Pump LT 60-24 (60 ml x 2 ml/hr), REF: 04434366, assembled in Mexico. The Easypump LT is indicated for continuous and/or intermittent infusion of medicatio
- Z-0585-06 · Jan 17, 2006 · TerminatedEasypump L T 60-24 infusion pump
- Z-0584-06 · Jan 17, 2006 · TerminatedEasypump ST 100-1 infusion pump
Other Pump, infusion, elastomeric devices
- EASYPUMP 4540018-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC313 reports
- Halyard 13568Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC86 reports
- ON-Q* Pain Relief System with SELECT-A-FLOW 13472Avanos Medical, Inc. · Pump, infusion, elastomeric73 reports
- Halyard 13472Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC68 reports
- EASYPUMP 4540008-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC54 reports
- DOSI-FUSER L25915-500D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- DOSI-FUSER L25915-600D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- Halyard 13561Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC42 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page