AVANOS 13560Recalls & reported problems
Avanos Medical, Inc. · HOMEPUMP C-SERIES*, 100 ml, 0.5 ml · Class 2
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00193494135607), received between Jun 7, 2022 and Oct 24, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (6 reports).
3 other recalls cover devices cleared under the same FDA submission (K052117); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Excess Flow or Over-Infusion6
Patient problems
- Insufficient Information5
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2024Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
FILL VOLUME: 100ML. FLOW RATE: 0.5 ML/HR. PROCEDURE: CHEMOTHERAPY. CATHPLACE: PORT CATHETER. DEVICE START DATE: (B)(6) 2024. DEVICE END DATE: (B)(6) 2024. IT WAS REPORTED, "THE PUMP EMPTIED ON DAY-3 INSTEAD OF SUPPOSED DURATION (5 DAYS)." THE PUMP AND TUBING WERE UNDER THE PATIENT'S CLOTHING, BLANKET AND THE FLOW RESTRICTOR WAS TAPED TO THE PATIENT'S SKIN. TUBING UNDER CLOTHING/BLANKET. THE DEVICE WAS FILLED USING A SYRINGE ON (B)(6) 2024. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20523749 ↗MalfunctionJun 7, 2022Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE FIFTH OF FIVE REPORTS. REFER TO 2026095-2022-00057 FOR THE FIRST EVENT. REFER TO 2026095-2022-00058 FOR THE SECOND EVENT. REFER TO 2026095-2022-00059 FOR THE THIRD EVENT. REFER TO 2026095-2022-00061 FOR THE FOURTH EVENT. FILL VOLUME: 84 ML, FLOW RATE: 0.5 ML/HR, PROCEDURE: UNKNOWN, CATHPLACE: UNKNOWN. IT WAS REPORTED THE PUMP DID NOT LAST THE EXPECTED 7-DAYS. THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 14628100 ↗MalfunctionJun 7, 2022Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE THIRD OF FIVE REPORTS. REFER TO 2026095-2022-00057 FOR THE FIRST EVENT. REFER TO 2026095-2022-00058 FOR THE SECOND EVENT. REFER TO 2026095-2022-00061 FOR THE FIFTH EVENT. FILL VOLUME: 84 ML. FLOW RATE: 0.5 ML/HR. PROCEDURE: UNKNOWN. CATHPLACE: UNKNOWN. IT WAS REPORTED THE PUMP DID NOT LAST THE EXPECTED 7-DAYS. THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 14628073 ↗MalfunctionJun 7, 2022Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE FOURTH OF FIVE REPORTS. REFER TO 2026095-2022-00057 FOR THE FIRST EVENT. REFER TO 2026095-2022-00058 FOR THE SECOND EVENT. REFER TO 2026095-2022-00059 FOR THE THIRD EVENT. FILL VOLUME: 84 ML, FLOW RATE: 0.5 ML/HR, PROCEDURE: UNKNOWN, CATHPLACE: UNKNOWN. IT WAS REPORTED THE PUMP DID NOT LAST THE EXPECTED 7-DAYS. THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 14628072 ↗MalfunctionJun 7, 2022Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE SECOND OF FIVE REPORTS. REFER TO 2026095-2022-00057 FOR THE FIRST EVENT. REFER TO 2026095-2022-00060 FOR THE FOURTH EVENT. REFER TO 2026095-2022-00061 FOR THE FIFTH EVENT. FILL VOLUME: 84 ML, FLOW RATE: 0.5 ML/HR, PROCEDURE: UNKNOWN, CATHPLACE: UNKNOWN. IT WAS REPORTED THE PUMP DID NOT LAST THE EXPECTED 7-DAYS. THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 14628070 ↗MalfunctionJun 7, 2022Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE FIRST OF FIVE REPORTS. REFER TO 2026095-2022-00059 FOR THE THIRD EVENT. REFER TO 2026095-2022-00060 FOR THE FOURTH EVENT. REFER TO 2026095-2022-00061 FOR THE FIFTH EVENT. FILL VOLUME: 84 ML, FLOW RATE: 0.5 ML/HR, PROCEDURE: UNKNOWN, CATHPLACE: UNKNOWN. IT WAS REPORTED THE PUMP DID NOT LAST THE EXPECTED 7-DAYS. THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 14628068 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K052117. One submission often covers many models, so check the product and lot details in each record.
- Z-2111-2009 · Feb 11, 2008 · TerminatedI-Flow Corporation Easy Pump LT 60-24 (60 ml x 2 ml/hr), REF: 04434366, assembled in Mexico. The Easypump LT is indicated for continuous and/or intermittent infusion of medicatio
- Z-0584-06 · Jan 17, 2006 · TerminatedEasypump ST 100-1 infusion pump
- Z-0585-06 · Jan 17, 2006 · TerminatedEasypump L T 60-24 infusion pump
Other Pump, infusion, elastomeric devices
- EASYPUMP 4540018-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC313 reports
- Halyard 13568Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC86 reports
- ON-Q* Pain Relief System with SELECT-A-FLOW 13472Avanos Medical, Inc. · Pump, infusion, elastomeric73 reports
- Halyard 13472Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC68 reports
- EASYPUMP 4540008-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC54 reports
- DOSI-FUSER L25915-500D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- DOSI-FUSER L25915-600D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- Halyard 13561Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC42 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.