The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00193494316143), received between Dec 8, 2020 and Aug 14, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K063530); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Excess Flow or Over-Infusion1
  • Infusion or Flow Problem1

Patient problems

  • No Consequences Or Impact To Patient1
  • Pain1
  • Swelling/ Edema1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2024Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    FILL VOLUME: 400ML. FLOW RATE: 7ML/HR. PROCEDURE: ABDOMINOPLASTY. CATHPLACE: ABDOMEN. START TIME: 12:00PM, 17-JUL-2024. STOP TIME: 8:00PM, 17-JUL-2024. IT WAS REPORTED, "PUMP WAS PLACED IN AM WITH FILL VOLUME OF 400 ML, SET AT 7ML/HR, AND THE PUMP INFUSED COMPLETELY BY 8PM." ADDITIONAL INFORMATION RECEIVED 24-JUL-2024 STATED THE PUMP WAS ABOVE THE CATHETER, 6 INCHES (DISTANCE BETWEEN THE PUMP AND CATHETER). THE TUBING WAS UNDER THE PATIENT'S CLOTHING AND/OR BLANKET. THERE WERE NO AIR BUBBLES NOTED AND THE ENVIRONMENTAL ROOM TEMPERATURE WAS 72-DEGREE FAHRENHEIT. THE PATIENT DESCRIBED AN INTENSE BURNING SENSATION IN ABDOMEN FOLLOWED BY NUMBNESS. THE DEVICE WAS USED IN THE PATIENT'S HOME SETTING. PATIENT NOTED EXCESSIVE SWELLING/ LEAKAGE AND REQUESTED AN APPOINTMENT THE FOLLOWING DAY. IT WAS NOTED THERE WERE 30-MINUTES BETWEEN FILLING AND USE OF THE DEVICE. THE PATIENT RECEIVED WRITTEN AND ORAL INSTRUCTIONS ON USE OF THE DEVICE. THERE WERE NO CHANGES IN THE FLOW RATE DURING THERAPY.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 19986129 ↗
  • MalfunctionDec 8, 2020Infusion or Flow Problem; Infusion or Flow Problem

    FILL VOLUME: 400 ML. FLOW RATE: 2 ML/HR. PROCEDURE: BILATERAL MASTECTOMY. CATHPLACE: UNKNOWN. INFUSION START TIME: (B)(6) 2020. INFUSION STOP TIME: UNKNOWN. FDA MEDWATCH / FDA USER FACILITY REPORT # MW5097398 RECEIVED ON 16-NOV-2020, AND THE FOLLOWING INFORMATION WAS PROVIDED: "PATIENT UNDERWENT BILATERAL MASTECTOMY ON (DATE REDACTED) 2020. TWO ON-Q PUMPS PLACED, ONE WHICH MALFUNCTIONED AND INFUSED ALL OF THE DRUG OVER A SHORTER PERIOD OF TIME THAN INTENDED. IT WAS A 550ML PUMP FILLED WITH 400ML ROPIVACAINE 0.5%, DUAL CATHETER, FLOW RATE 2ML/HR THROUGH EACH CATHETER. PUMP REMOVED FROM PATIENT ON UNKNOWN DATE." NO PATIENT INJURY OR ADVERSE EFFECTS WERE NOTED. ADDITIONAL INFORMATION RECEIVED 17-NOV-2020 INDICATED THE PROCEDURE DATE AND START DATE OF THE INFUSION WAS (B)(6) 2020. THE PUMP WAS FILLED TO 400ML. ADDITIONAL INFORMATION RECEIVED 18-NOV-2020 INDICATED THE EVENT OCCURRED AFTER THE PATIENT LEFT THE HOSPITAL AND NO FURTHER INFORMATION WAS AVAILABLE IN THE HOSPITAL RECORD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10967971 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K063530. One submission often covers many models, so check the product and lot details in each record.

  • Z-2860-2011 · Apr 15, 2011 · Terminated
    I-Flow ON-Q with Select-A-Flow Pumps, Intended for continuous and/or intermittent delivery of medication to and around surgical wound sites and/or close proximity to nerves for p
  • Z-2133-2008 · Feb 14, 2008 · Terminated
    On-Q C-bloc 400 ml, 1-7 ml/hr Select-A-Flow + 5 ml/60 min ONDEMAND, REF CB005 indicated for continuous and/or intermittent delivery of medication (such as local anesthetics or narc
  • Z-1869-2008 · Feb 12, 2008 · Terminated
    ON-Q PainBuster Infusion Pump (270 ml, 5 ml/hr), REF PS6504, Part No. 5001260. The pump is indicated for continuous and/or intermittent delivery of medication (such as local anest