Skip to content

CARESCAPE Canvas Smart Display 5897060Recalls & reported problems

GE Healthcare Finland Oy · CARESCAPE Canvas Smart Display · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K223531); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Apr 5, 2024Open, Classified

    GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.

    Z-2170-2024 · Root cause: Software design · Recall details · FDA record ↗

  2. Apr 5, 2024Open, Classified

    GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.

    Z-2169-2024 · Root cause: Software design · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K223531. One submission often covers many models, so check the product and lot details in each record.

  • Z-2168-2024 · Apr 5, 2024 · Open, Classified
    CARESCAPE Canvas 1000, Model Numbers: a) 5865770; b) 5865770-01040169; c) 5865770-01054142; d) 5865770-01054696; e) 5865770-01054698; f) 5865770-01054721; g) 5865770-010547
  • Z-2656-2023 · Sep 8, 2023 · Open, Classified
    F2-01 Frame, physiological patient monitor (with arrhythmia detection or alarms)

Other Burst suppression detection software for electroencephalograph devices

All Burst suppression detection software for electroencephalograph devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page