Skip to content

Shockwave IVL System with the Shockwave S4 Peripheral IVL Catheter S4IVLK3040Recalls & reported problems

Shockwave Medical, Inc. · The Shockwave S4 Intravascular Lithotripsy (IVL) Catheter consists of an array of miniaturized lithotripsy emitters, located inside a balloon, that generate acoustic pressure waves within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. Two radiopaque marker bands inside the balloon aid in positioning. The distal catheter and balloon are hydrophilic coated to increase lubricity. The proximal hub has three ports: one for inflation/deflation of the balloon with 50/50 saline/contrast medium, one for the ‘Over-the-wire’ (OTW) guidewire lumen, and one for connection to the IVL Connector Cable (and IVL Generator). · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00195451000324), received between Apr 10, 2023 and Aug 26, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Material Rupture1
  • Material Split, Cut or Torn1

Patient problems

  • Cardiac Arrest1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 26, 2025Difficult to Remove; Detachment of Device or Device Component; Material Split, Cut or Torn; Difficult to Remove; Detachment of Device or Device Component; Material Split, Cut or Torn

    A SHOCKWAVE S4 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED AS PART OF AN ENDOVASCULAR TREATMENT OF THE RIGHT LOWER LIMB (TIBIAL RECANALIZATION VIA ANTEGRADE +/- RETROGRADE PUNCTURE OF THE PEDAL ARTERY). AFTER DELIVERING 2 SHOCKWAVE IVL PULSES, A PERFORATION OF THE BALLOON WAS OBSERVED. THE DEFLATED CATHETER REMAINED STUCK IN THE TIBIAL ARTERY. DESPITE STRONG TRACTION, THEY WERE UNABLE TO REMOVE THE CATHETER AND THE SHAFT BROKE. THE IVL BALLOON AND PART OF THE SHAFT REMAINED IN THE ARTERY. SINCE THE CHANCE TO REMOVE THE CATHETER TIP IMMEDIATELY WAS NOT AVAILABLE DUE TO THE PATIENT EXPERIENCING AN ARREST ON THE TABLE (UNAUTHORIZED TO THE BALLOON), IT WAS UNABLE TO BE RETRIEVED. IT WAS REPORTED THAT THE PATIENT SUFFERED A CARDIORESPIRATORY ARREST DUE TO ELEVATED BLOOD POTASSIUM LEVELS UNRELATED TO SHOCKWAVE. THEY MANAGED TO RESUSCITATE HIM AND DECIDED NOT TO CONTINUE THE PROCEDURE AND TO LEAVE THE BALLOON FRAGMENT IN THE TIBIAL ARTERY. THE PATIENT WAS TAKEN BACK 5 DAYS POST INDEX PROCEDURE TO ATTEMPT TO REMOVE THE FRAGMENT FROM THE ARTERY, BUT WITHOUT SUCCESS. AS A RESULT, THE PATIENT HAD TO REMAIN HOSPITALIZED. THE PATIENT IS REPORTEDLY DOING WELL.

    Patient problems: Cardiac Arrest; Foreign Body In Patient

    MAUDE report 22900374 ↗
  • InjuryApr 10, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT LEG PERIPHERAL INTERVENTION USING SHOCKWAVE S4 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER. DUE TO POOR KIDNEY FUNCTION THE PATIENT PROCEDURE HAD TO BE DONE WITH C02 (CONTRAINDICATION/OFF LABEL) RATHER THAN CONTRAST. AFTER THE CATHETER SUCCESSFULLY DELIVERED 80 PULSES, THE BALLOON RUPTURED IN THE ARTERIES BELOW THE KNEE. AFTER RUPTURE, A THROMBUS WAS OBSERVED. IT WAS NOT NOTICED EARLIER BECAUSE OF THE CO2, AND IT IS UNKNOWN WHETHER THE THROMBUS WAS A RESULT OF THE PROCEDURE OR NOT. THE VESSEL COULD NOT BE VISUALIZED AS WELL. AN IN-DWELLING CATHETER WITH TISSUE PLASMINOGEN ACTIVATOR (TPA) DRIP WAS LEFT INSIDE THE PATIENT'S VESSEL TO DISSOLVE THE THROMBUS. THE PATIENT WAS REQUIRED TO HAVE AN ADDITIONAL HOSPITAL STAY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Thrombosis/Thrombus

    MAUDE report 16707363 ↗

Other Cardiovascular Procedure Kit devices

All Cardiovascular Procedure Kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page