Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00195860013137), received between Jul 26, 2023 and Jan 8, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Misassembly by Users1

Patient problems

  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2024Misassembly by Users; Misassembly by Users

    THE PRIMARY SURGERY OCCURRED ON (B)(6) 2023. EVOLUTION WAS CONDUCTING A CADAVER LAB WITH THE DOCTOR (ON AN UNRELATED PROJECT) ON (B)(6) 2023 (DATE OF AWARENESS) AND HE MENTIONED THAT HE HAD A COVER COME OFF OF AN EMERGE PLATE. THE DOCTOR DID NOT PROVIDE ANY IMAGES OR ADDITIONAL DETAILS OTHER THAN THAT THE LOWEST PLATE COVER WAS SHOWN AS DISPLACED ON A ROUTINE FOLLOW-UP X-RAY. THE DOCTOR SAID THAT THE PATIENT IS FUSED, AND NO REVISION IS PLANNED.

    Patient problems: Failure of Implant

    MAUDE report 18470274 ↗
  • MalfunctionJul 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AN EMERGE COVER DISLODGED FROM THE PLATE AFTER THE SURGEON ATTEMPTED TO TURN THE COVER PAST A SCREW THAT WAS BEYOND THE ACCEPTABLE ANGLE. THE SCREW HEAD WAS TOO PROUD AND LED TO THE COVER DISLODGING. THE SURGEON THEN SIMPLY PLACED THE DISLODGED COVER BACK INTO THE PLATE AND CLOSED THE PATIENT. SINCE THE COVER WAS DISLODGED, THE STRENGTH OF THE COVER HAS BEEN COMPROMISED AND MAY INCREASE THE PROBABILITY OF OCCURRENCE OF SCREW BACKOUT. THE SURGEON SHOULD HAVE REPLACED THE PLATE WITH A NEW PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17396826 ↗

Other Appliance, fixation, spinal intervertebral body devices