ENFit PU Feeding Tube PFTL6.5P-NCRecalls & reported problems

Avanos Medical, Inc. · ENFit PU Feeding Tube Purp 6.5Fr 90cm · Class 2

All models of ENFit PU Feeding Tube (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00350770001807), received between May 7, 2022 and Jan 15, 2024 — classified as 7 injuries, 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K082238); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 7 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem7
  • Malposition of Device3
  • Material Puncture/Hole1

Patient problems

  • Bowel Perforation7
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 15, 2024Malposition of Device; Adverse Event Without Identified Device or Use Problem; Malposition of Device; Adverse Event Without Identified Device or Use Problem

    AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 3011270181-2024-00008 FOR THE FIRST EVENT. REFER TO 3011270181-2024-00009 FOR THE SECOND EVENT. IT WAS REPORTED A GASTRIC PERFORATION OCCURRED DURING USE OF THE NASOGASTRIC TUBE.

    Patient problems: Bowel Perforation

    MAUDE report 18515040 ↗
  • InjuryJan 15, 2024Malposition of Device; Adverse Event Without Identified Device or Use Problem; Malposition of Device; Adverse Event Without Identified Device or Use Problem

    AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 3011270181-2024-00008 FOR THE FIRST EVENT. REFER TO 3011270181-2024-00010 FOR THE THIRD EVENT. IT WAS REPORTED A GASTRIC PERFORATION OCCURRED DURING USE OF THE NASOGASTRIC TUBE.

    Patient problems: Bowel Perforation

    MAUDE report 18515039 ↗
  • InjuryJan 15, 2024Malposition of Device; Adverse Event Without Identified Device or Use Problem; Malposition of Device; Adverse Event Without Identified Device or Use Problem

    AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO (B)(4) FOR THE SECOND EVENT. REFER TO (B)(4) FOR THE THIRD EVENT. IT WAS REPORTED A GASTRIC PERFORATION OCCURRED DURING USE OF THE NASOGASTRIC TUBE.

    Patient problems: Bowel Perforation

    MAUDE report 18515038 ↗
  • InjuryJan 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT EVENTS. THIS IS THE FIRST OF FOUR REPORTS. REFER TO 3011270181-2024-00004 FOR THE SECOND EVENT. REFER TO 3011270181-2024-00005 FOR THE THIRD EVENT. REFER TO 3011270181-2024-00006 FOR THE FOURTH EVENT. IT WAS REPORTED THERE WAS AN EVENT RELATING TO GASTRIC PERFORATION IN THE PATIENT WHO ALSO HAD AN AVANOS FEEDING TUBE.

    Patient problems: Bowel Perforation

    MAUDE report 18511070 ↗
  • InjuryJan 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT EVENTS. THIS IS THE FOURTH OF FOUR REPORTS. REFER TO 3011270181-2024-00003 FOR THE FIRST EVENT. REFER TO 3011270181-2024-00004 FOR THE SECOND EVENT. REFER TO 3011270181-2024-00005 FOR THE THIRD EVENT. IT WAS REPORTED THERE WAS AN EVENT RELATING TO GASTRIC PERFORATION IN THE PATIENT WHO ALSO HAD AN AVANOS FEEDING TUBE.

    Patient problems: Bowel Perforation

    MAUDE report 18511068 ↗
  • InjuryJan 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT EVENTS. THIS IS THE SECOND OF FOUR REPORTS. REFER TO 3011270181-2024-00003 FOR THE FIRST EVENT. REFER TO 3011270181-2024-00005 FOR THE THIRD EVENT. REFER TO 3011270181-2024-00006 FOR THE FOURTH EVENT. IT WAS REPORTED THERE WAS AN EVENT RELATING TO GASTRIC PERFORATION IN THE PATIENT WHO ALSO HAD AN AVANOS FEEDING TUBE.

    Patient problems: Bowel Perforation

    MAUDE report 18511066 ↗
  • InjuryJan 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT EVENTS. THIS IS THE THIRD OF FOUR REPORTS. REFER TO 3011270181-2024-00003 FOR THE FIRST EVENT. REFER TO 3011270181-2024-00004 FOR THE SECOND EVENT. REFER TO 3011270181-2024-00006 FOR THE FOURTH EVENT. IT WAS REPORTED THERE WAS AN EVENT RELATING TO GASTRIC PERFORATION IN THE PATIENT WHO ALSO HAD AN AVANOS FEEDING TUBE.

    Patient problems: Bowel Perforation

    MAUDE report 18511063 ↗
  • MalfunctionMay 7, 2022Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THE PATIENT'S NASOGASTRIC TUBE DEVELOPED A 'PINPOINT HOLE' MIDWAY OF TUBE AFTER 2 WEEKS OF BEING IN PLACE. THE DEVICE WAS REPLACED. THERE WAS NO REPORTED INJURY ADDITIONAL INFORMATION RECEIVED STATED THE 'DAY CARE NOTICED WATER FROM TUBE WHEN FLUSHING.'

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14315372 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K082238. One submission often covers many models, so check the product and lot details in each record.

  • Z-0093-2019 · Jun 4, 2018 · Terminated
    NeoMed Enteral Feeding Tubes with Enteral Only Connectors, French sizes 4.0Fr, 5.0Fr, 6.5Fr, 8.0Fr, and lengths 40cm, 60cm, 90cm. The product consists of a polyurethane feeding tu

Other TUBE, NASOGASTRIC devices

All TUBE, NASOGASTRIC devices →