CORGRIP* SR NG/NI Tube Retention System 26-010Recalls & reported problems
Avanos Medical, Inc. · CORGRIP* SR Nasogastric/Nasointestinal Feeding Tube Retention System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00350770005614), received between Jul 10, 2024 and Mar 17, 2026 — classified as 2 injuries, 2 malfunctions.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Hemorrhage/Blood Loss/Bleeding1
- Localized Skin Lesion1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2026
NEW BRIDLE IS LESS FLEXIBLE THAN PREVIOUS BRAND, STRINGS TO TIE IN A KNOT ARE WORSE QUALITY AND DO NOT STAY TIED, POSING A RISK FOR DISLODGMENT. OVERALL SEEMS WORSE QUALITY AND HARDER TO INSERT AND UTILIZE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616674 ↗MalfunctionMar 13, 2026
THE RETENTION SYSTEM HAS 2 STYLETS (YELLOW & WHITE), EACH WITH MAGNETS AT THE TIP. MAGNET FROM YELLOW STYLET CAME OFF AS IT WS BEING PULLED FROM PATIENT. MAGNET COULD HAVE FALLEN INSIDE NOSTRIL & TRAVEL INTO THE LUNGS, IT WASN'T SUPPOSED TO COME OFF FROM YELLOW STYLET. NORMALLY, MAGNET FROM WHITE TYLET WILL COME OFF & ATTACH TO YELLOW STYLET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24589343 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, CORGRIP SR PLACED ON THE (B)(6) 2024. KNOT BELOW CORGRIP SR IS COMPLETELY LOOSE. INSPECTION OF THE NOSTRIL AND SEE A LITTLE PRESSURE WOUND RIGHT BELOW THE COLUMELLA WHERE THE POINT OF THE CLAMP HAS PRESSED AGAINST THE NOSE. FURTHERMORE, A DEEPER, SKINNED LESION BELOW ON THE LOWER SIDE OF THE COLUMELLA WHERE THE CORGRIP SR TETHER HAS CUT THROUGH THE SKIN. CORGRIP SR HAS BEEN REMOVED AND NASOGASTRIC (NG) TUBE IS FIXATED WITH NOSE PLASTER. DEVICE WAS IN USE FOR FOUR DAYS. NO SEVERE PATIENT HARM. PATIENT IS STILL IN ICU FOR INJURIES NOT RELATED TO THIS INCIDENT.
Patient problems: Localized Skin Lesion
MAUDE report 19712387 ↗InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, CORGRIP CAUSED BLEEDING AND PAIN. DEVICE WAS REMOVED A DAY LATER. NO SERIOUS INJURY OR MEDICAL INTERVENTIONS REPORTED. PER ADDITIONAL INFORMATION RECEIVED ON 21JUN2024, THE PATIENT WAS REPORTED TO BE ¿GOOD.¿ THE DEVICE WAS PLACED ON (B)(6) 2024 AND WAS REMOVED ON (B)(6) 2024.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain
MAUDE report 19711560 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page