Skip to content

AVANOS 44099Recalls & reported problems

Avanos Medical, Inc. · MIC-KEY* Gastric-Jejunal Feeding Tube Kit Extension Sets · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00350770440996), received between Apr 8, 2026 and Apr 24, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Complete Blockage” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 31, 2024Open, Classified

    The IFU was missing some risk information requested by another country to be contained in the IFU pertaining to potential complications.

    Z-2422-2024 · Root cause: Error in labeling · FDA record ↗

Most reported problems

Device problems

  • Complete Blockage2
  • Break1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Vomiting1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2026Break; Complete Blockage

    IT WAS REPORTED THAT THERE WAS RESISTANCE WHEN FLUSHING AN NG TUBE. THEN ALL OF THE SUDDEN THE RESISTANCE WAS GONE. NO VOMITING BUT AGAIN, THEY COULD ASPIRATE FEED FROM GASTRIC PORT. LEAK INTO STOMACH CONFIRMED WHEN INJECTING CONTRAST INTO THE JEJUNAL LIMB UNDER X-RAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24982902 ↗
  • InjuryApr 8, 2026Complete Blockage; Fluid/Blood Leak

    IT WAS REPORTED THAT A PATIENT STARTED VOMITING FEED. THEY WERE ABLE TO ASPIRATE FEED WHICH WAS GIVEN VIA JEJUNAL PORT FROM THE GASTRIC PORT. UPON INJECTION OF THE JEJUNAL PORT UNDER X-RAY, THERE WAS CONTRAST LEAKAGE AROUND THE STOMA SITE AND INTO THE STOMACH IN KEEPING WITH JEJUNAL LIMB FRACTURE. MUM DESCRIBES ISSUES OCCURRED AFTER FORCEFULLY FLUSHING A TUBE BLOCKAGE.

    Patient problems: Vomiting

    MAUDE report 24821135 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.