Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00350770441092), received between Dec 31, 2020 and Jan 5, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 31, 2024Open, Classified
The IFU was missing some risk information requested by another country to be contained in the IFU pertaining to potential complications.
Z-2422-2024 · Root cause: Error in labeling · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak2
- Material Split, Cut or Torn2
- Material Integrity Problem1
Patient problems
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED, A PATIENT HAD A HALYARD GASTRIC-JEJUNAL FEEDING TUBE (GJ TUBE), 18FR 1.7CM 30CM TUBE THAT ¿BROKE OFF¿. (B)(6)2022, TO ¿GRAB¿ THE BROKEN TUBE FROM THE SMALL INTESTINE, WHERE IT WAS REPLACED BY THE DOCTOR. THERE WAS NO REPORTED INJURY TO THE PATIENT. ADDITIONAL INFORMATION RECEIVED (B)(6)2022 REPORTED, INTERVENTIONAL RADIOLOGY (IR) PLACED THE GJ TUBE ON (B)(6)2022 ; THE TUBE WAS BEING REPLACED WHEN THE ¿BREAK¿ WAS DISCOVERED. AN ENDOSCOPY, WITH GENERAL ANESTHESIA, WAS PERFORMED TO REMOVE THE BROKEN GJ TUBE AND TO PLACE THE NEW ONE. THERE WAS NO REPORTED INJURY.
Patient problems: Foreign Body In Patient
MAUDE report 16104772 ↗MalfunctionFeb 3, 2021Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS INITIALLY REPORTED THAT THE DEVICE WAS CHANGED DUE TO LEAKING. THE DEVICE WAS RETURNED AND ON (B)(6) 2021 IT WAS IDENTIFIED THAT THE INNER WALL OF THE TUBING WHICH SEPARATES THE JEJUNAL AND GASTRIC LUMENS HAD TORN, ALLOWING FOR INNER LUMEN CROSSOVER. ADDITIONAL INFORMATION RECEIVED 21-JAN-2021 INDICATED THERE WAS NO PATIENT HARM. NO FURTHER DETAILS WERE AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11276010 ↗MalfunctionDec 31, 2020Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS REPORTED THAT THE FEEDING TUBE WAS CHANGED DUE TO LEAKING. INJECTION OF CONTRAST BY RADIOLOGIST VIA JEJUNAL PORT SHOWED CONTRAST ENTERING THE STOMACH. THE TUBE WAS FOUND TO BE SPLIT. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11102567 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.