AVANOS 44561Recalls & reported problems
Avanos Medical, Inc. · * MIC* Bolus Gastrostomy Feeding Tube with ENFit® Connector · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00350770445618), received between Feb 4, 2026 and Apr 2, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Expulsion1
- Material Rupture1
- Migration1
Patient problems
- Abnormal Blood Gases1
- Feeding Problem1
- No Clinical Signs, Symptoms or Conditions1
- Peritonitis1
- Septic Shock1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 2, 2026Migration
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING ONE PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 9611594-2026-00167 FOR THE FIRST REPORT. IT WAS REPORTED THAT A "PATIENT HAD A NEW GASTROSTOMY TUBE INSERTED, AS REPLACEMENT, ON (B)(6) 2026, WITH RADIOLOGICAL CONTROL. ON THE MORNING OF (B)(6) 2026, WHILE ENTERAL FEEDING WAS BEING RESUMED, A STATE OF HYPOVIGILANCE OCCURRED. HEMODYNAMICS WERE STABLE. THERE WAS DIFFUSE ABDOMINAL CONTRACTURE. ARTERIAL BLOOD GAS ANALYSIS REVEALED A NORMAL PH BUT HYPERLACTATEMIA AT 9 MMOL/1. AN EMERGENCY ABDOMINAL CT SCAN REVEALED POOR POSITIONING OF THE GASTROSTOMY BALLOON, WHICH ENDED IN THE OMENTAL FAT, WITH THE PRESENCE OF A SHEET OF HEMOPERITONEUM SPREADING INTO THE DOUGLAS POUCH AND SUPRAMESOCOLIC PNEUMOPERITONEUM. THE PATIENT IS TRANSFERRED TO THE MEDICAL INTENSIVE CARE UNIT AND THEN TAKEN OVER BY THE DIGESTIVE SURGERY TEAM ON THE SAME DAY. PERITONITIS IN ALL FOUR QUADRANTS WAS CONFIRMED WITH THE PRESENCE OF ENTERAL FEEDING FLUID ASSOCIATED WITH FALSE MEMBRANES IN THE PELVIS. THE GASTROSTOMY TUBE WAS FOUND IN THE FREE PERITONEUM AND NOT IN THE STOMACH.
Patient problems: Abnormal Blood Gases; Peritonitis; Septic Shock
MAUDE report 24765645 ↗MalfunctionApr 2, 2026Material Rupture; Expulsion
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING ONE PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 9611594-2026-00167 FOR THE FIRST REPORT. REFER TO 9611594-2026-00168 FOR THE SECOND REPORT. IT WAS REPORTED THAT AN 82-YEAR-OLD PATIENT BEING TREATED FOR SPONDYLODISCITIS UNDERWENT GASTROSTOMY TUBE PLACEMENT ON (B)(6) 2026. A FEW HOURS LATER, THE TUBE WAS FOUND IN HIS BED WITH THE BALLOON BURST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765605 ↗MalfunctionFeb 4, 2026Detachment of Device or Device Component
IT WAS REPORTED, ¿A 70-YEAR-OLD PATIENT UNDERWENT GASTROSTOMY PLACEMENT ON (B)(6) 2025. UPON RETURNING HOME ON (B)(6) 2025, THE FEEDING TUBE COMPLETELY DETACHED FROM THE INSIDE. CURRENT PATIENT STATUS: THE PATIENT HAS BEEN RE-HOSPITALIZED TO CLOSE THE GASTROSTOMY OPENING.¿
Patient problems: Feeding Problem
MAUDE report 24259127 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page