Skip to content

AVANOS 44561Recalls & reported problems

Avanos Medical, Inc. · * MIC* Bolus Gastrostomy Feeding Tube with ENFit® Connector · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00350770445618), received between Feb 4, 2026 and Apr 2, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Expulsion1
  • Material Rupture1
  • Migration1

Patient problems

  • Abnormal Blood Gases1
  • Feeding Problem1
  • No Clinical Signs, Symptoms or Conditions1
  • Peritonitis1
  • Septic Shock1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2026Migration

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING ONE PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 9611594-2026-00167 FOR THE FIRST REPORT. IT WAS REPORTED THAT A "PATIENT HAD A NEW GASTROSTOMY TUBE INSERTED, AS REPLACEMENT, ON (B)(6) 2026, WITH RADIOLOGICAL CONTROL. ON THE MORNING OF (B)(6) 2026, WHILE ENTERAL FEEDING WAS BEING RESUMED, A STATE OF HYPOVIGILANCE OCCURRED. HEMODYNAMICS WERE STABLE. THERE WAS DIFFUSE ABDOMINAL CONTRACTURE. ARTERIAL BLOOD GAS ANALYSIS REVEALED A NORMAL PH BUT HYPERLACTATEMIA AT 9 MMOL/1. AN EMERGENCY ABDOMINAL CT SCAN REVEALED POOR POSITIONING OF THE GASTROSTOMY BALLOON, WHICH ENDED IN THE OMENTAL FAT, WITH THE PRESENCE OF A SHEET OF HEMOPERITONEUM SPREADING INTO THE DOUGLAS POUCH AND SUPRAMESOCOLIC PNEUMOPERITONEUM. THE PATIENT IS TRANSFERRED TO THE MEDICAL INTENSIVE CARE UNIT AND THEN TAKEN OVER BY THE DIGESTIVE SURGERY TEAM ON THE SAME DAY. PERITONITIS IN ALL FOUR QUADRANTS WAS CONFIRMED WITH THE PRESENCE OF ENTERAL FEEDING FLUID ASSOCIATED WITH FALSE MEMBRANES IN THE PELVIS. THE GASTROSTOMY TUBE WAS FOUND IN THE FREE PERITONEUM AND NOT IN THE STOMACH.

    Patient problems: Abnormal Blood Gases; Peritonitis; Septic Shock

    MAUDE report 24765645 ↗
  • MalfunctionApr 2, 2026Material Rupture; Expulsion

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING ONE PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 9611594-2026-00167 FOR THE FIRST REPORT. REFER TO 9611594-2026-00168 FOR THE SECOND REPORT. IT WAS REPORTED THAT AN 82-YEAR-OLD PATIENT BEING TREATED FOR SPONDYLODISCITIS UNDERWENT GASTROSTOMY TUBE PLACEMENT ON (B)(6) 2026. A FEW HOURS LATER, THE TUBE WAS FOUND IN HIS BED WITH THE BALLOON BURST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765605 ↗
  • MalfunctionFeb 4, 2026Detachment of Device or Device Component

    IT WAS REPORTED, ¿A 70-YEAR-OLD PATIENT UNDERWENT GASTROSTOMY PLACEMENT ON (B)(6) 2025. UPON RETURNING HOME ON (B)(6) 2025, THE FEEDING TUBE COMPLETELY DETACHED FROM THE INSIDE. CURRENT PATIENT STATUS: THE PATIENT HAS BEEN RE-HOSPITALIZED TO CLOSE THE GASTROSTOMY OPENING.¿

    Patient problems: Feeding Problem

    MAUDE report 24259127 ↗

Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices

All TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices →

Sources and updates