AVANOS 44567Recalls & reported problems
Avanos Medical, Inc. · MIC* Jejunal Feeding Tube with ENFit® Connector · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00350770445670), received between Sep 29, 2022 and Dec 1, 2023 — classified as 1 death, 1 injury.
The most frequently reported specific device problem is “Misconnection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Misconnection1
Patient problems
- Insufficient Information1
- Sepsis1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 1, 2023Misconnection; Misconnection
IT WAS REPORTED THE PATIENT UNDERWENT A RADICAL TOTAL GASTRECTOMY ON (B)(6) 2022, WITH JEJUNOSTOMY FEEDING TUBE INSERTED INTO THE JEJUNUM DIRECTLY THROUGH THE SKIN ON THE ABDOMEN. THE PATIENT WAS "DIAGNOSED WITH INTRAABDOMINAL SEPSIS AND SMALL BOWEL PRESSURE NECROSIS CAUSED BY PRESSURE FROM THE FEEDING JEJUNOSTOMY BALLOON AND 21ML OF FLUID WAS FOUND IN THE BALLOON." A COMPUTED TOMOGRAPHY (CT) WAS COMPLETED THAT WAS HIGHLY SUSPICIOUS FOR BOWEL ISCHEMIA. PATIENT HAD A LAPAROTOMY, WASHOUT, REMOVAL OF FEEDING JEJUNOSTOMY WITH SMALL BOWEL RESECTION AND ANASTOMOSIS. A REVIEW FOUND THAT AT SOME TIME DURING ADMISSION (TIME UNABLE TO BE DETERMINED) A SYRINGE WAS INSERTED INTO THE BALLOON PORT INSTEAD OF THE FEEDING PORT AND A WATER FLUSH ENTERED THE BALLOON, INFLATING IT TO 21MLS. THE INFLATED BALLOON WAS IMMOBILE AND EXERTED PRESSURE ON THE SMALL INTESTINE WALL WHICH CAUSED SMALL BOWEL PRESSURE NECROSIS AND LOCALIZED PERITONITIS.
Patient problems: Sepsis; Unspecified Infection
MAUDE report 18245590 ↗DeathSep 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE FEEDING TUBE WAS PLACED, AND THE PATIENT REMAINED IN THE HOSPITAL DUE TO OTHER NEEDS. IT WAS ADDITIONALLY REPORTED, THE FEEDING TUBE WAS FLUSHED UNDER PRESSURE AND HAD A LEAK JUST BELOW THE CONNECTOR ON THE TUBE; THE REPORTER CONFIRMED THERE WAS NO BLOCKAGE IN THE TUBE. THE PATIENT REPORTEDLY DIED. ADDITIONAL INFORMATION RECEIVED 09SEP2022 REPORTED, THE DEVICE REMAINED INSIDE THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 15505621 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page