Skip to content

AVANOS 44567Recalls & reported problems

Avanos Medical, Inc. · MIC* Jejunal Feeding Tube with ENFit® Connector · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00350770445670), received between Sep 29, 2022 and Dec 1, 2023 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Misconnection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Misconnection1

Patient problems

  • Insufficient Information1
  • Sepsis1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 1, 2023Misconnection; Misconnection

    IT WAS REPORTED THE PATIENT UNDERWENT A RADICAL TOTAL GASTRECTOMY ON (B)(6) 2022, WITH JEJUNOSTOMY FEEDING TUBE INSERTED INTO THE JEJUNUM DIRECTLY THROUGH THE SKIN ON THE ABDOMEN. THE PATIENT WAS "DIAGNOSED WITH INTRAABDOMINAL SEPSIS AND SMALL BOWEL PRESSURE NECROSIS CAUSED BY PRESSURE FROM THE FEEDING JEJUNOSTOMY BALLOON AND 21ML OF FLUID WAS FOUND IN THE BALLOON." A COMPUTED TOMOGRAPHY (CT) WAS COMPLETED THAT WAS HIGHLY SUSPICIOUS FOR BOWEL ISCHEMIA. PATIENT HAD A LAPAROTOMY, WASHOUT, REMOVAL OF FEEDING JEJUNOSTOMY WITH SMALL BOWEL RESECTION AND ANASTOMOSIS. A REVIEW FOUND THAT AT SOME TIME DURING ADMISSION (TIME UNABLE TO BE DETERMINED) A SYRINGE WAS INSERTED INTO THE BALLOON PORT INSTEAD OF THE FEEDING PORT AND A WATER FLUSH ENTERED THE BALLOON, INFLATING IT TO 21MLS. THE INFLATED BALLOON WAS IMMOBILE AND EXERTED PRESSURE ON THE SMALL INTESTINE WALL WHICH CAUSED SMALL BOWEL PRESSURE NECROSIS AND LOCALIZED PERITONITIS.

    Patient problems: Sepsis; Unspecified Infection

    MAUDE report 18245590 ↗
  • DeathSep 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE FEEDING TUBE WAS PLACED, AND THE PATIENT REMAINED IN THE HOSPITAL DUE TO OTHER NEEDS. IT WAS ADDITIONALLY REPORTED, THE FEEDING TUBE WAS FLUSHED UNDER PRESSURE AND HAD A LEAK JUST BELOW THE CONNECTOR ON THE TUBE; THE REPORTER CONFIRMED THERE WAS NO BLOCKAGE IN THE TUBE. THE PATIENT REPORTEDLY DIED. ADDITIONAL INFORMATION RECEIVED 09SEP2022 REPORTED, THE DEVICE REMAINED INSIDE THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 15505621 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page